Direct wave plus one scattered wave
The assumed scattering amplitude is strongest in the forward direction. The direct and scattered paths differ by the neighbour distance times the angular path factor.
E64 · Structural photoemission
What can photoelectron diffraction and standing-wave scans tell you about where atoms sit? An angle scan of photoelectron diffraction constrains the directions to nearby atoms, and a standing-wave scan constrains positions along one periodic direction. Neither tells you more than the measurement and its assumptions pin down. Counting noise, the range covered and fits that are not unique each need a separate check.
Take angle-resolved diffraction scans or standing-wave phase scans, fit the stored core-level counts, and see how much of the atomic structure those counts can really pin down.
Physics tutorial
BackgroundA photoelectron wave leaving an atom interferes with the part scattered by a neighbour, so the counts change with emission angle. In a standing-wave scan, the counts change with the phase of the standing wave instead. This Lab records both kinds of scan and fits the stored counts.
Why it mattersA fit can match the data well and still leave the structure undetermined. Scan coverage, count statistics and the model assumptions decide what a scan can actually tell you.
Start with the essentials
The assumed scattering amplitude is strongest in the forward direction. The direct and scattered paths differ by the neighbour distance times the angular path factor.
The depth distribution is a Gaussian cut off at the surface. The effective escape length includes the cosine of the takeoff angle. The coherent position does not have to equal the mean depth.
The constant, cosine and sine coefficients are fitted to the stored counts, using the calibrated reflectivity. The overall amplitude is an unknown nuisance parameter.
Equivalent positions are one period apart. The unknown layer abundance can absorb any change in attenuation, so one scan cannot pick out the absolute depth.
Typical misconceptionA low residual proves there is only one possible atomic structure.
Better mental modelThe models here are synthetic teaching models: one surface emitter and one scatterer with an assumed real scattering amplitude, or an ideal standing wave with an independently controlled phase at fixed reflectivity. They leave out element-specific phase shifts, multiple scattering, a realistic Bragg rocking curve and nondipole corrections. The layer signal is weighted by escape. The coherent position repeats every period, so it cannot give a unique absolute depth. The apparatus, atomic spacing and layer thickness are drawn at different schematic magnifications.
Typical misconceptionThe residual map gives a confidence interval or the experimental resolution.
Better mental modelThe unweighted least-squares map only describes the fit. The angular acceptance is calibrated independently, and energy-dependent scale factors are fitted. The standing-wave fit shows noisy coherent fractions above one as they are, and it rejects rank-deficient or grossly unphysical fits. All counts are synthetic, and no uncertainty or external certification is claimed.
Acquire an angular scan and move the selected point across the recorded angles.
What to observe: The collector follows the signed angle from the outward normal. The direct wave and the singly scattered wave interfere.Compare Completed angular scan with Narrow angle coverage and Too few counts.
What to observe: The search grid shows the relative residual over distance and tilt. A sharp minimum holds only within the assumed one-neighbour model.Acquire and fit a phase scan, then compare it with Broad layer distribution and No reflected wave.
What to observe: When the depths spread across several phases, the coherent modulation weakens. With no reflection there is no phase information.Compare Completed phase scan with One period deeper.
What to observe: The phase repeats while the absolute counts drop. Because the abundance is unknown, the absolute depth is not unique. The two examples use separate synthetic samples.