Why X-ray fluorescence depends on sample thickness

Two materials with the same copper fraction can produce different copper peaks. An atom must first absorb an incident photon and lose an inner electron. Its fluorescent photon must then cross the surrounding material. The detector counts the photons that complete that journey.

Sandbox Physics ·

Atoms emit photons at characteristic energies

Shine X-rays on an alloy and an incident photon can eject an inner electron. Another electron falls into the vacancy. The atom may release the energy difference as a new X-ray photon. This process is X-ray fluorescence, or XRF. Different elements have different electron energy levels, so their emitted photons collect around different energies.

An energy-sensitive detector sorts the arriving photons into a spectrum. Iron and copper produce peaks in different places. Peak positions help identify the elements; the counts under a peak describe how many photons were received. One element can produce several lines because electrons from different shells can fill the same inner vacancy. Naming the peaks starts the composition analysis.

Sources: [1][2]

The incident photon must cross an energy threshold

A prominent copper line lies near eight kiloelectronvolts, but removing a copper K-shell electron takes nearly nine. The emitted photon carries the difference between two shell energies. The threshold for removing the inner electron is called an absorption edge. Crossing it sharply increases the probability of creating that vacancy.

At an incident energy of eight kiloelectronvolts, iron can emit K-series fluorescence and copper cannot. A small fraction of harder photons can make copper fluoresce again. Comparing spectra therefore requires comparing the energy distributions of the incident beams. Even after a vacancy is created, an atom can relax through another process, so fluorescent emission has a probability below one.

Sources: [2][1]

Fluorescence from greater depth is harder to collect

Imagine adding more of the same material to a thin film. At first, doubling the material roughly doubles the number of atoms available to emit. Few incident or fluorescent photons are lost inside the layer. The line strength is nearly proportional to elemental mass per unit area, making known thin standards useful comparison scales.

With more material, the incident beam loses photons before reaching deeper atoms. Fluorescence from those atoms suffers further absorption on its outward path. Added material contributes less and less until the line intensity approaches saturation. Rigaku’s thin-film analysis guide uses this behaviour to distinguish thin-film and bulk ranges. The relevant thickness depends on the line and the surrounding material.

Adding another material changes how far the photons travel even if the iron-to-copper ratio stays fixed. Absorption and additional excitation caused by other components are called matrix effects. Photons at different energies suffer different losses. An absorption edge can even make a higher-energy fluorescent line more strongly absorbed by another element, shifting the peak ratio. Moving the collector toward a grazing direction also lengthens the outward path and can amplify the difference.

Sources: [3]

Known standards connect counts to content

Standards with known compositions connect detected counts to elemental signals. Dark and blank measurements remove electronic and holder backgrounds before peak comparison. The detector window absorbs some photons and its silicon layer captures only some of those transmitted. The sensitivity determined with standards includes these responses. Broad peaks that overlap can make the allocation of counts between elements unstable.

For a thicker sample, a calculation can follow fluorescent production and escape through depth before comparison with the standards. It needs assumptions about thickness, composition distribution and collection geometry. Known thickness and matrix help correct absorption in a homogeneous, single-layer primary-fluorescence approximation. More complex samples can include fluorescence from one element exciting another, which requires extending the calculation.

A bright peak therefore invites practical questions: was the excitation the same, what material did the photons cross, and which standard provides the comparison? Answering them turns peak strength into a conditional measurement of content.

Sources: [3][4]

Sources and further reading

  1. Bruker — How does XRF work?

    The photon-excitation and fluorescent-relaxation explanation gives the physical sequence.

  2. NIST — X-ray Transition Energies Database

    Element transition tables distinguish emitted line energies from the K absorption edge.

  3. Hikari Takahara — Thickness and composition analysis of thin film samples using FP method by XRF analysis, Rigaku Journal33(2),2017,17–21

    Printed pages17–19 explain the thin linear range, bulk saturation, matrix-dependent depth and standard-based spectrometer sensitivity.

  4. Amptek — FAST SDD detector specifications

    Detector structure, window transmission, silicon absorption and energy resolution explain the final counted spectrum.

