An optical trap pulls a bead toward the light focus. A tether pulls it away. Once the trap stiffness is known, the bead’s displacement measures force. Thermal motion provides part of that calibration: its amplitude and relaxation reveal how strongly the bead is confined.
Take rightward as positive. The trap center is X and the bead center is x, both in nanometers. Near the focus, the force on the bead is Ftrap=ktrap(X−x). Stiffness is measured in piconewtons per nanometer. A bead left of the focus feels a rightward force.
A tether attached on the left supplies the opposing force. At equilibrium the two forces balance. Without a tether, the mean bead position returns to the focus. Calibrating that simpler configuration helps us recognize the extra force when a molecule is attached.
1pN=10−12N,1nm=10−9m
Use compatible length, force and energy units throughout a calculation.
Water molecules continually kick the bead. Viscous drag also removes its momentum quickly. On the millisecond time scale used here, we neglect inertia: the mean velocity under a force is that force divided by the drag coefficient. Thermal kicks keep the position fluctuating around its mean.
The trap potential near the focus is a bowl. Displacement costs energy quadratically. At thermal equilibrium, this positional degree of freedom stores half of kBT on average. Here kB is Boltzmann’s constant and T is absolute temperature. The result converts position variance into stiffness.
U=21ktrap(x−X)2,⟨(x−X)2⟩=ktrapkBT
The record must sample equilibrium and retain the relevant thermal fluctuations. A slow detector suppresses the measured variance.
Use kBT=4.11pNnm near room temperature. A free bead has a true positional standard deviation of 10.0nm, giving a variance of 100nm2. Dividing thermal energy by that variance gives ktrap=0.0411pNnm−1.
Attach a tether. Its mean displacement from the trap center is now 40.0nm to the left. In the linear trap range, the equilibrium tension is 1.64pN. Before multiplying, check that trap stiffness and detector calibration have remained unchanged. Laser power, bead height and optical sensitivity can alter the calibration.
Many samples can still contain little new information
Neighboring samples often belong to the same unfinished fluctuation. After a displacement, the mean offset relaxes exponentially. Its time constant is drag divided by stiffness. Doubling the sampling rate does not automatically double the independent information.
To assess random uncertainty, divide a long record into blocks much longer than the relaxation time and compare their estimates. An error shared by the entire calibration survives this averaging. Attaching DNA adds another restoring force; we will need to reconsider the fluctuations.
τ=ktrapγ,C(t)=⟨δx2⟩e−∣t∣/τ
Drag in piconewton milliseconds per nanometer gives a time in milliseconds. This is a linear overdamped model; nearby surfaces alter drag.
Predict whether a stiffer trap increases or decreases the fluctuations. Change stiffness alone and compare the position record and power spectrum.
Keep the existing free-bead geometry. Record the calibration method before moving to the tether experiment in the next chapter.
Exercises and explanations
A detector reduces the true ten-nanometer standard deviation to eight nanometers. Does a direct equipartition calibration overestimate or underestimate stiffness?
Read the explanation
It overestimates stiffness. The variance falls from one hundred to sixty-four square nanometers, so the inferred stiffness is 1.56 times the true value. A longer record reduces sampling scatter but cannot restore frequencies removed by the detector.
What happens if you calibrate a tethered bead with the free-bead variance formula?
Read the explanation
The tether supplies an additional restoring force. The smaller variance makes the inferred trap stiffness include molecular stiffness. Chapter three separates the stiffnesses relevant to fluctuations and to pulling.
Figure 1 defines the surface–DNA–bead geometry; the elasticity analysis separates entropic and stretch compliance. Our collinear, inextensible teaching model uses its own parameters and synthetic records. No experimental points are digitized.
The apparatus and feedback discussion motivate separating applied load, bead response and detector response. The unfolding and motor experiments assume ideal prescribed-force control; they do not simulate this complete feedback apparatus.