Energy and momentum given to the sample
Positive energy transfer excites the sample. Wavevectors use radians per angstrom, so the wavelength is a full turn divided by the wavevector magnitude.
What does a neutron’s final energy tell us about motion inside a sample? A neutron that leaves with less energy has excited motion in the sample. Scanning its energy loss reveals excitation peaks, and the energy-gain partner shows how thermal motion populates those excitations.
Select the incident and outgoing neutron energies with two crystals. Scan energy transfer at fixed momentum transfer, record counts and fit an excitation with its thermal partner.
How can an arrival time reveal the energy a neutron exchanged with a sample? Choppers and independent standards first bound the incident speed. The scattered flight path and arrival time then bound the outgoing energy. Their difference gives the energy exchanged with the sample, and detector position also constrains momentum transfer.
Follow a neutron pulse through rotating slots and a sample. Detector pixels, arrival times and calibrated references bound the exchanged energy and momentum.
Schematic arm lengths and displacement density; transport slowed.
The sample cloud shows the stationary RMS displacement of the assumed thermal oscillator; shell shapes are schematic. Its RMS size scales with Bose occupation, with an arbitrary zero-point length. It is not an atomic trajectory. Count feedback uses the stored detector bin; all arm and crystal angles use the same solved geometry.
Capture a new scan or choose a demo.
Teal: counts. Purple: fit. Gold dashed: generating mean. Pink cursor: probe.
Purple: intrinsic profile. Gold: instrument-broadened response, not counts. Counting errors assume this known response and exclude calibration and model uncertainty.
Sample, empty container and electronics dark are separate exposures. Each compatible source origin and disconnected branch remains separate; a plotted region is not a measured point. Counting uncertainty and shared references retain their distinct roles. Negative corrected intervals remain.
Raw tags remain even when physical assignment fails. A union of source origins is retained as a union.
| Tag | Pixel | Raw clock code | Possible source | Energy transfer / meV | Momentum-transfer magnitude / inverse angstrom | Support |
|---|
The selected wavelength and emission time share slit constraints. Reference errors are common to many events and do not disappear by averaging more hits.
A 1.3 second observed beam-off history bounds earlier arrivals under the supplied finite chamber. A 120 millisecond observation still need not drain all allowed slow final flights.
Up to eight immutable measurements, with a cumulative decoded limit of 64 MiB. Raw events, independent references and units survive both formats; imports are checked before the session changes.
Open an empty sessionBackgroundSelect the incident and outgoing neutron energies with two crystals. Scan energy transfer at fixed momentum transfer, record counts and fit an excitation with its thermal partner.
Why it mattersDiffraction tells us about spatial order. Selecting the outgoing energy also reveals excitations that carry energy inside the sample.
Positive energy transfer excites the sample. Wavevectors use radians per angstrom, so the wavelength is a full turn divided by the wavevector magnitude.
The sample scattering angle is the full angle between the two beams. Each crystal uses its own Bragg half-angle to select energy.
The fitted amplitudes correct the final-to-incident wavevector factor. This relation uses integrated lines, not the two raw peak heights.
Typical misconceptionThe width of the measured peak is the sample’s intrinsic width.
Better mental modelThe instrument response also broadens the spectrum. The fit here uses a known Gaussian energy response and a fixed intrinsic width.
Move the energy-transfer probe and inspect the source and collector views.
What to observe: The monochromator changes incident energy; the analyzer keeps final energy fixed. The sample arm moves to keep momentum transfer fixed.Acquire and analyze Resolved excitation. Then compare Cool the sample.
What to observe: Cooling suppresses the energy-gain line because fewer excitations are already present.Choose Unreachable transfer and try to acquire.
What to observe: The scan stops when part of the chosen momentum and energy transfer cannot form a scattering triangle.A nonrelativistic neutron beam scatters from an illustrative thermal harmonic excitation. Final energy is fixed at 14.7 millielectronvolts. Every scan point solves the momentum triangle and the Bragg angles of a schematic 3.355-angstrom monochromator and analyzer. The sample spectrum contains an elastic Gaussian and one pair of excitation lines with a fixed intrinsic standard deviation of 0.18 millielectronvolts. Gaussian energy response, the outgoing-to-incoming wavevector factor, a flat background and Poisson counts are included. The fit uses those counts and known response widths. It searches excitation energies from one to 5.5 millielectronvolts. This is an energy-only response, not a full TAS resolution volume, a material identification or a reproduction of historical measurements. Magnetic scattering, absorption, multiple scattering, phonon eigenvectors and structure refinement are omitted. Hardware lengths are schematic and increase near backscattering to keep the two crystal mounts and open apertures apart. Particle motion is a slowed illustration; the spectra come from the recorded count model.
Read the instrument principle, sources and limits
BackgroundA neutron that gives energy to a sample leaves more slowly. At the same receiving surface, a later arrival can reveal this loss.
Why it mattersThe clock also includes the flight before the sample. Two rotating slots at separated positions select compatible combinations of incident speed and release time.
The right-handed axes follow the incident beam, horizontal transverse direction and vertical upward. Each post-sample flight begins with the actual new velocity.
Neutron mass and Planck’s constant connect speed with energy and wavelength. The actual trajectory calculation also retains gravity.
Positive energy transfer excites the sample. Both velocities refer to the sample; detector angle alone does not replace the vector relation.
Typical misconceptionThe most recent source trigger identifies every detector event.
Better mental modelA slow neutron or a neutron that lost substantial energy can arrive after a later pulse. Every allowed origin must be considered.
Typical misconceptionA narrower opening always makes every spectral feature clearer.
Better mental modelIt narrows the joint speed/release-time range but admits fewer neutrons. Finite pixels, geometry and reference uncertainty still contribute.
Typical misconceptionA negative corrected count must be a software error.
Better mental modelSample, container and dark observations each have finite counts. Their difference can be negative and retains its counting uncertainty.
Compare “Finite-sample energy exchange” with “Narrower slots”, then inspect the final chopper.
What to observe: Fewer particles pass. Transfer intervals change with the joint incident conditions selected by both slots.Choose “Source-pulse overlap” and expand the candidate table.
What to observe: One pixel and timestamp can retain several source origins, with separate energy and momentum regions.Change “Detector radius” and acquire again, then choose “Expired reference”.
What to observe: Radius changes the actual receiving position and flight path. Invalid references withhold dependent physical inference while recorded detector and monitor counts remain.Flight time reveals energy exchange through an independently anchored start, incident speed and receiving geometry. The finite range of compatible answers is part of the measurement.
You are hereNeutron Triple-Axis Spectrometer
How physics instruments measure · Reading path
Articles on this path that use this experiment, followed by the other experiments they suggest.
Probes and measurement · Reading path
Articles on this path that use this experiment, followed by the other experiments they suggest.
Fields & matter · Starter path
Map an electric field and steer a charged particle. Then find the electron's wavelength and use electron beams to see a crystal and a surface.
Start here →Make an Electric Field VisibleMove the probe around a dipole and check that the field always points downhill on the potential landscape.