Keep reading

What this page links to directly

See this series on the content map →

How physics instruments measure · Reading path

How physics instruments measure

  1. ArticleHow a vernier reads between millimetres
  2. ArticleAir helps hold up the object on a balance
  3. ArticleHow Cavendish Measured the Attraction Between Spheres
  4. ArticleWhich pressure does a liquid column measure?
  5. ArticleThe cup absorbs heat too
  6. ArticleRotational rheometers need both torque and speed
  7. ArticleWhy the air beyond a pipe changes its resonance
  8. ArticleHow a seismometer reads the ground through motion inside its case
  9. ArticleWhy does a meter change the circuit?
  10. ArticleHow a null detector reveals a resistance
  11. ArticleWhy a capacitor can behave like an inductor
  12. ArticleWhere does the slow wave on an oscilloscope come from?
  13. ArticleHow a lock-in amplifier finds a signal buried in noise
  14. ArticleWhere a small peak in a spectrum can come from
  15. ArticleHow a network analyzer separates the cable from the device
  16. ArticleWhen a slower quartz crystal can measure added mass
  17. ArticleHow an echelle spectrograph gives overlapping colors a second direction
  18. ArticleHow total brightness reveals many infrared colors
  19. ArticleMeasuring a Solution’s Concentration with Light
  20. ArticleHow an ellipsometer reads thin-film thickness from reflected light
  21. ArticleHow a polarimeter reveals mixed polarization
  22. ArticleHow individual photons reveal a fluorescence lifetime
  23. ArticleBrillouin Microscopy Listens to Sound with Light
  24. ArticleHow a white sphere compares a lamp’s total output
  25. ArticleHow powder rings reveal atomic spacing
  26. ArticleHow Small-Angle X-ray Scattering Reads a Particle Halo
  27. You are hereWhy X-ray fluorescence depends on sample thickness
  28. ArticleWhy an AFM makes a bump look wider
  29. ArticleHow a small indentation separates hardness from elasticity
  30. ArticleHow Differential Scanning Calorimetry Reads a Melting Peak
  31. ArticleHow thermogravimetry separates mass loss from the gas around it
  32. ArticleHow a diamond anvil cell reads pressure in ruby light
  33. ArticleHow diamond fluorescence reveals a magnetic-field direction
  34. ArticleReading a magnetic field in the rhythm of atoms
  35. ArticleKeeping a laser in time with atoms
  36. ArticleWhy a Kibble balance needs two experiments
  37. ArticleHow a dilution refrigerator keeps carrying heat away
  38. ArticleWhy a bigger vacuum pump may barely help the sample
  39. ArticleHow microwaves read a superconducting qubit
  40. ArticleHow current leaves a voltage across a thin sheet
  41. ArticleHow an ion flight becomes a mass spectrum
  42. ArticleHow four straight rods select ion mass
  43. ArticleHow a rapidly reversing field can hold ions
  44. ArticleHow a cyclotron keeps a proton on the accelerating beat
  45. ArticleWhy the drift tubes in an RF linac get longer
  46. ArticleThe directions hidden inside a small beam spot
  47. ArticleHow defocusing magnets help a beam stay in a ring
  48. ArticleWhy a collider cannot keep every collision
  49. ArticleGetting synchrotron light onto a sample
  50. ArticleHow a free-electron laser gathers electrons into a brighter beam
  51. ArticleFollowing energy exchange with neutron flight times
  52. ArticleHow a tokamak holds heat and measures where it goes
  53. ArticleWhy stellarator coils bend in three dimensions
  54. ArticleHow NIF turns a laser pulse into an implosion
  55. ArticleHow ELI brings an intense pulse to a target
  56. ArticleWhy a xenon detector needs liquid and gas
  57. ArticleHow LIGO measures a gravitational wave with mirrors that move
  58. ArticleHow delayed laser records let LISA measure gravitational waves
  59. ArticleHow FAST brings faint radio signals into focus
  60. ArticleHow ALMA keeps separated antennas in step
  61. ArticleHow SKA-Low steers a beam without moving its antennas
  62. ArticleHow EHT connects telescopes with clocks and hard drives
  63. ArticleHow Webb separates a faint spectrum from infrared background
  64. ArticleHow LHAASO finds an incoming direction from ground particles
Something broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy