Energy and momentum given to the sample
Positive energy transfer excites the sample. Wavevectors use radians per angstrom, so the wavelength is a full turn divided by the wavevector magnitude.
B-X4 · Measure energy exchange
What does a neutron’s final energy tell us about motion inside a sample? A neutron that leaves with less energy has excited motion in the sample. Scanning its energy loss reveals excitation peaks, and the energy-gain partner shows how thermal motion populates those excitations.
Select the incident and outgoing neutron energies with two crystals. Scan energy transfer at fixed momentum transfer, record counts and fit an excitation with its thermal partner.
Drag the gold analyzer handle or move the transfer below. Both energy selectors stay connected to the sample.
Neutron energies
Crystal and sample angles
Arm lengths are schematic; rotation angles follow the scattering geometry.
Arm directions and crystal planes follow the solved Bragg geometry. Hardware lengths and moving particles are schematic. The spectrum stays in the data row below.
Measurement sequence
One hundred sixty-one energy settings follow the same fixed momentum transfer. Each count keeps its own solved scattering geometry.
Ready. Choose a setup, then start the scan.
Start a scan to record counts. The spectrum will appear here as measurements arrive.
Recorded counts
No acquisition yet
Click to move the instrument probe. Teal dots: counts. Purple: fit. Gold dashed: separate generating mean.
Test the measurement model
Residuals use the same retained counts. The fit assumes a known Gaussian energy response.
Known instrument response
Purple: fixed intrinsic profile. Gold: instrument-broadened profile. These are known response shapes, not acquired counts.
The response width and the fixed intrinsic width combine before counts are generated. A broad, overlapping thermal pair can still give a visible peak while leaving temperature unresolved.
The thermal comparison uses integrated peak areas with the wavevector factor removed. No calibration or model uncertainty is included in the local counting errors.
Physics tutorial
BackgroundSelect the incident and outgoing neutron energies with two crystals. Scan energy transfer at fixed momentum transfer, record counts and fit an excitation with its thermal partner.
Why it mattersDiffraction tells us about spatial order. Selecting the outgoing energy also reveals excitations that carry energy inside the sample.
Start with the essentials
Positive energy transfer excites the sample. Wavevectors use radians per angstrom, so the wavelength is a full turn divided by the wavevector magnitude.
The sample scattering angle is the full angle between the two beams. Each crystal uses its own Bragg half-angle to select energy.
The fitted amplitudes correct the final-to-incident wavevector factor. This relation uses integrated lines, not the two raw peak heights.
Typical misconceptionThe width of the measured peak is the sample’s intrinsic width.
Better mental modelThe instrument response also broadens the spectrum. The fit here uses a known Gaussian energy response and a fixed intrinsic width.
Move the energy-transfer probe and inspect the source and collector views.
What to observe: The monochromator changes incident energy; the analyzer keeps final energy fixed. The sample arm moves to keep momentum transfer fixed.Acquire and analyze Resolved excitation. Then compare Cool the sample.
What to observe: Cooling suppresses the energy-gain line because fewer excitations are already present.Choose Unreachable transfer and try to acquire.
What to observe: The scan stops when part of the chosen momentum and energy transfer cannot form a scattering triangle.