How Differential Scanning Calorimetry Reads a Melting Peak

The area of a melting peak tells us how much extra heat a sample absorbed. The temperature at its top also remembers how quickly the sample was heated. A differential scanning calorimeter warms a sample and an empty reference pan in one furnace. The sample absorbs extra heat as it melts. The temperature difference temporarily grows into a peak.

Sandbox Physics ·

Two pans warm inside the same furnace

Even without a transition, the sample adds heat capacity to its pan. Raising that side through the same temperature interval requires more energy than warming an empty reference. Heat reaches both pans through finite conducting paths, so the sample side generally warms a little more slowly. During a steady ramp, the difference can settle to an approximately constant value.

Near melting, some incoming heat changes the sample’s state. An ideal pure substance can absorb heat at nearly constant temperature as it melts, while the reference continues warming. The sample temporarily falls further behind. Once melting finishes, its warming rate catches up again. The enlarged difference between the two temperature curves supplies a melting signal.

Sources: [1][3]

A calibrated thermal path converts difference into heat flow

Heat flow is energy transferred each second. A heat-flux instrument connects sample and reference to a common furnace through similar thermal paths and measures their temperature difference. Calibration converts that signal into a difference of heat flows. Here an endothermic process points upwards; another sign convention can draw the same melting process downwards.

A power-compensation instrument instead controls the two sides separately and compares their required electrical powers. Both architectures can measure transition heat, but the signal follows a different physical path.

Sources: [1]

Normal warming lies underneath the peak

The entire area between zero and the measured curve does not belong to melting. Ordinary warming requires heat, and the pans can store different amounts as their temperatures change. Instrument offsets can also add a blank signal. An empty-pan measurement removes the instrument background. The heat flow expected without a transition supplies the sensible-heat baseline. Removing both leaves the extra transition response.

Integrating power over elapsed time gives energy. Let Δh\Delta h be transition enthalpy per unit mass, mm the sample mass, Q˙\dot Q the calibrated heat-flow difference, Q˙base\dot Q_{\mathrm{base}} its baseline and tt time. At approximately constant pressure, a fully collected transition gives the relation below.

For example, a sample weighing ten milligrams that absorbs eighty joules per gram during a transition takes in an extra eight tenths of a joule. Doubling the mass and completing the same transition approximately doubles the heat. Enthalpy per unit mass stays the same. If the horizontal axis is temperature, integration must first convert temperature intervals to elapsed time using the actual heating history.

Δh=1m∫[Q˙(t)−Q˙base(t)]dt\Delta h=\frac{1}{m}\int\left[\dot Q(t)-\dot Q_{\mathrm{base}}(t)\right]dt
The baseline includes ordinary warming and the required instrument-storage correction. The integration endpoints must collect the transition signal, with an appropriate calibration for the finite thermal response.

Sources: [1][2][3]

A faster scan can move the peak

When the furnace warms faster, finite heat transfer lets the sample lag farther behind. Assigning the furnace or programmed temperature to the sample folds this delay into the apparent peak position. Recording the actual temperatures of both pans helps separate that thermal delay.

The material’s transformation can also take time. Melting and molecular rearrangement have finite rates. Even a curve plotted against actual sample temperature can shift or develop a longer tail during faster heating. The total heat of the completed transition can remain nearly unchanged while the timing changes. Peak temperature, a specified onset convention and equilibrium transition temperature describe different features.

Sources: [1][2]

Keep the tail and measure a reference

Ending a program as the peak rises collects only part of the transition heat. The sample may continue transforming during a final hold. Waiting for the response to return to a stable baseline helps collect the full area. More complex materials can also have temperature-dependent sensible heat, requiring a correspondingly different baseline.

Reference materials such as indium have known melting temperatures and fusion enthalpies that help check the temperature axis and heat-flow sensitivity. NIST’s 2024 certification of a new indium batch established its temperature and calorimetric values against traceable references. A useful measurement report also identifies heating rate, sample mass and baseline treatment, so its peak area and position can be interpreted under the conditions that produced them.

Sources: [2][1]

Sources and further reading

  1. NETZSCH, Functional principle of a heat-flux DSC

    Public furnace, sensor and twin-pan explanation; thermocouple temperature difference, heat-capacity baseline, melting response and the choice of peak sign. No proprietary sensor specifications adopted.

  2. Fortin, Herman, Koepke and Splett, NIST SP260-242, Certification of SRM2232a Indium (2024)

    Printed page1/Table1 for certified fusion values; section2.3.1 and figure2 for temperature/enthalpy calibration; printed page14/figure6 for baseline and time-area integration. These methods are distinct from the teaching model’s labelled peak and energy-quantile temperatures.

