What does an instrument record directly?

Trace a raw sensor reading through calibration and a blank comparison to the quantity it can support.

Before you start: Volume 1 Chapter 2 on calibration and Chapter 3 on residuals. The examples use Volume 3 Chapter 3 and Volume 5 Chapter 1.

After this chapter you can: Fit a calibration from reference readings and use a separate check and a blank to evaluate the result.

A detector records a signal, such as a pointer angle, a voltage or a count. Calibration connects the signal to a known reference. A background or blank measurement then helps separate what came from the sample from what came from the apparatus. You can check the result if the raw readings, settings and references are still available.

Volume 1, Chapters 2 and 3, used reference lengths and times to interpret camera positions. Volume 3, Chapter 3, showed how a meter changes a circuit. Here we trace a meter's raw signal to a voltage and use a calorimeter blank to isolate a sample contribution.

1.1 Record the signal along with the conditions that produced it

Choose the quantity you want to infer. For the meter that might be the potential difference between its leads. The direct signal in the simulated moving-coil meter is a coil angle in radians. A meter with a digital converter instead records an integer count. To interpret a row, keep the detector channel, time or trial number, unit, connection and range with it. These details tell you which conversion applies and whether the apparatus was in a usable state.

Record when each observation was taken. For equally spaced samples with interval Δt\Delta t, the times are tj=jΔtt_j=j\Delta t if the first sample starts at zero. The sampling rate is

fs=1Δt.f_s=\frac{1}{\Delta t}.

An observation can describe an instant or an interval. In the calorimeter the saved voltage and current belong to the preceding interval of constant power. Temperature samples represent the probe at the stated times. A temperature trace with one-second spacing can miss brief excursions between consecutive samples.

1.2 Calibration uses known inputs to map a detector signal

Suppose a simulated moving coil gives a raw angle rr that is nearly linear in voltage. At zero volts measure the mean raw angle r‾0\overline r_0. Apply a known span voltage VsV_s and measure the mean r‾s\overline r_s. A new reading gives the calibrated estimate

V^=Vsr−r‾0r‾s−r‾0.\widehat V=V_s\frac{r-\overline r_0}{\overline r_s-\overline r_0}.

The zero corrects an offset. The span sets the slope. A new reference between those two inputs checks how this conversion performs away from the points used to fit it. NIST/SEMATECH, Issues in Calibration.

Open Moving-Coil Meter and DC Multimeter and choose Reset. Open Input network, calibration and model. Keep the higher voltage range and the default voltage connections. Set Source resistance to zero and Source voltage to 4.00 V4.00\,\mathrm V. In this teaching circuit an ideal voltage source keeps the load at that voltage. Use Acquire zero and known span, let the needle settle and choose Keep this reading. Under Raw detector readings and zero/span calibration, compare the coil angle with the voltage it implies.

Figure 1.1 uses a separate reproducible run of the same model. It averages nine reference angles at each of 0 V0\,\mathrm V and 8.00 V8.00\,\mathrm V. The two raw means are about 0.004209 rad0.004209\,\mathrm{rad} and 0.841904 rad0.841904\,\mathrm{rad}. Their span gives a scale of about 9.5500 V rad−19.5500\,\mathrm{V\,rad^{-1}}. Five more samples use a simulated reference of 4.00 V4.00\,\mathrm V held out of the calibration.

  • Calibration reference
  • Held-out reading
An analog voltmeter calibration and five held-out measurementsThe upper panel shows the zero and eight-volt reference pairs and the calibration line. In the lower panel the five independent four-volt records have residuals of about minus two to plus two millivolts.

Figure 1.1 Two reference means set the angle-to-voltage calibration. The five residuals are calibrated voltages minus the separate 4 V check. The values come from a simulated meter.

Table 1.1 Five held-out readings of the simulated 4 V reference. The conversion uses one fixed calibration for all five.
RecordRaw angle (rad)Calibrated voltage (V)Residual (mV)
210.4231754.001134+1.134
220.4232044.001410+1.410
230.4232334.001686+1.686
240.4232624.001962+1.962
250.4228723.998238−1.762

1.3 Repeated readings and reference checks answer different questions

The mean of five held-out readings is about 4.00089 V4.00089\,\mathrm V. The sample standard deviation is about 1.51 mV1.51\,\mathrm{mV}. For nn independent readings of one stable quantity, let V‾\overline V be the mean and ss be the sample standard deviation. The repeatability contribution to the standard uncertainty of the mean is

s=∑i=1n(Vi−V‾)2n−1,urepeat=sn.s=\sqrt{\frac{\sum_{i=1}^{n}(V_i-\overline V)^2}{n-1}}, \qquad u_{\mathrm{repeat}}=\frac{s}{\sqrt n}.

The example gives about 0.676 mV0.676\,\mathrm{mV} for this contribution. It measures the scatter of an estimated mean under the repeatability assumptions. A shift in the zero, a reference error or a change in circuit loading can move every reading in the same direction. Repetition alone leaves such a shift in place. Use the held-out reference to look for a residual and keep separate information about the reference uncertainty and possible drift. NIST TN 1297, Type A Evaluation of Standard Uncertainty, NIST/SEMATECH, Analysis of Bias.

1.4 A matched blank measures what the apparatus contributes

Suppose you want the heat capacity of a metal sample. A warming curve also includes the water, cup and hardware. In Electrical Calorimeter, choose Reset and look at Saved warming and cooling curves. The model acquired a water-only blank with the same amount of water, the same vessel and the same heater pulse. Compare it with the sample run. The JSON export stores the original voltage, current and probe observations, plus calibration and the separate ideal model states.

The model fits the blank and sample records separately, accounting for heat loss and probe lag. Let C^sample\widehat C_{\mathrm{sample}} be the fitted bath heat capacity with the sample present and C^blank\widehat C_{\mathrm{blank}} the fitted blank bath heat capacity. Divide their difference by sample mass msm_s to estimate specific heat.

c^=C^sample−C^blankms.\widehat c=\frac{\widehat C_{\mathrm{sample}}-\widehat C_{\mathrm{blank}}}{m_s}.

The default fits are about 327.02 J K−1327.02\,\mathrm{J\,K^{-1}} and 281.95 J K−1281.95\,\mathrm{J\,K^{-1}}, respectively. A 0.0500 kg0.0500\,\mathrm{kg} sample gives about 901 J kg−1 K−1901\,\mathrm{J\,kg^{-1}\,K^{-1}}. Those fitted heat capacities come from the measured traces and the model. If you change the water mass between the blank and sample runs, the difference would also include the change in the heat capacity of the water. A matched blank lets the two fitted capacities share those apparatus contributions. OpenStax, University Physics Volume 2, Section 1.4.

1.5 Exercises

  1. A calibrated meter displays a voltage inferred from a movable coil. What is the direct signal? Name two other record fields needed to convert it.
  2. A sensor reads 0.020 rad0.020\,\mathrm{rad} at zero volts and 0.820 rad0.820\,\mathrm{rad} at 8.0 V8.0\,\mathrm V. Find the calibrated voltage for a raw reading of 0.420 rad0.420\,\mathrm{rad}.
  3. Two calibrations are possible for a meter, one for a lower range and one for a higher range. Why should a new raw record say which range was used?
  4. A thermometer takes one sample every 0.20 s0.20\,\mathrm s. Give the sampling rate. Can it reconstruct every change lasting much less than that interval?
  5. Compare the averages of one thousand and ten readings made with the same instrument bias. What happens to random scatter of the mean? What happens to the bias?
  6. A sample of mass 0.060 kg0.060\,\mathrm{kg} raises a calorimeter's fitted capacity from 250250 to 280 J K−1280\,\mathrm{J\,K^{-1}}. Find its specific heat, assuming matched conditions.
  7. In Table 1.1, which reading has the largest absolute residual? What extra information would help decide whether the calibration is biased near 4 V?
  8. A blank used less water than the loaded run. If you subtract the blank heat capacity to estimate the sample capacity, what bias in the sample estimate do you expect?

1.6 Answers and explanations

Exercise 1

The direct reading is the coil angle. The voltage range and calibration record are needed to convert it. The connections also identify the two nodes whose voltage difference was measured.

Exercise 2

The unknown angle lies halfway from zero to span. The calibration gives 8.0(0.420−0.020)/(0.820−0.020) V=4.0 V8.0(0.420-0.020)/(0.820-0.020)\,\mathrm V=4.0\,\mathrm V.

Exercise 3

The selected range changes how much current passes through the coil for a given voltage. A fixed angle can therefore correspond to different voltages. The raw record needs the matching calibration.

Exercise 4

The sampling rate is 1/(0.20 s)=5.0 Hz1/(0.20\,\mathrm s)=5.0\,\mathrm{Hz}. A brief change can fall between readings or leave too few points to show its shape.

Exercise 5

If the samples are independent with the same spread, the larger mean has one tenth the random standard error because its sample size is one hundred times larger. Both averages retain the same bias.

Exercise 6

The sample adds heat capacity 30 J K−130\,\mathrm{J\,K^{-1}}. Dividing by mass gives 500 J kg−1 K−1500\,\mathrm{J\,kg^{-1}\,K^{-1}}.

Exercise 7

Record 24 has the largest absolute residual, about 1.962 mV1.962\,\mathrm{mV}. Repeat measurements of the same independent reference and information about its uncertainty and the meter's drift would help assess bias.

Exercise 8

The estimated sample heat capacity is too large. Some of the added heat capacity belongs to the extra water in the loaded run. The subtraction attributes all of it to the sample.

Test it in the Lab

Change one setting at a time, then compare what you see with the tables and answers in this chapter.

Moving-Coil Meter and DC Multimeter →Connect physical probes, compare moving-coil and digital meters, and retain calibrated DC readings to measure loading and current burden.Electrical Calorimeter →Heat a sample in a stirred water bath. Record electrical input and thermometer response, compare a separate water-only blank, and recover specific heat with heat-loss and probe-response corrections.

Sources and reading scope

  1. NIST/SEMATECH Engineering Statistics Handbook, §2.3.1 Issues in Calibration

    The handbook explains how calibration relates an instrument reading to reference standards and why checks cover the range of use.

  2. NIST/SEMATECH Engineering Statistics Handbook, §2.4.5 Analysis of Bias

    The handbook compares measured values with a reference and discusses how to assess different sources of measurement bias.

  3. NIST Technical Note 1297, §3 Type A Evaluation of Standard Uncertainty

    The note evaluates a mean from repeated observations and gives the standard uncertainty for independent repeated readings.

  4. OpenStax, University Physics Volume 2, §1.4 Heat Transfer, Specific Heat, and Calorimetry

    Heat capacity connects an energy transfer to a temperature change, and calorimetry uses the energy balance to infer the final temperature.

Keep reading

What this page links to directly

See this series on the content map →

Interactive Physics Textbook · Reading path

Learning the Microscopic World from Experiments

  1. You are hereWhat does an instrument record directly?
  2. ArticleHow does diffraction reveal crystal plane spacing?
  3. ArticleHow do two-dimensional images reveal a particle’s three-dimensional shape?
  4. ArticleHow does an X-ray spectrum identify elements?
  5. ArticleHow do counts reveal neutrino oscillations?
  6. ArticleHow does the end of a beta spectrum constrain neutrino mass?
1 more group

Other articles that use the same experiments

Each experiment from this article that another series also uses, followed by the articles there.

  1. LabMoving-Coil Meter and DC Multimeter
  2. ArticleWhy does a meter change the circuit?
  3. LabElectrical Calorimeter
  4. ArticleThe cup absorbs heat too
Something broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy