Spin orientation dynamics
Torque sets precession, relaxation removes orientation, and the pump replenishes it along the optical axis. Pumping also increases the total relaxation rate. Rates and optical gain are declared teaching assumptions.
IN34 · Read the spin rhythm
How can light read atomic precession without mistaking brightness drift for a magnetic field? A magnetic field sets the atomic precession frequency. A weak probe turns that motion into a changing polarization; two optical channels and an independent brightness monitor separate common probe drift. Pump drift changes the spins themselves and needs a separate check.
Drive polarized cesium spins with an RF field and sweep through resonance. Compare two counted optical channels, probe drift and the separate static weak-field response.
Cosine channel Negative-sine channel
·
Collector ACollector B
· photons per gate
CSV retains every time gate, two detectors, probe and pump monitors, signed references, frequency or coil readback, dark observations and three known analyzer angles. The fit reads these observations only. Mx fits a common centre and absorptive width with nuisance amplitudes and phases. Static inference assumes independently known pumping and relaxation and stays inside the monotonic branch.
| Gate | A | B | Probe monitor | Pump monitor |
|---|
The ideal reference is generating context stored separately in JSON. It is excluded from the raw CSV and from every inference.
Physics tutorial
BackgroundCircular pump light orients cesium spins. A transverse RF field drives their precession, and a weak probe reads the resulting optical rotation.
Why it mattersThe detectors count photons. A field interpretation needs to distinguish atomic motion from changes in the light source.
Start with the essentials
Torque sets precession, relaxation removes orientation, and the pump replenishes it along the optical axis. Pumping also increases the total relaxation rate. Rates and optical gain are declared teaching assumptions.
One ground-state hyperfine manifold is selected. A resolved resonance centre becomes a field magnitude through the known gyromagnetic ratio.
Dark observations and independent rotation standards convert the detector counts to rotation. The saved cosine and negative-sine references then give RMS-scaled signed channels. A cosine signal leading the reference gives positive quadrature.
RF is off, the pump points along the positive longitudinal axis and the field is transverse. Rotation is approximately linear near zero and eventually turns over. Inference uses only the declared monotonic interval.
Typical misconceptionNormalization removes every laser-power change.
Better mental modelCommon probe brightness multiplies both outputs and can cancel. Pump power changes the spins and linewidth, so it needs a separate check.
Typical misconceptionSetting the resonance frequency to zero makes a zero-field magnetometer.
Better mental modelThe rotating-wave approximation needs sufficiently fast precession. The static branch solves the RF-off equation directly; collision physics for SERF is absent.
Hold the gold bias dial and drag horizontally.
What to observe: The resonance centre moves with the field; the spin arrow and newly acquired counts update during the drag.Choose Probe brightness drifts, then Pump power drifts.
What to observe: Normalization preserves the resonance under common probe drift. The independent pump monitor detects pump drift and withholds a stationary inference.Choose Look along the field, then return to Driven precession.
What to observe: A probe along the static field cannot see transverse precession. Photon counts remain available, but the magnetic conclusion is withheld.Choose Static weak field and slowly change the bias field.
What to observe: With RF off, rotation changes across the calibrated coil sweep. Leaving the monotonic branch withholds a unique field.Keep a sweep, change pumping, select the earlier record and export CSV.
What to observe: Its time gates, standards and counts stay fixed. The CSV contains no generating field and can still reproduce the inference.Continue exploring
How does prepared atomic orientation become an optical magnetic-field measurement?