When a slower quartz crystal can measure added mass

A thin rigid coating moves with a quartz surface, adds inertia and lowers its shear resonance. A quartz crystal microbalance can use that change to measure very small deposits. Water and compliant attachments also change the motion, so a lower frequency needs a suitable blank and a measure of loss before it supports a mass estimate.

Sandbox Physics ·

A coating must turn around with the surface

Place electrodes on opposite faces of a quartz plate and apply an alternating voltage. Its piezoelectric response repeatedly deforms the crystal. With a suitable cut, the two faces move sideways in opposite directions. This thickness-shear motion resembles a thin stack of paper whose top and bottom slide opposite ways. A common sensor repeats the motion about five million times each second.

A uniform, rigidly attached coating must accelerate whenever the surface reverses direction. The elastic restoring action remains, but the moving inertia increases, so the resonance becomes slower. The coating must be thin and stiff enough to follow the quartz. A loosely attached particle or a compliant connection can have relative motion of its own.

Sources: [1]

Find the mechanical resonance in the current

A voltage drives a larger mechanical response near resonance. An instrument can sweep a narrow frequency interval and save the current at each point. It also records whether the current peak arrives before or after the voltage peak. This position within a cycle is its phase. Current size and phase together preserve the shape of the response.

The electrodes also charge capacitively. That current can flow without driving the same mechanical motion, and it contributes to the total reading. An equivalent circuit separates a motional path representing inertia, elasticity and loss from a parallel capacitance representing electrodes and wiring. Fitting the two current components avoids identifying the largest total current with the mechanical series resonance.

The peak position identifies a preferred frequency. Its width reflects loss: more rapid decay generally gives a wider resonance. Counting frequency precisely is therefore only part of a mass measurement.

Sources: [1]

One microgram per square centimetre gives about 56.6 hertz

For a five-megahertz crystal, adding a uniform rigid film of one microgram per square centimetre lowers frequency by about 56.6 hertz under the Sauerbrey relation. A microgram is a millionth of a gram. This is mass per area; converting it to total mass also needs the effective coated area.

Write the added areal mass as mAm_A, the positive sensitivity as CfC_f, and the frequency change as Δf\Delta f. The relation below expresses the downward shift caused by added rigid inertia.

A thicker film can deform in shear itself. Uneven coverage can also weight some surface regions more strongly than others. Even for a small frequency shift, the conversion requires a thin, stiff coating that follows the surface.

Δf=−CfmA\Delta f=-C_f m_A
Frequency change uses hertz and areal mass uses micrograms per square centimetre. The sensitivity is about 56.6 hertz per unit areal mass for this five-megahertz AT-cut quartz. The linear conversion assumes a uniform thin, rigidly attached load.

Sources: [1][2]

Water slows a bare crystal too

Let one face of a bare crystal touch water. The moving surface drags nearby liquid sideways. More distant liquid follows less closely, and viscosity dissipates energy within that shear layer. The crystal is now moving against more than a solid deposited film.

At twenty degrees Celsius, ideal one-face water loading lowers a five-megahertz resonance by about 715 hertz. Applying the dry-film sensitivity would suggest about 12.6 micrograms per square centimetre of added material even though no film was deposited. This liquid model depends on the product of density and viscosity, so one response cannot determine them separately.

The resonance also broadens, providing another clue. Compliant attachments can change both coupled inertia and loss. Measuring position and width helps select a load model. Separating mass from mechanical properties requires further frequencies or independent information about the material.

Sources: [1][2]

The blank must describe comparable conditions

The crystal frequency itself changes with temperature. Clean-crystal scans at several temperatures provide a baseline for the sample temperature and can remove calibrated static drift. The current receiver needs known references for scale and phase, and the frequency clock needs its own scale reference. Saving those standards with the raw sweep gives the measured shift a clear origin.

Mass measurement in liquid remains possible. If the liquid properties and temperature stay fixed and an added layer remains sufficiently rigid, a matched liquid blank and a suitable load model can support a mass increment. Changing liquid, swelling films or relative motion require another interpretation. The instrument measures how a surface drags its load; the mass estimate rests on that physical picture.

Sources: [1]

Sources and further reading

  1. SRS QCM200 / QCM25 Operation and Service Manual · Revision 2.5 (6/2018)

    Printed pp.15–16 introduce rigid-film sensitivity and the BVD circuit; Eq.12 on p.27 gives Sauerbrey sensitivity, Eq.15 on p.29 gives liquid loading and p.30 gives the approximately 715 Hz water example. Pages35–37 discuss rigid increments and viscoelastic changes in solution; pp.42 and49–55 describe dissipation, thickness shear and temperature. These adopted sections were checked on 28 October 2026.

  2. openQCM Advanced Calculator · Newtonian liquid model and model-selection guide

    The Sauerbrey and Kanazawa–Gordon guide sections distinguish rigid gas-phase films from one-face Newtonian liquid loading. The liquid section states that frequency shift and half-bandwidth increment have equal magnitudes and identifies density times viscosity. Consulted 28 October 2026; no stable page revision is exposed. Its automatic multi-parameter inversion claims are not adopted here.

Keep reading

What this page links to directly

See this series on the content map →

How physics instruments measure · Reading path

How physics instruments measure

  1. ArticleHow a vernier reads between millimetres
  2. ArticleAir helps hold up the object on a balance
  3. ArticleHow Cavendish Measured the Attraction Between Spheres
  4. ArticleWhich pressure does a liquid column measure?
  5. ArticleThe cup absorbs heat too
  6. ArticleRotational rheometers need both torque and speed
  7. ArticleWhy the air beyond a pipe changes its resonance
  8. ArticleHow a seismometer reads the ground through motion inside its case
  9. ArticleWhy does a meter change the circuit?
  10. ArticleHow a null detector reveals a resistance
  11. ArticleWhy a capacitor can behave like an inductor
  12. ArticleWhere does the slow wave on an oscilloscope come from?
  13. ArticleHow a lock-in amplifier finds a signal buried in noise
  14. ArticleWhere a small peak in a spectrum can come from
  15. ArticleHow a network analyzer separates the cable from the device
  16. You are hereWhen a slower quartz crystal can measure added mass
  17. ArticleHow an echelle spectrograph gives overlapping colors a second direction
  18. ArticleHow total brightness reveals many infrared colors
  19. ArticleMeasuring a Solution’s Concentration with Light
  20. ArticleHow an ellipsometer reads thin-film thickness from reflected light
  21. ArticleHow a polarimeter reveals mixed polarization
  22. ArticleHow individual photons reveal a fluorescence lifetime
  23. ArticleBrillouin Microscopy Listens to Sound with Light
  24. ArticleHow a white sphere compares a lamp’s total output
  25. ArticleHow powder rings reveal atomic spacing
  26. ArticleHow Small-Angle X-ray Scattering Reads a Particle Halo
  27. ArticleWhy X-ray fluorescence depends on sample thickness
  28. ArticleWhy an AFM makes a bump look wider
  29. ArticleHow a small indentation separates hardness from elasticity
  30. ArticleHow Differential Scanning Calorimetry Reads a Melting Peak
  31. ArticleHow thermogravimetry separates mass loss from the gas around it
  32. ArticleHow a diamond anvil cell reads pressure in ruby light
  33. ArticleHow diamond fluorescence reveals a magnetic-field direction
  34. ArticleReading a magnetic field in the rhythm of atoms
  35. ArticleKeeping a laser in time with atoms
  36. ArticleWhy a Kibble balance needs two experiments
  37. ArticleHow a dilution refrigerator keeps carrying heat away
  38. ArticleWhy a bigger vacuum pump may barely help the sample
  39. ArticleHow microwaves read a superconducting qubit
  40. ArticleHow current leaves a voltage across a thin sheet
  41. ArticleHow an ion flight becomes a mass spectrum
  42. ArticleHow four straight rods select ion mass
  43. ArticleHow a rapidly reversing field can hold ions
  44. ArticleHow a cyclotron keeps a proton on the accelerating beat
  45. ArticleWhy the drift tubes in an RF linac get longer
  46. ArticleThe directions hidden inside a small beam spot
  47. ArticleHow defocusing magnets help a beam stay in a ring
  48. ArticleWhy a collider cannot keep every collision
  49. ArticleGetting synchrotron light onto a sample
  50. ArticleHow a free-electron laser gathers electrons into a brighter beam
  51. ArticleFollowing energy exchange with neutron flight times
  52. ArticleHow a tokamak holds heat and measures where it goes
  53. ArticleWhy stellarator coils bend in three dimensions
  54. ArticleHow NIF turns a laser pulse into an implosion
  55. ArticleHow ELI brings an intense pulse to a target
  56. ArticleWhy a xenon detector needs liquid and gas
  57. ArticleHow LIGO measures a gravitational wave with mirrors that move
  58. ArticleHow delayed laser records let LISA measure gravitational waves
  59. ArticleHow FAST brings faint radio signals into focus
  60. ArticleHow ALMA keeps separated antennas in step
  61. ArticleHow SKA-Low steers a beam without moving its antennas
  62. ArticleHow EHT connects telescopes with clocks and hard drives
  63. ArticleHow Webb separates a faint spectrum from infrared background
  64. ArticleHow LHAASO finds an incoming direction from ground particles
Something broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy