How physics instruments measure

How LIGO measures a gravitational wave with mirrors that move

LIGO compares the changing lengths of two long arms with suspended mirrors and laser light. The ground moves, and the control system moves the mirrors too. A strain measurement needs both a quiet mechanical reference and a calibrated account of the corrections made to it.

Sandbox Physics ·

One laser compares two paths

A beamsplitter sends laser light along perpendicular arms and brings the returning beams together. Light waves have crests and troughs. Their relative alignment, called phase, determines how they add. A small change in the difference between the paths changes the output power, which a photodetector converts to an electrical signal.

Each LIGO arm is about four kilometres long. A suitably oriented and polarized gravitational wave lengthens one arm as it shortens the other. The fractional change is called strain. Waves from other directions produce a different response: the instrument measures the projection that its two arms can sense.

Sources: [1][3]

A suspension separates the mirror from the moving ground

A mirror bolted to the floor follows the floor. A hanging mirror responds differently. When its support shakes quickly, inertia prevents it from following closely. Several suspension stages reduce the transmitted motion in succession. Well above the suspension resonances and below internal modes, longitudinal motion approaches that of a free mass.

Slow motion can still move the whole assembly, and motion near a resonance can grow. Sensors measure the supporting platform; actuators correct its motion, and damping controls the suspensions. Passive suspension and active control give the optics a quieter place to operate.

Sources: [2][3]

Stored light returns to a small reference field

Two mirrors form an optical cavity in each arm. Light makes repeated trips, increasing its response to an arm-length change. Storage also takes time. At higher frequencies, contributions from different trips no longer reinforce the signal equally, so storage affects bandwidth as well as sensitivity.

Most carrier light cancels at the measurement port. A small length offset lets some carrier through as a reference. Its interference with the light produced by the changing arm length creates a power variation at the photodetector. An output mode cleaner filters unwanted spatial patterns and modulation components before detection.

Sources: [3][4]

Feedback leaves information in two recorded signals

The cavities need to stay close to their operating point. A feedback controller reads a deviation and applies a corrective force to the mirrors. The error signal tracks the residual motion through the sensing system.

The control signal records the other part. Converting the error through the sensing response gives a residual displacement. Converting the control through the actuator and suspension response gives the motion used to compensate it. Adding these contributions and dividing by the arm length recovers an equivalent strain along the detector. Each conversion includes delay and frequency response.

h(f)=C−1(f)derr(f)+A(f)dctrl(f)Lh(f)=\frac{C^{-1}(f)d_{\mathrm{err}}(f)+A(f)d_{\mathrm{ctrl}}(f)}{L}
Both signals are analyzed at the same vibration frequency. The left side is equivalent strain; the denominator is one arm’s length. The numerator converts the saved error through the inverse sensing response to residual length, and the saved control through the actuation response to compensating motion. These add under the convention that feedback opposes the external length change.

Sources: [5][6]

A separate laser supplies a calibrated push

Electrical counts need a length reference. A photon calibrator reflects a separate, power-calibrated laser from a mirror. Reflection transfers momentum. Modulating the power applies a varying force; the measured reflected power, incidence angle and mechanical response let researchers estimate the resulting motion.

The power standard, optical losses and beam positions all contribute uncertainty. Off-centre illumination can also rotate the mirror. Excitations at different frequencies check the recovered amplitude, sign and delay, with further excitations used to test the calibration. This connects measured optical power to motion and ultimately to strain.

Sources: [7][6]

The calibrated record still contains noise

The mirrors and their coatings retain thermal motion. Detected photon numbers fluctuate too. More laser power can reduce relative counting fluctuations, but fluctuations in the light also push the mirrors. The limiting noise changes with frequency. A sensitivity comparison must refer every contribution to the same strain output.

Calibration delivers a noisy strain time series. Researchers then compare it with expected waveforms, examine environmental monitors and other observatories, and estimate properties of the source. Public strain products support that later analysis. The isolation, optical readout and calibration determine what the delivered curve means.

Sources: [8][3][9]

Sources and further reading

  1. LIGO Laboratory · What is an Interferometer?

    Read 14 November 2026: the basic Michelson paths, recombination and response to length changes. Uses the principle; no image or proton-size analogy is copied.

  2. LIGO Laboratory · Seismic Isolation

    Read 14 November 2026: the suspension and active-isolation passages on quadruple suspensions, main/reaction chains and active platforms. No dimensions, suppression factor or present sensitivity is adopted.

  3. LIGO Scientific Collaboration · Advanced LIGO (2015)

    Selected reading, 14 November 2026: arXiv:1411.4547v1 PDF pages 2–5 on configuration/readout, page 7 on quantum noise, selected passages on pages 17–18 and 20–21 on suspensions/isolation, and pages 33–34 on the output mode cleaner. Component roles only; design curves are not current measurements. Not a full-paper reading.

  4. Fricke et al. · DC readout experiment in Enhanced LIGO (2012)

    Read 14 November 2026: v2 PDF pages 2–4, section 2, on signal sidebands, a carrier-frequency reference and static offset; selected output-mode-cleaner passages. Used for the readout mechanism, without transferring Enhanced LIGO settings or measured noise to a present instrument.

  5. LIGO Laboratory · Feedback and Control Systems

    Read 14 November 2026: What is Feedback? and the response to environmental motion. Uses error comparison and corrective actuation, without adopting the page’s sample-rate number or literal perfect-stillness wording.

