Model parameters need calibration
Typical misconceptionA preset that works here is a recipe I can use on a real instrument.
Better mental modelThis teaching model stores both expected dose and a seeded stochastic exposure. In stochastic mode, electron arrivals at write sites and resist reactions are sampled separately; the assigned reaction yield is 0.02, followed by a Gaussian resist-response correlation. Development thresholds this response on a 10 nm grid. A new write number produces an independent exposure. Line-edge positions use linear interpolation along 16 rows over 160 nm, and RMS is computed about each edge mean; interrupted lines do not receive a complete-line roughness value. These are calculated resist responses, without microscope metrology noise or material calibration. Scattering is modeled by two normalized Gaussians, with the broad one assigned a weight of 0.6 relative to the narrow one, and their widths are not measured material parameters. The positive resist opens wherever the resist response reaches a sharp threshold, on a 10 nm grid. The correction evens out the dose inside the pattern before coarse shot sampling is applied; it is neither an exact inversion nor a general proximity correction. The model leaves out Monte Carlo electron transport, chemical development kinetics, charging, drift and pattern transfer. Display gain does not change the data. At fixed dose, the beam current changes the writing time. The beam paths show the geometry at the start of exposure, and the relief shows the developed resist. The hardware and the lateral and vertical scales are enlarged independently.