When does a thermometer represent the sample temperature?

Check thermal equilibrium, calibration and probe response before treating a thermometer reading as the sample temperature.

Before you start: Volume 1 Chapter 2 and Chapter 3 on measurement, calibration and model checks. Basic algebra.

After this chapter you can: Use a thermometer scale, calculate an equilibrium temperature and explain the lag between a probe and its sample.

A thermometer represents a sample when it is at the same temperature as the part of the sample it touches. Give the probe time to exchange heat, and check that the probe scarcely changes the sample's temperature. The reading also needs a calibration that converts the sensor's signal to a temperature.

Volume 1, Chapters 2 and 3, compared instrument readings with reference values and tracked changes in time. Here we apply the same ideas to a thermometer in a bath.

1.1 Bodies in thermal equilibrium have the same temperature

Put a cooler thermometer into a warmer bath. Energy passes into the thermometer as heat. The temperature difference shrinks as the thermometer warms. When the two are in thermal equilibrium, there is no net heat transfer between them.

The zeroth law says that if two bodies are each in thermal equilibrium with a third body, they are in thermal equilibrium with each other. A thermometer can serve as that third body. Two baths that bring it to the same equilibrium reading have the same temperature at the points of contact. For this comparison, the thermometer's surroundings and contact must let it settle to the bath temperature. A probe attached to a cold support may settle to a steady reading even while heat continues to cross it. OpenStax, University Physics Volume 2, Section 1.1.

Heat can travel through a contact by conduction. Moving fluid can transport energy by convection. Radiation carries energy across space, even without a material connection. Stirring a bath can reduce temperature differences within it. A probe at one position measures a local temperature, so check more than one position if the bath is uneven. OpenStax, University Physics Volume 2, Section 1.6.

1.2 Calibration turns a sensor reading into degrees

A thermometer may report a voltage, a resistance or a raw number that changes with temperature. In a narrow working range, suppose its raw reading rr depends linearly on the numerical Celsius value TCT_{\mathrm C}. Then r=aTC+br=aT_{\mathrm C}+b, where the gain aa sets the change per degree and the intercept bb sets the reading at zero Celsius. Two reference temperatures determine both. For a new reading,

TC=r−ba.T_{\mathrm C}=\frac{r-b}{a}.

The kelvin value is TK=TC+273.15T_{\mathrm K}=T_{\mathrm C}+273.15. A change of one Celsius degree has the same size as a change of one kelvin. The two scales put zero at different points, so use the absolute kelvin temperature when a law calls for a ratio of absolute temperatures. OpenStax, University Physics Volume 2, Section 1.2.

The simulated sensor in Section 1.4 reads 20.2420.24 at a reference of 20 ∘C20\,^\circ\mathrm C and 40.3040.30 at 40 ∘C40\,^\circ\mathrm C. The change in raw reading divided by the reference change gives a=1.003a=1.003. The first reference gives b=0.18b=0.18. The calibration corrects scale and zero. A reading can still lag while the probe warms.

1.3 Heat capacity sets how much a sample changes during contact

A body with heat capacity CC needs energy Q=CΔTQ=C\Delta T to change its temperature by ΔT\Delta T, when its heat capacity is nearly constant, the body stays in the same phase and mechanical work is negligible. Heat capacity is measured in joules per kelvin. Specific heat cc is the heat capacity per unit mass, so a body of mass mm has C=mcC=mc. OpenStax, University Physics Volume 2, Section 1.4.

Bring two bodies starting at T1T_1 and T2T_2 into contact and insulate them from the room. With nearly constant heat capacities C1C_1 and C2C_2, one body's energy loss is the other's gain. Their common final temperature TfT_f satisfies

C1(Tf−T1)+C2(Tf−T2)=0.C_1(T_f-T_1)+C_2(T_f-T_2)=0.

Solving for the final temperature weights each starting temperature by its heat capacity.

Tf=C1T1+C2T2C1+C2.T_f=\frac{C_1T_1+C_2T_2}{C_1+C_2}.

For example, mix 0.100 kg0.100\,\mathrm{kg} of water at 60.0 ∘C60.0\,^\circ\mathrm C with 0.200 kg0.200\,\mathrm{kg} at 20.0 ∘C20.0\,^\circ\mathrm C. With a cup of negligible heat capacity, the two amounts have the same specific heat. Their final temperature is about 33.3 ∘C33.3\,^\circ\mathrm C. It lies closer to the colder starting temperature because there is twice as much cold water.

The same balance applies to a sample and its probe. If the probe has much less heat capacity than the sample, their final temperature stays near the sample's initial temperature. For a tiny sample, the probe can shift the final temperature noticeably.

1.4 A probe follows a changing bath with a delay

Hold a bath at temperature TbT_b. Suppose a probe at TpT_p absorbs heat at a rate G(Tb−Tp)G(T_b-T_p), with thermal conductance GG in watts per kelvin. The probe has heat capacity CpC_p, so its temperature follows

CpdTpdt=G(Tb−Tp),Tp(t)=Tb+[Tp(0)−Tb]e−t/τ,τ=CpG.\begin{aligned} C_p\frac{\mathrm dT_p}{\mathrm dt} &= G(T_b-T_p),\\ T_p(t)&=T_b+[T_p(0)-T_b]\mathrm e^{-t/\tau},\qquad \tau=\frac{C_p}{G}. \end{aligned}

The temperature gap loses the same fraction in each equal time interval. After time τ\tau, the remaining gap is about thirty-seven percent of the initial gap. For a probe that starts at 22 ∘C22\,^\circ\mathrm C and enters a bath held at 32 ∘C32\,^\circ\mathrm C, it reads about 28.32 ∘C28.32\,^\circ\mathrm C after one response time. The curve assumes constant conductance and a fixed bath temperature.

  • Maintained bath
  • Four-second probe
  • Twenty-four-second probe
Fast and slow thermometers after entering a maintained bathThe bath stays at 32 degrees Celsius. The two thermometers begin at 22 degrees. The four-second probe approaches the bath temperature before the twenty-four-second probe.

Figure 1.1 Calculated thermometer response in a maintained 32 °C bath. Both probes start at 22 °C. Their response times are four and twenty-four seconds.

Open Electrical Calorimeter and choose Reset. Under Raw measurements, probe calibration and model, find Probe response in a fixed-temperature bath. The default record gives a fitted response time of about 4.00 s4.00\,\mathrm s. Choose A slow thermometer to acquire a second record and compare the step curve. Its fitted response time is about 24.0 s24.0\,\mathrm s. The bath inside this separate calibration stays at the same target temperature.

Table 1.1 Calibrated probe temperatures from the simulated fixed-bath step records. The bath stays at 32 °C. The probe samples include a small reproducible perturbation.
Time (s)Default probe (°C)Slow probe (°C)
428.32123.535
2431.97628.322
12031.99931.932

Return to the main warming curve. The heater now brings the bath temperature up gradually. In the model, the sample, water and container share a single bath temperature. The probe has its own temperature. After 120 s120\,\mathrm s the ideal bath is about 24.57 ∘C24.57\,^\circ\mathrm C for either probe. The calibrated detector readings are about 24.49 ∘C24.49\,^\circ\mathrm C and 24.10 ∘C24.10\,^\circ\mathrm C. The lab's JSON export keeps the ideal temperatures and the separate raw measurements. During heating, a calibrated thermometer with a longer response time can read farther below the sample temperature.

1.5 Exercises

  1. A thermometer settles to the same reading in two large, well-stirred baths. State the law that relates the bath temperatures and what contact condition is needed.
  2. Convert 25.0 ∘C25.0\,^\circ\mathrm C to kelvins. If the temperature rises by 5.05.0 Celsius degrees, what is the rise in kelvins?
  3. A sensor gives raw numbers of 15.515.5 and 35.735.7 at reference temperatures of 15.0 ∘C15.0\,^\circ\mathrm C and 35.0 ∘C35.0\,^\circ\mathrm C. Find the gain and intercept. What Celsius temperature corresponds to a raw value of 25.625.6?
  4. An object of heat capacity 240 J K−1240\,\mathrm{J\,K^{-1}} absorbs 1200 J1200\,\mathrm J of heat without doing appreciable work. Find its temperature change.
  5. Mix equal masses of water at 20.0 ∘C20.0\,^\circ\mathrm C and 50.0 ∘C50.0\,^\circ\mathrm C in an insulated container of negligible heat capacity. Find the final temperature and explain why the arithmetic average applies here.
  6. Two probes start at 22 ∘C22\,^\circ\mathrm C and move to a bath held at 32 ∘C32\,^\circ\mathrm C. Their response times are four and twenty-four seconds. Which probe is closer to the bath after four seconds? Use Table 1.1 to compare.
  7. A small sample feels heat from both a probe and the room. The probe reading is steady. What else must you check before treating it as the sample temperature?
  8. A probe trails below a warming sample's temperature, then the heater turns off and the bath cools. Can the probe later read higher than the bath? Explain using the temperature gap.

1.6 Answers and explanations

Exercise 1

The zeroth law relates the two baths by thermal equilibrium with the same thermometer. The probe must come to the bath temperature at each contact with negligible influence from its surroundings.

Exercise 2

The absolute temperature is 298.15 K298.15\,\mathrm K, or 298.2 K298.2\,\mathrm K to the precision of the Celsius reading. The increase is 5.0 K5.0\,\mathrm K because the offset cancels in a difference.

Exercise 3

The gain is the raw change divided by the reference change, so a=20.2/20.0=1.01a=20.2/20.0=1.01. Using the first reference gives b=0.35b=0.35. The calibrated value is (25.6−0.35)/1.01=25.0 ∘C(25.6-0.35)/1.01=25.0\,^\circ\mathrm C.

Exercise 4

The increase is ΔT=Q/C=5.0 K\Delta T=Q/C=5.0\,\mathrm K. A positive heat input raises the temperature when the body remains in the same phase and work can be neglected.

Exercise 5

The final temperature is 35.0 ∘C35.0\,^\circ\mathrm C. Equal masses of water have equal heat capacities here. Losing and gaining the same energy therefore require the same magnitude of temperature change.

Exercise 6

The default probe is closer. Its gap after four seconds is about 3.68 K3.68\,\mathrm K. The slow probe is still about 8.47 K8.47\,\mathrm K below the bath because it has completed only one sixth of its response time.

Exercise 7

Check that the probe is in good thermal contact with the sample and that little heat flows from it to the room or its mount. Check that the probe did not appreciably change the sample temperature. A steady reading can arise when ongoing heat flows balance at the probe.

Exercise 8

Yes. While the bath cools, its temperature can drop below the slower probe. The gap then points from the probe to the bath. The probe gives heat to the bath as it approaches the new temperature.

Test it in the Lab

Change one setting at a time, then compare what you see with the tables and answers in this chapter.

Electrical Calorimeter →Heat a sample in a stirred water bath. Record electrical input and thermometer response, compare a separate water-only blank, and recover specific heat with heat-loss and probe-response corrections.

Sources and reading scope

  1. OpenStax, University Physics Volume 2, §1.1 Temperature and Thermal Equilibrium

    The section uses the zeroth law to explain how two bodies in thermal equilibrium with a thermometer share a temperature.

  2. OpenStax, University Physics Volume 2, §1.2 Thermometers and Temperature Scales

    The section explains how thermometers use a measurable property and compares Celsius and Kelvin scales.

  3. OpenStax, University Physics Volume 2, §1.4 Heat Transfer, Specific Heat, and Calorimetry

    Heat capacity connects an energy transfer to a temperature change, and calorimetry uses the energy balance to infer the final temperature.

  4. OpenStax, University Physics Volume 2, §1.6 Mechanisms of Heat Transfer

    The section distinguishes heat transfer by conduction, convection and radiation and gives examples of what sets the transfer rate.

Keep reading

What this page links to directly

You are hereWhen does a thermometer represent the sample temperature?

Articles

See this series on the content map →

Interactive Physics Textbook · Reading path

Heat, Probability and Matter

  1. You are hereWhen does a thermometer represent the sample temperature?
  2. ArticleHow do molecular collisions create pressure?
  3. ArticleHow do heat and work change internal energy?
  4. ArticleWhy does heat flow spontaneously from hot to cold?
  5. ArticleHow much work can a heat engine deliver in one cycle?
  6. ArticleHow do probabilities determine a system's thermal averages?
1 more group

Other articles that use the same experiments

Each experiment from this article that another series also uses, followed by the articles there.

  1. LabElectrical Calorimeter
  2. ArticleThe cup absorbs heat too