Tunable Laser & Mode-Hop-Free Scan

Why does rotating a grating alone cause mode hops? The grating peak moves away from the current cavity resonance until a neighboring mode is favored. Coordinating cavity length with the grating moves the same integer order smoothly. It must stay inside the gain and etalon passbands; a fixed selector can still support a finite scan within its bandwidth.

Tune a Littrow grating, the external cavity length, the gain center and an intracavity etalon together. Scan the frequency out and back, and find where the laser hops to another mode or stops emitting.

Interactive modelTunable Laser & Mode-Hop-Free Scan
Actuator frequency travel—\text{—}
Grating target offset—\text{—}
Selected optical offset—\text{—}
Selected longitudinal order—\text{—}
Littrow angle—\text{—}
External cavity length change—\text{—}
Gain center offset—\text{—}
Relative output—\text{—}
Mode hops during emitted segments—\text{—}
Longest continuous same-order interval—\text{—}
Selected line minus grating target—\text{—}
Captured scan samples—\text{—}
Captured actuator travel—\text{—}

Physics tutorial

Keep one cavity order inside the transmission bands

BackgroundAn external-cavity diode laser picks its frequency with several elements at once: a Littrow grating, the external cavity, the gain center and an intracavity etalon. Scan the frequency out and back, and look for mode hops or places where the laser stops emitting.

Why it mattersRotating the grating looks like the obvious way to tune the laser, yet the frequency jumps in steps instead of moving smoothly.

Start with the essentials

Focus question
Why does rotating a grating alone cause mode hops?
One-sentence intuition
A mode-hop-free scan keeps one cavity order inside the grating, gain and etalon passbands throughout the sweep.

Core mathematical model

Littrow condition for the grating

λg=2dsin⁡θνg=c/λg\begin{aligned}\lambda_g&=2d\sin\theta\\\nu_g&=c/\lambda_g\end{aligned}

The grating spacing is fixed. Rotating its normal changes which wavelength returns along the incoming beam; the specular zero order goes toward a separate output collector.

Longitudinal modes of the external cavity

νq=qc2LΔνFSR=c2L\begin{aligned}\nu_q&=\frac{qc}{2L}\\\Delta\nu_{\rm FSR}&=\frac{c}{2L}\end{aligned}

Each mode's frequency is set by its integer order and the optical length of the cavity.

Shortening the cavity in step with the scan

L(x)=L0ν0ν0+rLxrL=1 ⇒ νq0=ν0+x\begin{aligned}L(x)&=L_0\frac{\nu_0}{\nu_0+r_Lx}\\r_L&=1\ \Rightarrow\ \nu_{q_0}=\nu_0+x\end{aligned}

A fixed integer order counts the same number of half-wavelengths between the endpoints. Raising its frequency shortens those half-wavelengths, so the cavity must shorten too.

Etalon transmission (Airy function)

Te=11+Asin⁡2(πu)u=ν−νeDeA=4Re(1−Re)2\begin{aligned}\mathcal T_e&=\frac{1}{1+A\sin^2(\pi u)}\\u&=\frac{\nu-\nu_e}{D_e}\\A&=\frac{4R_e}{(1-R_e)^2}\end{aligned}

If the etalon's narrow transmission band is not moved along, it can interrupt the scan.

A simplified score for each mode

Mq=3.6−12ug2−12ua2+ln⁡Teug=(νq−νg)/σgua=(νq−νa)/σa\begin{aligned}M_q&=3.6-\tfrac12 u_g^2-\tfrac12 u_a^2\\&\quad+\ln\mathcal T_e\\u_g&=(\nu_q-\nu_g)/\sigma_g\\u_a&=(\nu_q-\nu_a)/\sigma_a\end{aligned}

A mode can emit only when its score is positive. Saturation and absolute power are not modeled.

The laser keeps its current mode

Mbest−Mold≤hMold>0 ⇒ q=qold\begin{aligned}M_{\rm best}-M_{\rm old}&\le h\\M_{\rm old}&>0\ \Rightarrow\ q=q_{\rm old}\end{aligned}

Because of this empirical rule, the laser can switch modes at different points on the way out and on the way back.

Common difficulties

Tuning always means hopping

Typical misconceptionTo change the frequency, the laser has to jump to another longitudinal mode.

Better mental modelChanging the cavity length moves the same mode order smoothly to a new frequency.

A broader grating filter fixes everything

Typical misconceptionA broader grating selection guarantees a stable single-mode scan.

Better mental modelThis model always picks exactly one mode, so it cannot show whether a real laser stays single-mode. Coordinate the cavity length, gain and etalon instead.

Run the experiment

  1. 01

    Rotate only the grating

    The default already contains an outward and return sweep with Cavity tracking ratio at zero. Compare the steps in the frequency trajectory with the sloping target line.

    What to observe: The laser frequency jumps from one fixed cavity resonance to the next.
  2. 02

    Follow one mode order

    Predict whether the steps will disappear, then choose Coordinated cavity. The same sweep is reacquired automatically.

    What to observe: The same integer order now moves smoothly with the grating.
  3. 03

    Find the next limit

    Choose Fixed gain center or Stationary etalon, then choose Track every selector.

    What to observe: The output can drop out even when the cavity tracking is exact.

Related experiments and articles

Probes and measurement · Reading path

Controlled light and precision measurement

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleFinding sharp atomic lines inside a broad thermal spectrum
FeedbackSomething broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy