Can you stop charging without hiding the surface? Only if the charge has a path to ground. A conductive coating drains charge only when it is actually connected to ground. Waiting, or pumping the chamber to a lower pressure, will not remove a layer that has already been deposited on the surface, and that layer can hide the detail you wanted to see.
Compare a grounded coating with the same coating disconnected from ground. Run one scan starting from an uncharged surface, then blank the beam: the charge drains away, but the distortion in the scan and the deposited layer stay.
Why can the sample chamber stay at higher pressure than the pump inlet? Gas must cross the pipe before the pump can remove it. A narrow pipe limits the effective pumping speed. Two calibrated gauges reveal that pressure difference; helium counts, compared with standards that cross the same route, test a tracer-responsive leak path.
Follow gas through the chamber, valve and pump. Switch pressure-gauge principles and compare pressures with saved helium ion gates.
Early readable interval and the first withheld point. The readable-point plot expands the first 10 seconds; the raw-voltage plot keeps the full run.
Actual acquisition times · diaphragm indication
Time / s
Pressure / Pa
Standard uncertainty / Pa
Raw channels and calibration
Exports retain voltages, integer ion gates, gas and valve commands, exact gate times, reference flow temperatures, blank and heldout observations. JSON also labels the separate generating reference. CSV contains raw observations and calibrations only.
Scroll the table sideways to see every column.
Last six gates in the selected retained run
End / s
Chamber / V
Pump / V
He ions
Valve
Diaphragm / V
Supported / Pa
Physics tutorial
Can you stop charging without hiding the surface?
BackgroundCompare a grounded coating with the same coating disconnected from ground. Run one scan starting from an uncharged surface, then blank the beam: the charge drains away, but the distortion in the scan and the deposited layer stay.
Why it mattersA conductive coating needs a path to ground. Waiting after an unstable scan does not correct the image already formed.
Start with the essentials
Focus question
Does a coating still drain charge when its ground contact is open?
One-sentence intuition
What matters for the image is the charge during the scan, not what is left after blanking, and neither waiting nor lower pressure removes a layer already deposited.
Core mathematical model
Charging up from an uncharged surface
CdtdV=−Inet−GV,V(t)=−GInet(1−e−t/τc),τc=GC
The capacitance is fixed at 2 pF. The net negative current is 35 percent of the incident current, times the duty fraction. That averaged duty fraction comes from the dwell time and an assumed 10 microsecond blanking interval per pixel; the rapid on-off switching itself is not resolved.
Charge draining after the beam is off
V(T+toff)=V(T)e−toff/τc,G=G0+cGcoat
The contact switch is one when the coating is connected and zero when it is open. The bare specimen's own leakage stays in both cases. Conductance here means the effective path to ground, not the bulk conductivity. Waiting changes the final potential, not the scan that came before.
Deposited layer and surface masking
h(t)=rdepmin(t,T),S(t)=exp[−15nmbcoat+h(t)]
Only beam-induced deposition is modeled, and material already deposited stays during recovery. Each scan row uses the state of the surface at the time that row was scanned. The masking and the displacement scale are illustrative; they are not a resolution calibration or an electrostatic transport calculation.
Common difficulties
Waiting afterwards does not fix the image
Typical misconceptionA small final potential means the preceding scan was stable.
Better mental modelThe target uses the potential at the end of the scan. The recovery time does not change the scan image, and it does not clean away deposited material.
What the model covers
Typical misconceptionThis is a calibrated SEM charging or coating calculation.
Better mental modelThis is an RC circuit model, averaged over the beam duty cycle and started from a fresh surface; it is not recorded SEM data. The capacitance is fixed and the net current is negative. The coating's continuity and conductance are assumed. Beam-induced deposition stops during recovery, but material already deposited stays. Gas transmission, scan displacement, surface masking and the spectral markers are illustrative. Ripple from individual pixel pulses, neutralization by gas ions, positive charging, self-consistent fields and material calibration are not modeled.
Run the experiment
01
Compare a grounded and a floating coating
Select Thin carbon coat, then Disconnected coat. The film thickness and masking stay the same, but the charge drainage changes.
What to observe: A floating coating still shows its spectral marker and still masks the surface, but it no longer adds a path to ground.
02
Blank the beam after the scan
Select Bare scan, then blank. Compare the potential at the end of the scan with the final residual potential, then change the Beam-off recovery duration.
What to observe: The charge decays once the beam is off, but the scan image and the contamination stay the same, and an unstable exposure still fails the target.
03
Find a stable preparation
During the scan, keep the potential's magnitude below 2 V, the contamination layer below 1 nm and the surface contrast above 75 percent. Recovering afterwards cannot make an unstable exposure pass.
What to observe: A thin, grounded carbon coat meets this teaching target. Thick gold hides surface detail, and a poor vacuum raises the assumed deposition rate. Lowering the pressure alone does not ground an insulator.
BackgroundMolecules must find the chamber exit and cross valves and pipes before the pump removes them. Walls and trapped pockets can keep supplying gas.
Why it mattersThe pump inlet can have low pressure while the sample chamber stays higher. Two gauges reveal the connecting route.
Start with the essentials
Focus question
Does a larger pump or a wider pipe lower chamber pressure more?
One-sentence intuition
Effective speed depends on the entire gas route. Gauge response depends on composition, and helium flow requires standards that travel the same path.
Core mathematical model
Gas inventory in each volume
piVi=niRT,Vip˙i=Qi+j∑Cij(pj−pi)−Sipi
At fixed gas temperature, nitrogen and helium inventories evolve separately. Every internal flow enters one volume and leaves its neighbour; pumping removes gas at the inlet.
Connecting the pump to the chamber
Seff=C+SCS,QpV=n˙RT
A narrow route makes effective speed approach its conductance. Converting molar flow to pressure-volume throughput uses the recorded gas temperature. Throughput is not heating power.
Common difficulties
Nitrogen-equivalent pressure
Typical misconceptionA voltage trace directly gives the true total pressure of any mixture.
Better mental modelIonization response depends on gas species. An unidentified mixture retains a nitrogen-equivalent pressure until composition is independently known.
Allowing the signal to recover
Typical misconceptionResidual helium can be ignored as soon as tracer delivery stops.
Better mental modelThe acquired reference tail must lie within two percent of the late-pulse response relative to the blank, allowing for both signs and count uncertainty. Incomplete recovery withholds helium-flow inference.
Run the experiment
01
Compare a narrow route with a larger pump
Compare Open path, Narrow path and Larger pump.
What to observe: Conductance changes the pressure difference and evacuation curve.
02
Move the valve while the apparatus runs
Drag the gold handle, calibrate the path again and keep the run.
What to observe: The network updates before release. Old plots and exports retain their original settings.
03
Mark gas outside the chamber
Compare External leak with Trapped gas and select helium ion counts.
What to observe: An external tracer path produces helium response. Absence of that response does not establish a virtual leak from pressure alone.