B-X4 · Measure energy exchange

Neutron Triple-Axis Spectrometer

What does a neutron’s final energy tell us about motion inside a sample? A neutron that leaves with less energy has excited motion in the sample. Scanning its energy loss reveals excitation peaks, and the energy-gain partner shows how thermal motion populates those excitations.

Select the incident and outgoing neutron energies with two crystals. Scan energy transfer at fixed momentum transfer, record counts and fit an excitation with its thermal partner.

Interactive modelNeutron Triple-Axis Spectrometer

Drag the gold analyzer handle or move the transfer below. Both energy selectors stay connected to the sample.

Neutron energies

Ei=17.7 meVEf=14.7 meVQ=2.00 A˚−1\begin{aligned}E_i&=17.7\,\mathrm{meV}\\E_f&=14.7\,\mathrm{meV}\\Q&=2.00\,\mathring{\mathrm A}^{-1}\end{aligned}

Crystal and sample angles

α=41.6∘θM=18.7∘θA=20.6∘\begin{aligned}\alpha&=41.6^\circ\\\theta_M&=18.7^\circ\\\theta_A&=20.6^\circ\end{aligned}

Arm lengths are schematic; rotation angles follow the scattering geometry.

Apparatus scale and particle motion

Arm directions and crystal planes follow the solved Bragg geometry. Hardware lengths and moving particles are schematic. The spectrum stays in the data row below.

Measurement sequence

Scan the energy. Find the excitation.

One hundred sixty-one energy settings follow the same fixed momentum transfer. Each count keeps its own solved scattering geometry.

  1. 01Set the instrument
  2. 02Record the scan
  3. 03Infer from counts

Ready. Choose a setup, then start the scan.

Start a scan to record counts. The spectrum will appear here as measurements arrive.

Recorded counts

Neutron energy transfer

No acquisition yet

ε/meV,Nbin\varepsilon/\mathrm{meV},\quad N_{\rm bin}

Click to move the instrument probe. Teal dots: counts. Purple: fit. Gold dashed: separate generating mean.

Test the measurement model

What remains after fitting?

ε/meV,N−N^\varepsilon/\mathrm{meV},\quad N-\widehat N

Residuals use the same retained counts. The fit assumes a known Gaussian energy response.

Known instrument response

Peak width needs a resolution scale

δE/meV,R/Rmax⁡\delta E/\mathrm{meV},\quad R/R_{\max}

Purple: fixed intrinsic profile. Gold: instrument-broadened profile. These are known response shapes, not acquired counts.

σint=0.18 meV\sigma_{\rm int}=0.18\,\mathrm{meV}

The response width and the fixed intrinsic width combine before counts are generated. A broad, overlapping thermal pair can still give a visible peak while leaving temperature unresolved.

The thermal comparison uses integrated peak areas with the wavevector factor removed. No calibration or model uncertainty is included in the local counting errors.

Physics tutorial

Neutron Triple-Axis Spectrometer

BackgroundSelect the incident and outgoing neutron energies with two crystals. Scan energy transfer at fixed momentum transfer, record counts and fit an excitation with its thermal partner.

Why it mattersDiffraction tells us about spatial order. Selecting the outgoing energy also reveals excitations that carry energy inside the sample.

Start with the essentials

Focus question
What does a neutron’s final energy tell us about motion inside a sample?
One-sentence intuition
The energy the neutron loses is the energy the sample gains.

Core mathematical model

Energy and momentum given to the sample

Q=ki−kf,ε=Ei−Ef,E=ℏ2k22mn\mathbf Q=\mathbf k_i-\mathbf k_f,\qquad\varepsilon=E_i-E_f,\qquad E=\frac{\hbar^2k^2}{2m_n}

Positive energy transfer excites the sample. Wavevectors use radians per angstrom, so the wavelength is a full turn divided by the wavevector magnitude.

The scattering triangle must close

cos⁡α=ki2+kf2−Q22kikf,2dsin⁡θB=λ\cos\alpha=\frac{k_i^2+k_f^2-Q^2}{2k_i k_f},\qquad2d\sin\theta_B=\lambda

The sample scattering angle is the full angle between the two beams. Each crystal uses its own Bragg half-angle to select energy.

Thermal occupation sets the two integrated areas

AgainAloss=e−ε0/(kBT)\frac{A_{\rm gain}}{A_{\rm loss}}=e^{-\varepsilon_0/(k_B T)}

The fitted amplitudes correct the final-to-incident wavevector factor. This relation uses integrated lines, not the two raw peak heights.

Common difficulties

A reconstructed result depends on the measurement model

Typical misconceptionThe width of the measured peak is the sample’s intrinsic width.

Better mental modelThe instrument response also broadens the spectrum. The fit here uses a known Gaussian energy response and a fixed intrinsic width.

Run the experiment

  1. 01

    Follow both energy selectors

    Move the energy-transfer probe and inspect the source and collector views.

    What to observe: The monochromator changes incident energy; the analyzer keeps final energy fixed. The sample arm moves to keep momentum transfer fixed.
  2. 02

    Measure the thermal pair

    Acquire and analyze Resolved excitation. Then compare Cool the sample.

    What to observe: Cooling suppresses the energy-gain line because fewer excitations are already present.
  3. 03

    Test the geometry

    Choose Unreachable transfer and try to acquire.

    What to observe: The scan stops when part of the chosen momentum and energy transfer cannot form a scattering triangle.