How physics instruments measure

How a polarimeter reveals mixed polarization

A Mueller polarimeter sends several polarization states into a sample and measures each output through several analyzers. These intensities distinguish a common rotation of every contribution from a mixture whose unresolved contributions reduce the overall polarization.

Sandbox Physics ·

Turn a polarizing filter and brightness becomes a question

Rotate polarizing sunglasses in front of an LCD screen. The screen changes brightness because its electric field favors one direction and the sunglasses transmit one direction more strongly. An ordinary photodetector measures total intensity. Add a rotating filter and it can also ask which field direction is favored.

The difference between horizontal and vertical intensities gives one linear-polarization contrast. Two diagonal settings give another. A wave plate changes the relative timing of perpendicular field components so an analyzer can also distinguish the two circular states. Total intensity and these three contrasts are the four Stokes parameters. Each coordinate comes from an intensity comparison.

Sources: [1]

A wave plate can change polarization at constant power

Send diagonally polarized light through an ideal quarter-wave plate. Its perpendicular field components begin in step. After the plate, one is a quarter-cycle behind the other, so their combined field rotates in a circle. The plate absorbs no light in this ideal example. Total intensity remains unchanged even though polarization changes.

Before the sample, a polarimeter uses polarizing elements to prepare different inputs. After the sample, analyzer elements turn different output polarizations into measurable intensities. Cycling both sets produces a table. A four-by-four Mueller matrix recovered from that table describes how the sample transforms the four input Stokes coordinates into the four output coordinates.

One input asks about one direction in this four-coordinate description. Linear states alone leave part of the circular response undetermined. Both preparation and analysis must span all four coordinates. Additional measurements provide repeated information that can be compared.

Sources: [2]

Small regions can preserve polarization while their sum loses it

Imagine two equal ideal wave-plate regions. One leaves polarization unchanged; the other turns horizontal polarization into vertical polarization. Horizontal illumination then produces equal horizontal and vertical contributions. If the regions have no stable relative phase, their intensities add. A detector that cannot resolve the regions receives a beam with no favored polarization. This averaging is called depolarization.

Now use a diagonal input that both regions send to the same diagonal output. Their sum remains fully polarized. The difference between the two experiments reveals a directional property of the sample. Horizontal illumination alone would suggest a more general scrambling than actually occurs.

A uniform wave plate gives a definite relation between two field components. An unresolved incoherent mixture needs the averaged intensity description supplied by a Mueller matrix. Spatial averaging of differently transformed light is enough to produce depolarization, even for a sample that stays still.

Sources: [2][3]

The instrument transforms polarization too

Polarizers leak some light along the rejected direction. Real wave plates miss the exact intended delay. Treating a commanded angle as an exact state would assign these instrument errors to the sample. Independently known reference light can characterize each analyzer response. An empty-path measurement then determines the states the generator actually sends.

Fresh air and known-wave-plate measurements test predictions beyond the fitted reference observations. Photon counts fluctuate, so small departures of a fitted matrix from physicality must be compared with count uncertainty. A resolved departure from passive-optics requirements calls for checking alignment, calibration or the model. The reference light and the unresolved regions together determine what the final measurement describes.

Sources: [1][3]

Sources and further reading

  1. NIST — Optical Polarization Metrology

    Four Stokes coordinates describe intensity and polarization; modulation of input and analyzer reveals sample structure.

  2. NIST SCATMECH — MuellerMatrix

    Mueller transformations, depolarization measures and the distinction between cone mapping and physically realizable Jones ensembles.

  3. San José and Gil — Characterization of passivity in Mueller matrices (2019 preprint)

    A physically realizable Mueller matrix must also satisfy both passivity inequalities to represent a mixture of passive components.

