Skip to main content
Sandbox Physics
How physics instruments measure

How EHT connects telescopes with clocks and hard drives

EHT records voltages at separate telescopes and later identifies samples that belong to the same incoming wave. Atomic frequency references keep the sampling rhythm stable. Time tags locate the samples, and correlation corrects their remaining misalignment. The links in this Earth-scale instrument are relationships between saved signals.

Sandbox Physics ·

Recording the wave before combining the observations

Each receiver converts incoming radiation into electrical signals suitable for digitisation. The recorded voltage samples contain astronomical emission mixed with receiving noise. EHT brings these separate digital recordings to correlators for comparison.

The separation between two telescopes is a baseline. An incoming wave usually travels different distances to its ends, and Earth’s rotation changes that difference. Station coordinates, Earth orientation and the source direction provide the first geometric-delay prediction.

Sources: [4][5]

A stable rhythm still needs a measured offset

A station hydrogen maser supplies the frequency reference for receiving and sampling. Its short-term stability lets the waves remain comparable over many samples. The clocks still have relative offsets and drifts, which must be monitored and corrected.

A timestamp locates a recording approximately. Atmospheric paths and electronic delays require further work. The atmosphere also changes phase, the position of the wave within its cycle. These changes limit how long signals can be averaged coherently, with the common wave reinforcing instead of cancelling.

Sources: [1][2]

Searching for the remaining delay

After applying the geometric model, processing searches over residual delay and its rate of change. Suitable corrections make the common astronomical signal survive multiplication and averaging. This is fringe fitting. Frequency channels and short time segments help distinguish an offset from a delay that is changing.

Noise can also produce a peak. One check is to find the peak in one set of samples and test its delay and fringe rate against independent samples. Calibration of receiving sensitivity, sampling and gains then relates the correlation coefficient to astronomical flux density.

Sources: [2][1]

A triangle cancels station phase terms

At the same time and frequency, an error assigned to one station enters one directed baseline with a positive sign and the next with a negative sign. Adding phases around three stations cancels those station terms. The combination is called closure phase.

Four stations give a second combination. Multiply the amplitudes of one pair of baselines and divide by the product for another pair. Each station occurs once in each product, cancelling its gain. This closure amplitude and the closure phase still carry noise and baseline-specific errors.

Φ123=ϕ12+ϕ23+ϕ31(mod2π)\Phi_{123}=\phi_{12}+\phi_{23}+\phi_{31}\pmod{2\pi}
The baselines run from 1 to 2, 2 to 3 and 3 to 1. ϕij\phi_{ij} is a directed baseline phase and Φ123\Phi_{123} the closure phase. Angles differing by a complete turn are equivalent.

Sources: [2][3]

Testing an image against the measurements

A finite station network leaves spatial information unmeasured; missing scans leave further gaps. Image reconstruction supplies explicit assumptions to fill them. Useful tests compare an image’s predicted signals with the recordings and examine which features survive changed assumptions or omitted stations.

Sources: [3]

Sources and further reading

  1. EHT Collaboration — First M87 Results II: Array and Instrumentation (2019)

    Read on 14 November 2026: selected frequency-standard passages in section 3.2, opening digital-backend/recording passages in 3.4–3.5, and correlation/calibration passages in section 5. Used for the 2017 system principles, not current station counts or performance records.

  2. EHT Collaboration — First M87 Results III: Data Processing and Calibration (2019)

    Read on 14 November 2026: section 3 and equations 1–2, section 5.1 on fringe processing and leave-one-band-out phase estimation, and section 7 equations 15–20 on closure quantities. No event data downloaded or measured thresholds adopted.

  3. EHT Collaboration — First M87 Results IV: Imaging the Central Supermassive Black Hole (2019)

    Read on 14 November 2026: opening section 2.2 and section 2.2.2 on incomplete sampling, extra image constraints and forward comparison with data. No reproduction of M87 imaging or complete-paper review is claimed.

  4. MIT Haystack Observatory — Event Horizon Telescope

    Read on 14 November 2026: hard-drive transport and correlation passages in The EHT Array, and downconversion/digital-backend/playback in Instrumentation. Current network, resolution and future-plan figures are not adopted.

  5. NRAO Essential Radio Astronomy — Chapter 3

    Read on 14 November 2026: sections 3.7.1–3.7.2 on geometric delay, correlation and complex visibility, to explain the comparison between separated stations.

