Where does the slow wave on an oscilloscope come from?

A stable slow trace can come from a much faster input. An oscilloscope records voltage at clocked instants. The motion between those instants may leave no samples that distinguish it. Triggering can steady the picture, but it cannot supply the missing observations.

Sandbox Physics ·

Look once every millisecond

Start with ideal instantaneous sampling at one thousand samples per second. Give sinusoids at 100 Hz100\,\mathrm{Hz} and 1100 Hz1100\,\mathrm{Hz} the same starting phase. Between observations, the slow signal advances one tenth of a cycle. The fast one advances a whole cycle and another tenth. At every clock tick they arrive at the same phase.

The extra turn fits entirely between samples. Join the recorded points and the fast signal can look like the slow one. This merging of different frequencies in a sampled record is aliasing. The apparent wave can be clean and repeatable because the ambiguity repeats with the clock.

sin⁡ ⁣(2π1100n1000)=sin⁡ ⁣(2π100n1000)\sin\!\left(2\pi\frac{1100n}{1000}\right)=\sin\!\left(2\pi\frac{100n}{1000}\right)
The integer nn numbers the sample instants. This equality isolates sampling before analog filtering. A real front end also changes amplitude and phase.

Sources: [1][2]

A trigger gives the edge a repeatable position

Draw a repeating pulse from a different starting point each time and it slides across the display. A trigger selects a shared event, such as voltage rising through a chosen level. Putting that event at the same horizontal position makes the edge stand still.

One digital implementation finds the crossing among the numerical voltage samples already acquired. Sampling continues before the trigger, and a buffer preserves the preceding part of the stream. That is how a record can show events before its trigger point. If no qualifying edge arrives, normal mode retains the previous record; auto mode can save an untriggered segment after a timeout.

An aliased slow wave can repeatedly cross that level too. Steadiness shows that these samples align consistently. Distinguishing the fast input from the slow one requires information about sampling as well.

Sources: [1]

The edge can change before sampling begins

Voltage passes through the probe and analog input before it reaches the sampling clock. The probe adds a load to the source, and the input responds imperfectly to rapid changes. A sharp pulse can emerge lower and rounder. Bandwidth describes how that analog path responds to different frequencies.

A resistor and capacitor provide a simple teaching front end: voltage on the capacitor approaches a changed input gradually. With a bandwidth of 20 kHz20\,\mathrm{kHz}, an ideal step takes about 17.5 μs17.5\,\mathrm{\mu s} to rise from ten to ninety percent. That is less than one fifty-thousandth of a second. The number follows from this response shape and is not a universal oscilloscope specification.

A faster clock records the rounded edge in more detail. It does not restore the sharp edge that entered the analog path. Conversely, a fast analog input can pass changes that a slow clock still fails to distinguish.

Sources: [4][2]

Measurements must follow the recorded points

The converter stores each voltage as an integer code within a finite range. Observing zero and a known voltage supplies a voltage scale; counting clock ticks during a known interval checks the time scale. The connecting line is a display built from those discrete readings.

For example, dividing a full span of 4 V4\,\mathrm V into the two hundred and fifty-six codes available with eight bits gives steps of about 15.6 mV15.6\,\mathrm{mV}. Twelve bits provide four thousand and ninety-six codes, with steps of about 0.977 mV0.977\,\mathrm{mV}. Finer voltage steps reduce rounding error. Recording the motion between clock ticks still requires a sufficiently fast clock.

The record must also last long enough. A hundred-hertz wave fits less than half a cycle into a four-millisecond window. A forty-millisecond window spans roughly four cycles, allowing several same-direction crossings to be compared. A longer record supplies more repetitions while leaving the spacing between samples set by the clock.

Several crossings in the same direction can supply a period. Sampled ten- and ninety-percent crossings can supply an edge time. A short record or an edge represented by only a few sparse samples can leave those estimates poorly determined. If a peak exceeds the code range and clips, its original height is absent from the samples.

The earlier eleven-hundred-hertz input leaves more of its oscillations visible under a faster sampling clock. Keeping both sets of raw codes, their ranges and their trigger positions lets you investigate how the slow trace arose.

Sources: [3][2]

Sources and further reading

  1. Tektronix — Oscilloscope Systems and Controls

    Read Real-time Sampling, Figs27–29, and Trigger Position, Trigger Level and Slope, Trigger Modes, Fig35. The linked teaching instrument specifically uses a digital trigger operating on acquired codes.