Keep reading

What this page links to directly

See this series on the content map →

How physics instruments measure · Reading path

How physics instruments measure

  1. ArticleHow a vernier reads between millimetres
  2. ArticleAir helps hold up the object on a balance
  3. ArticleHow Cavendish Measured the Attraction Between Spheres
  4. ArticleWhich pressure does a liquid column measure?
  5. ArticleThe cup absorbs heat too
  6. ArticleRotational rheometers need both torque and speed
  7. ArticleWhy the air beyond a pipe changes its resonance
  8. ArticleHow a seismometer reads the ground through motion inside its case
  9. ArticleWhy does a meter change the circuit?
  10. ArticleHow a null detector reveals a resistance
  11. ArticleWhy a capacitor can behave like an inductor
  12. ArticleWhere does the slow wave on an oscilloscope come from?
  13. ArticleHow a lock-in amplifier finds a signal buried in noise
  14. ArticleWhere a small peak in a spectrum can come from
  15. ArticleHow a network analyzer separates the cable from the device
  16. ArticleWhen a slower quartz crystal can measure added mass
  17. ArticleHow an echelle spectrograph gives overlapping colors a second direction
  18. ArticleHow total brightness reveals many infrared colors
  19. ArticleMeasuring a Solution’s Concentration with Light
  20. ArticleHow an ellipsometer reads thin-film thickness from reflected light
  21. You are hereHow a polarimeter reveals mixed polarization
  22. ArticleHow individual photons reveal a fluorescence lifetime
  23. ArticleBrillouin Microscopy Listens to Sound with Light
  24. ArticleHow a white sphere compares a lamp’s total output
  25. ArticleHow powder rings reveal atomic spacing
  26. ArticleHow Small-Angle X-ray Scattering Reads a Particle Halo
  27. ArticleWhy X-ray fluorescence depends on sample thickness
  28. ArticleWhy an AFM makes a bump look wider
  29. ArticleHow a small indentation separates hardness from elasticity
  30. ArticleHow Differential Scanning Calorimetry Reads a Melting Peak
  31. ArticleHow thermogravimetry separates mass loss from the gas around it
  32. ArticleHow a diamond anvil cell reads pressure in ruby light
  33. ArticleHow diamond fluorescence reveals a magnetic-field direction
  34. ArticleReading a magnetic field in the rhythm of atoms
  35. ArticleKeeping a laser in time with atoms
  36. ArticleWhy a Kibble balance needs two experiments
  37. ArticleHow a dilution refrigerator keeps carrying heat away
  38. ArticleWhy a bigger vacuum pump may barely help the sample
  39. ArticleHow microwaves read a superconducting qubit
  40. ArticleHow current leaves a voltage across a thin sheet
  41. ArticleHow an ion flight becomes a mass spectrum
  42. ArticleHow four straight rods select ion mass
  43. ArticleHow a rapidly reversing field can hold ions
  44. ArticleHow a cyclotron keeps a proton on the accelerating beat
  45. ArticleWhy the drift tubes in an RF linac get longer
  46. ArticleThe directions hidden inside a small beam spot
  47. ArticleHow defocusing magnets help a beam stay in a ring
  48. ArticleWhy a collider cannot keep every collision
  49. ArticleGetting synchrotron light onto a sample
  50. ArticleHow a free-electron laser gathers electrons into a brighter beam
  51. ArticleFollowing energy exchange with neutron flight times
  52. ArticleHow a tokamak holds heat and measures where it goes
  53. ArticleWhy stellarator coils bend in three dimensions
  54. ArticleHow NIF turns a laser pulse into an implosion
  55. ArticleHow ELI brings an intense pulse to a target
  56. ArticleWhy a xenon detector needs liquid and gas
  57. ArticleHow LIGO measures a gravitational wave with mirrors that move
  58. ArticleHow delayed laser records let LISA measure gravitational waves
  59. ArticleHow FAST brings faint radio signals into focus
  60. ArticleHow ALMA keeps separated antennas in step
  61. ArticleHow SKA-Low steers a beam without moving its antennas
  62. ArticleHow EHT connects telescopes with clocks and hard drives
  63. ArticleHow Webb separates a faint spectrum from infrared background
  64. ArticleHow LHAASO finds an incoming direction from ground particles