Keep reading

What this page links to directly

See this series on the content map →

How physics instruments measure · Reading path

How physics instruments measure

  1. ArticleHow a vernier reads between millimetres
  2. ArticleAir helps hold up the object on a balance
  3. ArticleHow Cavendish Measured the Attraction Between Spheres
  4. ArticleWhich pressure does a liquid column measure?
  5. ArticleThe cup absorbs heat too
  6. ArticleRotational rheometers need both torque and speed
  7. ArticleWhy the air beyond a pipe changes its resonance
  8. ArticleHow a seismometer reads the ground through motion inside its case
  9. ArticleWhy does a meter change the circuit?
  10. ArticleHow a null detector reveals a resistance
  11. ArticleWhy a capacitor can behave like an inductor
  12. ArticleWhere does the slow wave on an oscilloscope come from?
  13. ArticleHow a lock-in amplifier finds a signal buried in noise
  14. ArticleWhere a small peak in a spectrum can come from
  15. ArticleHow a network analyzer separates the cable from the device
  16. ArticleWhen a slower quartz crystal can measure added mass
  17. ArticleHow an echelle spectrograph gives overlapping colors a second direction
  18. ArticleHow total brightness reveals many infrared colors
  19. ArticleMeasuring a Solution’s Concentration with Light
  20. ArticleHow an ellipsometer reads thin-film thickness from reflected light
  21. ArticleHow a polarimeter reveals mixed polarization
  22. ArticleHow individual photons reveal a fluorescence lifetime
  23. ArticleBrillouin Microscopy Listens to Sound with Light
  24. ArticleHow a white sphere compares a lamp’s total output
  25. ArticleHow powder rings reveal atomic spacing
  26. ArticleHow Small-Angle X-ray Scattering Reads a Particle Halo
  27. ArticleWhy X-ray fluorescence depends on sample thickness
  28. ArticleWhy an AFM makes a bump look wider
  29. ArticleHow a small indentation separates hardness from elasticity
  30. ArticleHow Differential Scanning Calorimetry Reads a Melting Peak
  31. ArticleHow thermogravimetry separates mass loss from the gas around it
  32. ArticleHow a diamond anvil cell reads pressure in ruby light
  33. ArticleHow diamond fluorescence reveals a magnetic-field direction
  34. ArticleReading a magnetic field in the rhythm of atoms
  35. ArticleKeeping a laser in time with atoms
  36. ArticleWhy a Kibble balance needs two experiments
  37. ArticleHow a dilution refrigerator keeps carrying heat away
  38. ArticleWhy a bigger vacuum pump may barely help the sample
  39. ArticleHow microwaves read a superconducting qubit
  40. ArticleHow current leaves a voltage across a thin sheet
  41. ArticleHow an ion flight becomes a mass spectrum
  42. ArticleHow four straight rods select ion mass
  43. ArticleHow a rapidly reversing field can hold ions
  44. ArticleHow a cyclotron keeps a proton on the accelerating beat
  45. ArticleWhy the drift tubes in an RF linac get longer
  46. ArticleThe directions hidden inside a small beam spot
  47. ArticleHow defocusing magnets help a beam stay in a ring
  48. ArticleWhy a collider cannot keep every collision
  49. ArticleGetting synchrotron light onto a sample
  50. ArticleHow a free-electron laser gathers electrons into a brighter beam
  51. ArticleFollowing energy exchange with neutron flight times
  52. ArticleHow a tokamak holds heat and measures where it goes
  53. ArticleWhy stellarator coils bend in three dimensions
  54. ArticleHow NIF turns a laser pulse into an implosion
  55. ArticleHow ELI brings an intense pulse to a target
  56. ArticleWhy a xenon detector needs liquid and gas
  57. ArticleHow LIGO measures a gravitational wave with mirrors that move
  58. ArticleHow delayed laser records let LISA measure gravitational waves
  59. ArticleHow FAST brings faint radio signals into focus
  60. ArticleHow ALMA keeps separated antennas in step
  61. ArticleHow SKA-Low steers a beam without moving its antennas
  62. You are hereHow EHT connects telescopes with clocks and hard drives
  63. ArticleHow Webb separates a faint spectrum from infrared background
  64. ArticleHow LHAASO finds an incoming direction from ground particles
1 more group

Other articles that use the same experiments

Each experiment from this article that another series also uses, followed by the articles there.

  1. LabStellar Interferometry and Sparse Imaging
  2. ArticleHow scattered telescopes make an image
  3. ArticleFrom wave delay to complex visibility
  4. ArticleEarth rotation and closure phase
  5. ArticleDirty beams and nonunique images
  6. ArticleFit a real EHT closure triangle
  7. ArticleRead EHT imaging and dynamic priors
  8. ArticleHow Ryle built a large radio telescope from smaller ones
Something broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy