Two readouts, one update
Subtract the calibrated intercept from the high-minus-low excitation fraction and divide by the measured local slope. Gain sets the fraction corrected per pair. Far from the line the discriminator loses direction.
IN35 · Ask atoms which way to tune
What turns a narrow atomic line into a working clock? Atoms tell a laser which way to tune when successive probes on either side of a resonance excite different fractions. Feedback keeps the laser near the line centre. Independent measurements of environmental shifts are still needed to connect that centre to an unperturbed atomic frequency.
Interrogate trapped atoms on both sides of a resonance, retain classified atom counts and servo commands, and calibrate one thermal shift.
| Pair | Low excited / total | High excited / total | Before / after (Hz) | Start (s) |
|---|
Continue with Ramsey spectroscopy · the optical frequency ruler
Physics tutorial
BackgroundLattice light confines the atoms. A separate clock laser probes whether they absorb light. A longer coherent pulse distinguishes closer frequencies.
Why it mattersA low excitation fraction can occur on either side of a resonance. Two probes give the direction needed for feedback.
Start with the essentials
Subtract the calibrated intercept from the high-minus-low excitation fraction and divide by the measured local slope. Gain sets the fraction corrected per pair. Far from the line the discriminator loses direction.
Use the uniformly spaced latter half of retained commands, normalized by the external nominal carrier. Differences between overlapping adjacent averages measure command fluctuations and cancel constant offsets. No confidence interval, dead-time correction or actual optical-output measurement is supplied.
Separate reference observations fit the fourth-power coefficient, excluding a withheld temperature. Apply the static correction only within the calibrated temperature interval and for a homogeneous enclosure.
Typical misconceptionA quiet command trace proves an absolutely accurate atomic frequency.
Better mental modelDifferencing adjacent averages cancels constant errors. This statistic describes servo commands; continuous output and omitted shifts require independent measurements.
Watch classified atom counts and the next command. Drag the gold frequency dial on the apparatus.
What to observe: The excitation imbalance changes and feedback moves the command. Preview counts and atom colours update during the drag.Choose the few-atoms demonstration and compare excitation fractions and commands.
What to observe: Binomial fluctuations increase when fewer atoms make each decision.Lengthen the pulse, retain a mismatched record, then acquire matching standards.
What to observe: The old calibration belongs to the old pulse. Inference rejects the mismatch and keeps the raw counts.Warm the enclosure and inspect independent thermal references and the corrected command.
What to observe: Temperature moves the whole resonance. Allan deviation is insensitive to a constant offset.How physics instruments measure · Reading path
Articles on this path that use this experiment, followed by the other experiments they suggest.