Phase-only stretching
The Fourier transform uses a negative exponent, and the carrier uses the positive sign convention. With positive group-delay dispersion, frequencies above the carrier arrive later.
Can a short pulse be amplified without its peak power getting dangerously high? It can, if you stretch the pulse first. Dispersion spreads it out in time: same energy, much lower peak power. The amplifier can then add energy without the peak getting dangerously high, and a compressor afterwards undoes the stretch to give back a short, intense pulse.
Follow one complex pulse through four probe planes as it is stretched, amplified and compressed. Record its energy and peak power, then spoil the compression with leftover dispersion or a narrow spectral acceptance.
Why can the same pulse power give a different intensity at the target? The target receives energy where the beam comes together, and at the times the light arrives. An error in the wavefront can split the focal spot, even at the same power. An early pulse puts some of that energy on the target before the main peak.
Follow a compressed pulse through a square pupil. Change the wavefront or early pulse and compare the focal field with what a finite camera receives.
The phase is wrapped. Only bins above one ten-thousandth of the maximum spectral power are shown in the phase trace. The file retains every complex coefficient.
The file stores the computed field, camera words, pixel scale and teaching calibration separately. Select the current shot to save your latest settings.
Both images show the same field of view. The target image calculates the optical field. The synthetic camera integrates each pixel and digitizes its energy. Its reported range covers the receiving face and ADC rounding.
BackgroundChirped pulse amplification stretches a short pulse, amplifies it and compresses it again. Here one complex pulse goes through every step, and a probe can look at it at a physical plane after each one.
Why it mattersBecause every step acts on the same complex pulse and each one can be checked at a physical plane, you can see where the peak drops, where the energy grows and where the phase is removed.
The Fourier transform uses a negative exponent, and the carrier uses the positive sign convention. With positive group-delay dispersion, frequencies above the carrier arrive later.
The Lab checks the propagated field against this analytic intensity FWHM. Stretching leaves the energy and the spectral magnitude unchanged.
The Frantz–Nodvik map is applied to each time bin using the energy that has already passed. The energy taken from the initial reservoir can never exceed the saturation energy times the log gain.
This lossy step narrows the spectrum that gets through. It stands in for gain narrowing as a separate step in the chain and does not resolve inversion depletion by frequency. The rejected energy is reported.
Cancelling the second-order dispersion leaves any third-order or nonlinear phase in place, and it cannot bring back spectral energy that was already rejected.
Stretching lowers the peak intensity but not the incident fluence. The phase in the effective plate is a teaching diagnostic: it does not certify that an optic survives, and it is not integrated along the amplifier path.
Typical misconceptionIf the main peak is short, the pulse must be good.
Better mental modelAlso check the satellite pulses, the RMS duration, the energy near the edges of the numerical grid and the separate ideal reference.
Choose Matched CPA and move Physical probe plane through the four positions. All four fields are recorded automatically.
What to observe: The stretcher and the compressor each pass on all the energy they receive.Compare the Matched CPA and Bypass stretcher presets.
What to observe: The amplifier holds about the same stored energy in both cases, but the peak intensity and the plate phase are very different.Drag the compressor dial, then compare Dispersion mismatch, Cubic phase and Narrow acceptance. Check the short output pulse against its dashed ideal reference below.
What to observe: Cancelling one order of dispersion is not enough to restore the pulse fully.Read the instrument principle, sources and limits
BackgroundA short pulse emerges from the compressor. The square pupil passes the portion of the beam inside its opening. The phase across the pupil decides where that light interferes on the focal plane.
Why it mattersA camera records energy over a pixel and a time window. Its bright spot is useful even when the sequence of arrival times remains unknown.
The square pupil fixes a normalized focal pattern. The two time shifted pulse copies share this pattern because the model uses one carrier wavelength. Integrating intensity over the whole focal plane gives the power at that time.
Typical misconceptionThe highest camera code gives the peak intensity directly.
Better mental modelA pixel collects light over an area and the full simulated time window. The focal intensity shown here comes from the computed spatial field times the sampled instantaneous power.
Choose Matched pulse and then Split focus. Compare the computed focal plane and the camera pixels.
What to observe: The phase step makes light from the two halves cancel on axis. Energy reaches the off-axis lobes.Try Residual chirp, then Bright precursor.
What to observe: Residual chirp widens the main peak. The brighter early pulse changes the power before it while both pulses retain the same shape in space.Save the Narrow pupil shot, then switch to Matched pulse and save again. Download either record from the list.
What to observe: The file contains the original camera codes, their teaching calibration and the separate computed field.You are hereChirped Pulse Amplification
How physics instruments measure · Reading path
Articles on this path that use this experiment, followed by the other experiments they suggest.
Probes and measurement · Reading path
Articles on this path that use this experiment, followed by the other experiments they suggest.
Probes and measurement · Reading path
Articles on this path that use this experiment, followed by the other experiments they suggest.