  3. OpenStax University Physics Volume2, section1.4: Heat Transfer, Specific Heat, and Calorimetry

    Heat as energy transfer and sensible heat. Used for the thermal picture; sensor geometry and calibration come from the separate instrument sources.

Keep reading

What this page links to directly

See this series on the content map →

How physics instruments measure · Reading path

How physics instruments measure

  1. ArticleHow a vernier reads between millimetres
  2. ArticleAir helps hold up the object on a balance
  3. ArticleHow Cavendish Measured the Attraction Between Spheres
  4. ArticleWhich pressure does a liquid column measure?
  5. ArticleThe cup absorbs heat too
  6. ArticleRotational rheometers need both torque and speed
  7. ArticleWhy the air beyond a pipe changes its resonance
  8. ArticleHow a seismometer reads the ground through motion inside its case
  9. ArticleWhy does a meter change the circuit?
  10. ArticleHow a null detector reveals a resistance
  11. ArticleWhy a capacitor can behave like an inductor
  12. ArticleWhere does the slow wave on an oscilloscope come from?
  13. ArticleHow a lock-in amplifier finds a signal buried in noise
  14. ArticleWhere a small peak in a spectrum can come from
  15. ArticleHow a network analyzer separates the cable from the device
  16. ArticleWhen a slower quartz crystal can measure added mass
  17. ArticleHow an echelle spectrograph gives overlapping colors a second direction
  18. ArticleHow total brightness reveals many infrared colors
  19. ArticleMeasuring a Solution’s Concentration with Light
  20. ArticleHow an ellipsometer reads thin-film thickness from reflected light
  21. ArticleHow a polarimeter reveals mixed polarization
  22. ArticleHow individual photons reveal a fluorescence lifetime
  23. ArticleBrillouin Microscopy Listens to Sound with Light
  24. ArticleHow a white sphere compares a lamp’s total output
  25. ArticleHow powder rings reveal atomic spacing
  26. ArticleHow Small-Angle X-ray Scattering Reads a Particle Halo
  27. ArticleWhy X-ray fluorescence depends on sample thickness
  28. ArticleWhy an AFM makes a bump look wider
  29. ArticleHow a small indentation separates hardness from elasticity
  30. You are hereHow Differential Scanning Calorimetry Reads a Melting Peak
  31. ArticleHow thermogravimetry separates mass loss from the gas around it
  32. ArticleHow a diamond anvil cell reads pressure in ruby light
  33. ArticleHow diamond fluorescence reveals a magnetic-field direction
  34. ArticleReading a magnetic field in the rhythm of atoms
  35. ArticleKeeping a laser in time with atoms
  36. ArticleWhy a Kibble balance needs two experiments
  37. ArticleHow a dilution refrigerator keeps carrying heat away
  38. ArticleWhy a bigger vacuum pump may barely help the sample
  39. ArticleHow microwaves read a superconducting qubit
  40. ArticleHow current leaves a voltage across a thin sheet
  41. ArticleHow an ion flight becomes a mass spectrum
  42. ArticleHow four straight rods select ion mass
  43. ArticleHow a rapidly reversing field can hold ions
  44. ArticleHow a cyclotron keeps a proton on the accelerating beat
  45. ArticleWhy the drift tubes in an RF linac get longer
  46. ArticleThe directions hidden inside a small beam spot
  47. ArticleHow defocusing magnets help a beam stay in a ring
  48. ArticleWhy a collider cannot keep every collision
  49. ArticleGetting synchrotron light onto a sample
  50. ArticleHow a free-electron laser gathers electrons into a brighter beam
  51. ArticleFollowing energy exchange with neutron flight times
  52. ArticleHow a tokamak holds heat and measures where it goes
  53. ArticleWhy stellarator coils bend in three dimensions
  54. ArticleHow NIF turns a laser pulse into an implosion
  55. ArticleHow ELI brings an intense pulse to a target
  56. ArticleWhy a xenon detector needs liquid and gas
  57. ArticleHow LIGO measures a gravitational wave with mirrors that move
  58. ArticleHow delayed laser records let LISA measure gravitational waves
  59. ArticleHow FAST brings faint radio signals into focus
  60. ArticleHow ALMA keeps separated antennas in step
  61. ArticleHow SKA-Low steers a beam without moving its antennas
  62. ArticleHow EHT connects telescopes with clocks and hard drives
  63. ArticleHow Webb separates a faint spectrum from infrared background
  64. ArticleHow LHAASO finds an incoming direction from ground particles
Something broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy