IN28 · Follow a surface through force

Atomic Force Microscope

Does a scanned bump reveal the surface, the tip shape or a feedback error? An AFM image records how the scanner moves to hold a chosen force signal. A finite tip broadens narrow features, indentation lowers the apparent surface, and slow feedback leaves a tracking error. Known standards and scans at different speeds help separate these effects.

Acquire split-detector counts, raster commands and force curves, then reconstruct apparent height from calibrated feedback records.

Interactive modelAtomic Force Microscope

A retained feedback scan

Calibration, force inference and retained records
Withheld deflection error δd/nm\delta d/\mathrm{nm}00
Withheld force relative error δF/F\delta F/F00
Calibrated scanner scale szs_z00
Conditional reduced modulus E∗/GPaE^*/\mathrm{GPa}00
Approach/retract relative difference δE∗/E∗\delta E^*/E^*00
Observed contact duty DD00
Independent specimen reference

The retained feedback row is compared with a separate model specimen and finite-tip envelope. Neither reference supplies image pixels or fitted material values.

Independent raw calibration and measurement acquisitions
AcquisitionFramesPhotoelectronsSeed
Immutable retained scans
RecordModeCalibrationSupported image

Rigid known deflections calibrate the nonlinear split-detector signal. A reference cantilever calibrates stiffness; known plane and step scans calibrate height. Separate deflection, force and step checks are withheld. A force curve fits reduced modulus and an unknown contact origin from measured commands and deflection, conditional on the known tip radius.

Apparent height includes indentation and feedback lag. The local Hertz regression error treats the force axis as exact and omits calibration/count correlations, radius, stiffness-standard and material-model uncertainty. It applies to the flat nonadhesive patch; sharp-edge geometry and real adhesion require a richer model. CSV and JSON retain full commands, counted detector frames, clocks, definitions, seeds and hardware. CSV alone replays the inverse.

For lateral stick–slip measured with a probe, see single-contact AFM friction.

Physics tutorial

A height image made by a force-control loop

BackgroundA sharp tip pushes against the sample and bends a spring-like cantilever. A reflected laser beam converts its slope into a shift on a split photodetector.

Why it mattersThe scanner must move vertically to hold a force signal steady. Its recorded movement depends on the sample, tip and ability of the loop to follow changes.

Start with the essentials

Focus question
Does a scanned bump reveal the surface, the tip shape or a feedback error?
One-sentence intuition
Calibrate the detector and scanner with known standards, then compare acquired height with the simultaneously recorded deflection error.

Core mathematical model

Repulsive elastic contact on a flat patch

F=kq=43E∗R δ3/2F=kq=\frac{4}{3}E^*\sqrt{R}\,\delta^{3/2}

Upward deflection qq is positive; repulsive load is FF. The known-radius spherical apex uses reduced modulus E∗E^* and indentation δ\delta. The Hertz approximation assumes elastic nonadhesive contact and shallow indentation relative to RR.

The counted optical-lever signal

u=C+−C−C++C−u=\frac{C_+-C_-}{C_++C_-}

Dark-corrected photoelectron counts from the two detector halves are C+C_+ and C−C_-. Independent known deflections calibrate this nonlinear imbalance; the displayed gain does not enter the physical counts.

The sampled integral controller

zn+1=zn+g(qobs,n−qset)z_{n+1}=z_n+g(q_{\mathrm{obs},n}-q_{\mathrm{set}})

The scanner command zz increases upward. Excess observed deflection raises the head and reduces the load. Gain gg depends on bandwidth and update period. In cyclic-force mode the observation is the measured peak within one off-resonance cycle.

Finite-tip geometry before force detection

htip(x,y)=max⁡a,b[hs(a,b)−t(a−x,b−y)]h_{\mathrm{tip}}(x,y)=\max_{a,b}\left[h_s(a,b)-t(a-x,b-y)\right]

The undeformed specimen is hsh_s and the tip height above its apex is tt. This forward geometric envelope determines where contact begins. Image pixels are generated separately from the retained scanner commands.

Common difficulties

The image does not directly give the sharp surface

Typical misconceptionA perfectly calibrated height signal removes the tip shape and all feedback error.

Better mental modelCalibration fixes the scale. Finite-tip contact, indentation and tracking still change the relationship between the recorded movement and the undeformed surface.

One low drive frequency is not a complete dynamic guarantee

Typical misconceptionOperating below the free resonance automatically makes every contact quasi-static.

Better mental modelA brief contact can change force much faster than the drive period. This mode restricts contact duty and compares its supported settings with a separate inertial contact calculation.

Run the experiment

  1. 01

    Follow a bump

    Keep a scan and switch between acquired height and command/deflection plots.

    What to observe: The same recorded scan produces both channels. Where the scanner falls behind, the deflection error grows.
  2. 02

    Change the tip

    Choose a blunt tip or change radius, then acquire matching standards.

    What to observe: A wider tip touches a feature with its side before the apex reaches it, broadening the apparent feature.
  3. 03

    Outrun the loop

    Lower bandwidth or increase line rate, then compare the measured image with the separate specimen reference.

    What to observe: Lag reduces the apparent height of narrow features and shifts their edges. Slower scans approach the finite-tip profile.
  4. 04

    Measure a force curve

    Open approach/retract data and calibration details. Try the softer-surface or cyclic-force preset.

    What to observe: A flat-patch Hertz fit estimates reduced modulus conditional on the known radius and observed calibration. Cyclic force uses peak deflection at low frequency; the force curves themselves are acquired with the drive off.

Related experiments and articles

How physics instruments measure · Reading path

How physics instruments measure