  2. Tektronix — Evaluating Oscilloscopes

    Read Bandwidth, Fig44, Rise Time, Sample Rate and Record Length. The rise-time coefficient depends on response shape; the numerical example here is independently calculated for a declared single pole.

  3. Walt Kester, Analog Devices MT-001, Rev.A (10/08)

    Read pp1–3, Figs1–2 and Eqs1–4: ideal quantization error and the assumptions behind a noise approximation. No universal white-noise or effective-bit claim is made here.

  4. Tektronix — ABCs of Probes

    Read chapters2–3 on source loading, probe transmission and input resistance, and chapter5 on source impedance. The linked Lab chooses a simple buffered one-to-one probe with finite resistance.

Keep reading

What this page links to directly

See this series on the content map →

How physics instruments measure · Reading path

How physics instruments measure

  1. ArticleHow a vernier reads between millimetres
  2. ArticleAir helps hold up the object on a balance
  3. ArticleHow Cavendish Measured the Attraction Between Spheres
  4. ArticleWhich pressure does a liquid column measure?
  5. ArticleThe cup absorbs heat too
  6. ArticleRotational rheometers need both torque and speed
  7. ArticleWhy the air beyond a pipe changes its resonance
  8. ArticleHow a seismometer reads the ground through motion inside its case
  9. ArticleWhy does a meter change the circuit?
  10. ArticleHow a null detector reveals a resistance
  11. ArticleWhy a capacitor can behave like an inductor
  12. You are hereWhere does the slow wave on an oscilloscope come from?
  13. ArticleHow a lock-in amplifier finds a signal buried in noise
  14. ArticleWhere a small peak in a spectrum can come from
  15. ArticleHow a network analyzer separates the cable from the device
  16. ArticleWhen a slower quartz crystal can measure added mass
  17. ArticleHow an echelle spectrograph gives overlapping colors a second direction
  18. ArticleHow total brightness reveals many infrared colors
  19. ArticleMeasuring a Solution’s Concentration with Light
  20. ArticleHow an ellipsometer reads thin-film thickness from reflected light
  21. ArticleHow a polarimeter reveals mixed polarization
  22. ArticleHow individual photons reveal a fluorescence lifetime
  23. ArticleBrillouin Microscopy Listens to Sound with Light
  24. ArticleHow a white sphere compares a lamp’s total output
  25. ArticleHow powder rings reveal atomic spacing
  26. ArticleHow Small-Angle X-ray Scattering Reads a Particle Halo
  27. ArticleWhy X-ray fluorescence depends on sample thickness
  28. ArticleWhy an AFM makes a bump look wider
  29. ArticleHow a small indentation separates hardness from elasticity
  30. ArticleHow Differential Scanning Calorimetry Reads a Melting Peak
  31. ArticleHow thermogravimetry separates mass loss from the gas around it
  32. ArticleHow a diamond anvil cell reads pressure in ruby light
  33. ArticleHow diamond fluorescence reveals a magnetic-field direction
  34. ArticleReading a magnetic field in the rhythm of atoms
  35. ArticleKeeping a laser in time with atoms
  36. ArticleWhy a Kibble balance needs two experiments
  37. ArticleHow a dilution refrigerator keeps carrying heat away
  38. ArticleWhy a bigger vacuum pump may barely help the sample
  39. ArticleHow microwaves read a superconducting qubit
  40. ArticleHow current leaves a voltage across a thin sheet
  41. ArticleHow an ion flight becomes a mass spectrum
  42. ArticleHow four straight rods select ion mass
  43. ArticleHow a rapidly reversing field can hold ions
  44. ArticleHow a cyclotron keeps a proton on the accelerating beat
  45. ArticleWhy the drift tubes in an RF linac get longer
  46. ArticleThe directions hidden inside a small beam spot
  47. ArticleHow defocusing magnets help a beam stay in a ring
  48. ArticleWhy a collider cannot keep every collision
  49. ArticleGetting synchrotron light onto a sample
  50. ArticleHow a free-electron laser gathers electrons into a brighter beam
  51. ArticleFollowing energy exchange with neutron flight times
  52. ArticleHow a tokamak holds heat and measures where it goes
  53. ArticleWhy stellarator coils bend in three dimensions
  54. ArticleHow NIF turns a laser pulse into an implosion
Something broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy