Tunable Laser & Mode-Hop-Free Scan

Coordinate a Littrow grating, external cavity, gain center and intracavity etalon. Acquire outward and return frequency scans and diagnose hops or loss of emission.

Interactive modelTunable Laser & Mode-Hop-Free Scan
Actuator frequency travel—\text{—}
Grating target offset—\text{—}
Selected optical offset—\text{—}
Selected longitudinal order—\text{—}
Littrow angle—\text{—}
External cavity length change—\text{—}
Gain center offset—\text{—}
Relative output—\text{—}
Mode hops during emitted segments—\text{—}
Longest continuous same-order interval—\text{—}
Selected line minus grating target—\text{—}
Captured scan samples—\text{—}
Captured actuator travel—\text{—}

Physics tutorial

A mode-hop-free scan moves one cavity resonance together with every frequency selector.

BackgroundCoordinate a Littrow grating, external cavity, gain center and intracavity etalon. Acquire outward and return frequency scans and diagnose hops or loss of emission.

Why it mattersWhy does rotating a grating alone cause mode hops?

Start with the essentials

Focus question
Why does rotating a grating alone cause mode hops?
One-sentence intuition
A mode-hop-free scan moves one cavity resonance together with every frequency selector.

Core mathematical model

Littrow selection

λg=2dsin⁡θνg=c/λg\begin{aligned}\lambda_g&=2d\sin\theta\\\nu_g&=c/\lambda_g\end{aligned}

The first diffracted order retraces the incident beam.

External longitudinal modes

νq=qc2LΔνFSR=c2L\begin{aligned}\nu_q&=\frac{qc}{2L}\\\Delta\nu_{\rm FSR}&=\frac{c}{2L}\end{aligned}

Integer mode order and optical length jointly determine the frequency.

Coordinated length

L(x)=L0ν0ν0+rLxrL=1 ⇒ νq0=ν0+x\begin{aligned}L(x)&=L_0\frac{\nu_0}{\nu_0+r_Lx}\\r_L&=1\ \Rightarrow\ \nu_{q_0}=\nu_0+x\end{aligned}

Shorten the cavity when the commanded frequency rises.

Airy etalon

Te=11+Asin⁡2[π(ν−νe)/De]A=4Re(1−Re)2\begin{aligned}\mathcal T_e&=\frac{1}{1+A\sin^2[\pi(\nu-\nu_e)/D_e]}\\A&=\frac{4R_e}{(1-R_e)^2}\end{aligned}

An untracked narrow transmission band can interrupt the scan.

Reduced selection merit

Mq=3.6−12ug2−12ua2+ln⁡Teug=(νq−νg)/σgua=(νq−νa)/σa\begin{aligned}M_q&=3.6-\tfrac12 u_g^2-\tfrac12 u_a^2\\&\quad+\ln\mathcal T_e\\u_g&=(\nu_q-\nu_g)/\sigma_g\\u_a&=(\nu_q-\nu_a)/\sigma_a\end{aligned}

Positive merit permits output; saturation and absolute power are not modeled.

Explicit memory rule

Mbest−Mold≤hMold>0 ⇒ q=qold\begin{aligned}M_{\rm best}-M_{\rm old}&\le h\\M_{\rm old}&>0\ \Rightarrow\ q=q_{\rm old}\end{aligned}

Forward and reverse switches can differ under this phenomenological rule.

Common difficulties

Tuning means hopping

Typical misconceptionChanging optical frequency must change longitudinal order.

Better mental modelChanging the cavity length moves a fixed order continuously.

A broad filter fixes everything

Typical misconceptionBroadening the grating guarantees a stable single-mode scan.

Better mental modelThis model selects one mode by construction; it cannot certify real multimode stability. Coordinate length, gain and etalon instead.

Run the experiment

  1. 01

    Rotate only the grating

    Acquire an outward and return sweep with zero cavity tracking.

    What to observe: Selected frequency steps between fixed external resonances.
  2. 02

    Track one order

    Choose coordinated cavity and repeat the same sweep.

    What to observe: The same integer order moves continuously with the grating.
  3. 03

    Find the next bottleneck

    Hold gain or etalon fixed, then restore tracking.

    What to observe: Output may disappear even when cavity tracking is exact.