  6. Abbott et al. · Calibration of the Advanced LIGO detectors for GW150914 (2017)

    Read 14 November 2026: v2 sections I–II, equations (1)–(5), PDF pages 4–7 on feedback, sensing/actuation and strain reconstruction; also sections IV–V, equations (9)–(10), on photon calibration and actuator comparison. Keeps both reconstruction paths and delays; no event-specific uncertainty figure is adopted.

  7. Karki et al. · The Advanced LIGO Photon Calibrators (2016)

    Read 14 November 2026: section I equation (1), II.3, III.1–III.2 and V.4 on radiation pressure, beam positions, power-standard transfer, losses and delay. No precision record or figure is reproduced. An automatically generated HTML header date is not treated as the publication date.

  8. McCuller et al. · Frequency-Dependent Squeezing for Advanced LIGO (2020)

    Read 14 November 2026: v1 PDF page 1 introduction and the opening of section II on page 2, for readout fluctuations, radiation-pressure back action and their tradeoff. The article does not implement squeezing, repeat experimental values or infer the current upgrade status.

  9. Gravitational Wave Open Science Center · About and public data policy

    Read 14 November 2026: the public-data role and CC BY 4.0 data licence; also checked the tutorials introduction. Used only to distinguish strain products from later analysis. No event data was downloaded and no matched-filter result was reproduced.

Keep reading

What this page links to directly

See this series on the content map →

How physics instruments measure · Reading path

How physics instruments measure

  1. ArticleHow a vernier reads between millimetres
  2. ArticleAir helps hold up the object on a balance
  3. ArticleHow Cavendish Measured the Attraction Between Spheres
  4. ArticleWhich pressure does a liquid column measure?
  5. ArticleThe cup absorbs heat too
  6. ArticleRotational rheometers need both torque and speed
  7. ArticleWhy the air beyond a pipe changes its resonance
  8. ArticleHow a seismometer reads the ground through motion inside its case
  9. ArticleWhy does a meter change the circuit?
  10. ArticleHow a null detector reveals a resistance
  11. ArticleWhy a capacitor can behave like an inductor
  12. ArticleWhere does the slow wave on an oscilloscope come from?
  13. ArticleHow a lock-in amplifier finds a signal buried in noise
  14. ArticleWhere a small peak in a spectrum can come from
  15. ArticleHow a network analyzer separates the cable from the device
  16. ArticleWhen a slower quartz crystal can measure added mass
  17. ArticleHow an echelle spectrograph gives overlapping colors a second direction
  18. ArticleHow total brightness reveals many infrared colors
  19. ArticleMeasuring a Solution’s Concentration with Light
  20. ArticleHow an ellipsometer reads thin-film thickness from reflected light
  21. ArticleHow a polarimeter reveals mixed polarization
  22. ArticleHow individual photons reveal a fluorescence lifetime
  23. ArticleBrillouin Microscopy Listens to Sound with Light
  24. ArticleHow a white sphere compares a lamp’s total output
  25. ArticleHow powder rings reveal atomic spacing
  26. ArticleHow Small-Angle X-ray Scattering Reads a Particle Halo
  27. ArticleWhy X-ray fluorescence depends on sample thickness
  28. ArticleWhy an AFM makes a bump look wider
  29. ArticleHow a small indentation separates hardness from elasticity
  30. ArticleHow Differential Scanning Calorimetry Reads a Melting Peak
  31. ArticleHow thermogravimetry separates mass loss from the gas around it
  32. ArticleHow a diamond anvil cell reads pressure in ruby light
  33. ArticleHow diamond fluorescence reveals a magnetic-field direction
  34. ArticleReading a magnetic field in the rhythm of atoms
  35. ArticleKeeping a laser in time with atoms
  36. ArticleWhy a Kibble balance needs two experiments
  37. ArticleHow a dilution refrigerator keeps carrying heat away
  38. ArticleWhy a bigger vacuum pump may barely help the sample
  39. ArticleHow microwaves read a superconducting qubit
  40. ArticleHow current leaves a voltage across a thin sheet
  41. ArticleHow an ion flight becomes a mass spectrum
  42. ArticleHow four straight rods select ion mass
  43. ArticleHow a rapidly reversing field can hold ions
  44. ArticleHow a cyclotron keeps a proton on the accelerating beat
  45. ArticleWhy the drift tubes in an RF linac get longer
  46. ArticleThe directions hidden inside a small beam spot
  47. ArticleHow defocusing magnets help a beam stay in a ring
  48. ArticleWhy a collider cannot keep every collision
  49. ArticleGetting synchrotron light onto a sample
  50. ArticleHow a free-electron laser gathers electrons into a brighter beam
  51. ArticleFollowing energy exchange with neutron flight times
  52. ArticleHow a tokamak holds heat and measures where it goes
  53. ArticleWhy stellarator coils bend in three dimensions
  54. ArticleHow NIF turns a laser pulse into an implosion
  55. ArticleHow ELI brings an intense pulse to a target
  56. ArticleWhy a xenon detector needs liquid and gas
  57. You are hereHow LIGO measures a gravitational wave with mirrors that move
  58. ArticleHow delayed laser records let LISA measure gravitational waves
  59. ArticleHow FAST brings faint radio signals into focus
  60. ArticleHow ALMA keeps separated antennas in step
  61. ArticleHow SKA-Low steers a beam without moving its antennas
  62. ArticleHow EHT connects telescopes with clocks and hard drives
  63. ArticleHow Webb separates a faint spectrum from infrared background
  64. ArticleHow LHAASO finds an incoming direction from ground particles
1 more group

Other articles that use the same experiments

Each experiment from this article that another series also uses, followed by the articles there.

  1. LabLIGO Interferometer
  2. ArticleThe short rising signal that reached both LIGO detectors
Something broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy