How does diffraction reveal crystal plane spacing?

Use diffraction angles and ring radii to calculate crystal plane spacings. Check the angular scale against a silicon powder reference.

Before you start: Volume 4 Chapter 2 on imaging, Chapter 3 on interference and Chapter 4 on diffraction. Volume 6 Chapter 2 connects wavelength to momentum. Chapter 1 covers calibration.

After this chapter you can: Calculate a crystal spacing from an electron ring or X-ray peak and identify which calibration sets the result.

A diffraction peak gives a spacing once the wavelength and scattering angle are known. An area detector records counts at pixel positions. The analysis converts the positions to angles, locates peaks and uses Bragg's law. Calibration fixes the angular scale before the sample is interpreted.

2.1 The peak comes from waves adding in phase

Consider two parallel crystal planes separated by a perpendicular distance dd. For elastic scattering, the incident and outgoing waves have the same wavelength λ\lambda. In the Bragg geometry both rays make an angle θ\theta with the planes. The ray scattered from the deeper plane travels an extra dsin⁡θd\sin\theta on the way in and again on the way out. Contributions from successive planes add in phase when

2dsin⁡θ=nλ,n=1,2,…2d\sin\theta=n\lambda,\qquad n=1,2,\ldots

The scattering angle α\alpha between the incident and outgoing beam directions is 2θ2\theta. For a first-order indexed peak,

d=λ2sin⁡(α/2).d=\frac{\lambda}{2\sin(\alpha/2)}.

The magnitude of momentum transfer divided by ℏ\hbar is q=4πsin⁡(α/2)/λq=4\pi\sin(\alpha/2)/\lambda. At a first-order peak, q=2π/dq=2\pi/d. This reciprocal relation explains why a smaller spacing sends the same wavelength to a larger scattering angle.

2.2 Counts become angles only after calibration

A powder contains crystallites in many orientations. Some grains satisfy the Bragg condition at each allowed angle. Their outgoing directions lie on a cone about the incident beam, so a detector perpendicular to the beam can cut out a ring. With sample-to-detector distance LL and ring radius rr measured from the direct-beam position, the geometry is

α=arctan⁡(r/L).\alpha=\arctan(r/L).

At small angles the ring also gives a direct approximation for the spacing.

d≃λLr(r/L≪1).d\simeq\frac{\lambda L}{r}\quad(r/L\ll 1).

The last approximation follows from tan⁡α≃α\tan\alpha\simeq\alpha and 2sin⁡(α/2)≃α2\sin(\alpha/2)\simeq\alpha with angles in radians. Doubling the camera length doubles the radius at the same angle. A tilted panel requires the direction from the specimen to each pixel instead of this perpendicular-panel formula.

A calibration specimen with known spacings gives known peak angles at a stated wavelength. NIST's SRM 640g certificate gives the silicon lattice parameter a=0.5431109 nma=0.5431109\,\mathrm{nm}. Its table gives a calculated first peak at 2θ=28.441∘2\theta=28.441^\circ for λ=0.15405929 nm\lambda=0.15405929\,\mathrm{nm}. The powder Lab uses the same lattice parameter to generate a synthetic standard at its own wavelength. It fits the detector distance and angle offsets from the standard image, then tests them with a separate exposure at another detector pose. The peak locations in the sample are inferred after that check.

2.3 A sequence of peaks tests the indexing

A cubic cell of side aa has plane spacings dhkl=a/h2+k2+l2d_{hkl}=a/\sqrt{h^2+k^2+l^2}, where (hkl)(hkl) indexes the plane family. Within a cell, atoms at different positions can scatter out of phase. In the model with identical atoms and a face-centered cubic lattice, the sum cancels for mixed odd and even indices. The first permitted families are (111)(111), (200)(200) and (220)(220). With a different basis, a different sequence may be allowed. Even an allowed peak may be too weak, overlap its neighbors, or lie outside the detector.

The simulated X-ray sample in Figure 2.1 uses λ=0.071 nm\lambda=0.071\,\mathrm{nm}. Its first fitted peak is at 2θ≃17.69∘2\theta\simeq17.69^\circ. For the (111)(111) assignment, Bragg's law gives d111≃0.231 nmd_{111}\simeq0.231\,\mathrm{nm} and the cubic lattice parameter follows as a=3 d111≃0.400 nma=\sqrt3\,d_{111}\simeq0.400\,\mathrm{nm}. The other visible peaks test this assignment at spacings different from the first.

  • Integrated simulated record
  • Nominal scattering angle
Powder diffraction profile and nominal scattering anglesThe calibrated simulated profile has six resolved peaks between 13 and 43 degrees. The first peaks near 17.69 and 20.45 degrees coincide with the first two vertical markers from the nominal cubic model.

Figure 2.1 Simulated X-ray counts integrated by scattering angle after detector calibration. The separate Bragg markers use the nominal face-centered cubic cell with a=0.400 nma=0.400\,\mathrm{nm}. The first two observed peaks lie near 17.69∘17.69^\circ and 20.45∘20.45^\circ.

2.4 Check a spacing two ways in the Labs

In Electron Diffraction & Reciprocal Space, choose Powder rings and Acquire counts. Keep the energy at 200 keV200\,\mathrm{keV}. The relativistic de Broglie wavelength is about 2.508 pm2.508\,\mathrm{pm}. Reduce Calibrated camera length from 600 mm600\,\mathrm{mm} to 500 mm500\,\mathrm{mm} and acquire again. The first indexed ring should shrink from about 3.77 mm3.77\,\mathrm{mm} to 3.14 mm3.14\,\mathrm{mm}. Convert both count-derived radii to angles and check that the spacing stays near 0.40 nm0.40\,\mathrm{nm}. The simulation knows which ring to inspect and uses a specified camera length. At 300 mm300\,\mathrm{mm} the neighboring powder rings overlap enough that the analysis withholds the spacing.

In Powder X-ray Diffractometer, start with One cubic powder. Use Keep exposure and note the first two angles under Peaks fitted in the observed angular profile. Switch to A shifted encoder zero, keep the new exposure, and compare the fitted angles as well as Withheld reference error. The raw counts are drawn at different detector poses relative to the encoder indications. The inferred lattice parameter in both cases should be near 0.400 nm0.400\,\mathrm{nm}. The record includes the simulated dark, flat-field, standard and independent check frames. The reduction uses the saved pixel counts, wavelength and encoder readings to infer the sample cell.

2.5 Exercises

  1. A first-order peak lies at 2θ=30∘2\theta=30^\circ for λ=0.150 nm\lambda=0.150\,\mathrm{nm}. Find dd.
  2. Keeping the wavelength and order fixed, does a smaller spacing give a larger or smaller scattering angle?
  3. A ring radius of 4 mm4\,\mathrm{mm} is measured at L=400 mmL=400\,\mathrm{mm}. Where should the ring appear at L=600 mmL=600\,\mathrm{mm}?
  4. At small angles, what happens to the inferred spacing if the camera length used in the calculation is too large by 5%5\%?
  5. Why does the powder pattern contain a ring, although one crystallite contributes in a small set of directions?
  6. A (200)(200) peak gives d200=0.200 nmd_{200}=0.200\,\mathrm{nm}. What aa and d111d_{111} does the cubic assignment predict?
  7. Could (100)(100) be the first line of the equal-scatterer face-centered cubic model? Explain.
  8. A standard agrees with the expected peaks at the pose used in the fit. What extra check tests whether the geometry also transfers to another pose?
Exercise 1

Use θ=15∘\theta=15^\circ. Then d=0.150 nm/(2sin⁡15∘)≃0.290 nmd=0.150\,\mathrm{nm}/(2\sin15^\circ)\simeq0.290\,\mathrm{nm}.

Exercise 2

For a permitted reflection, sin⁡θ=nλ/(2d)\sin\theta=n\lambda/(2d) grows as dd shrinks. The scattering angle increases.

Exercise 3

At fixed scattering angle r/Lr/L stays constant. The radius scales to 6 mm6\,\mathrm{mm}.

Exercise 4

The estimate is too large by about 5%5\%, since d≃λL/rd\simeq\lambda L/r is proportional to LL at fixed radius and wavelength.

Exercise 5

Different grains satisfy the same Bragg angle at different azimuths. Together they populate a cone, which makes a ring on a perpendicular detector.

Exercise 6

a=2d200=0.400 nma=2d_{200}=0.400\,\mathrm{nm}. This gives d111=a/3≃0.231 nmd_{111}=a/\sqrt3\simeq0.231\,\mathrm{nm}.

Exercise 7

No. (100)(100) mixes odd and even indices, so the amplitudes of equal scatterers in the face-centered cell cancel.

Exercise 8

Measure the known standard again at the other pose. Apply the original calibration without refitting it, then compare the recovered angles with the standard's predicted angles.

Test it in the Lab

Change one setting at a time, then compare what you see with the tables and answers in this chapter.

Electron Diffraction & Reciprocal Space →Tilt a thin crystal, record electron counts and measure the radius of a known diffraction spot or powder ring. Compare it with the geometric prediction, then halve the camera length and expose again. Low counts, overlapping peaks and saturated pixels limit what you can measure.Powder X-ray Diffractometer →Move an area detector, calibrate its geometry with a known powder and integrate saved photon pixels to test one or two cubic lattices.

Sources and reading scope

  1. International Union of Crystallography, Bragg’s Law

    The dictionary derives the constructive-interference condition and defines the angle measured relative to the lattice planes.

  2. NIST Certificate, SRM 640g Silicon Powder

    The certificate gives a calibrated silicon lattice parameter and explains its use for checking line positions and line shape.

  3. C. Davisson and L. H. Germer, Diffraction of Electrons by a Crystal of Nickel

    The original experiment records how electron scattering from a nickel crystal varies with angle and incident energy.

Keep reading

What this page links to directly

See this series on the content map →

Interactive Physics Textbook · Reading path

Learning the Microscopic World from Experiments

  1. ArticleWhat does an instrument record directly?
  2. You are hereHow does diffraction reveal crystal plane spacing?
  3. ArticleHow do two-dimensional images reveal a particle’s three-dimensional shape?
  4. ArticleHow does an X-ray spectrum identify elements?
  5. ArticleHow do counts reveal neutrino oscillations?
  6. ArticleHow does the end of a beta spectrum constrain neutrino mass?
1 more group

Other articles that use the same experiments

Each experiment from this article that another series also uses, followed by the articles there.

  1. LabElectron Diffraction & Reciprocal Space
  2. ArticleLaue and the Braggs: from spots to a crystal model
  3. ArticleElectron diffraction: two experiments, one wavelength test
  4. LabPowder X-ray Diffractometer
  5. ArticleHow powder rings reveal atomic spacing