Diamond Anvil Cell

How can we tell the pressure inside a tiny compressed sample chamber? Ruby fluorescence shifts toward red under pressure. Comparing a grain with its own ambient spectrum gives a local pressure when temperature and stress conditions match the pressure scale. Several grains reveal spatial differences that the applied load alone cannot tell us.

Illuminate five ruby grains between opposed diamonds, retain their photon spectra and infer local pressures from independently checked wavelength references.

Interactive modelDiamond Anvil Cell

Simulated photon counts · room-temperature ruby scale

Five grains test whether the pressure agrees

Selected grain: actual binned counts and fitted doublet

N/countN/\mathrm{count}

λair/nm\lambda_{\mathrm{air}}/\mathrm{nm}

Recorded countsBlind doublet fit

Separate local pressure estimates

P/GPaP/\mathrm{GPa}

x/μmx/\mathrm{\mu m}

Counting errors only; five separate sampled positions.

Reference checks, every grain and raw channels

Bars show conditional counting standard errors. They omit pressure-scale, grain, temperature and model systematics; the five points do not reconstruct an arbitrary pressure field.

Local fits from the selected observations
x/μmx/\mathrm{\mu m}λ1/nm\lambda_1/\mathrm{nm}P/GPaP/\mathrm{GPa}ucount/GPau_{\mathrm{count}}/\mathrm{GPa}bT/GPab_T/\mathrm{GPa}

The temperature allowance is separate from counting precision. Ambient spectra belong to these same grains; independent wavelength and pressure checks remain excluded from the fitted references.

All raw channels of the selected grain
ChannelPhotonsClipped
Separate synthetic source reference

The load and gasket response generate this teaching specimen. They never enter the spectral pressure inverse. A saved JSON keeps this source description separately; the raw CSV replays without it.

Physics tutorial

Pressure read from a tiny ruby grain

BackgroundOpposed transparent diamonds compress a gasket around a small sample chamber. A ruby grain inside the chamber emits a red doublet when illuminated by green light.

Why it mattersThe force on the outside passes through diamonds, seats and a deforming gasket. A local optical gauge helps establish what reached the enclosed sample.

Start with the essentials

Focus question
How can we tell the pressure inside a tiny compressed sample chamber?
One-sentence intuition
Use the same grain’s ambient spectrum and independent wavelength references, then compare several local readings under qualified temperature and stress conditions.

Core mathematical model

The adopted room-temperature pressure scale

P=AB[(λR1λR1,0)B−1]P=\frac{A}{B}\left[\left(\frac{\lambda_{R_1}}{\lambda_{R_1,0}}\right)^B-1\right]

The Mao–Xu–Bell 1986 quasi-hydrostatic scale uses A=1904,mathrmGPaA=1904,mathrm{GPa} and B=7.665B=7.665. The ambient wavelength λR1,0\lambda_{R_1,0} is measured for the same grain. This teaching model supports zero through twenty gigapascals and temperature matched to the reference.

A counted channel integrates part of a spectral line

E[Cj]=b+∑iNiπ[tan⁡−1 ⁣2(λj,+−λi)Γi−tan⁡−1 ⁣2(λj,−−λi)Γi]\mathbb E[C_j]=b+\sum_i\frac{N_i}{\pi}\left[\tan^{-1}\!\frac{2(\lambda_{j,+}-\lambda_i)}{\Gamma_i}-\tan^{-1}\!\frac{2(\lambda_{j,-}-\lambda_i)}{\Gamma_i}\right]

Each detector channel has finite wavelength edges. Total line photons NiN_i, center λi\lambda_i and Lorentzian width Γi\Gamma_i are fitted to the observed counts with a baseline. Line centers are found before any pressure is assigned.

Only a finite cone reaches the receiver

nmsin⁡θm=sin⁡θa,ηcol=12∫0θm,max⁡TmdTdasin⁡θm dθmn_m\sin\theta_m=\sin\theta_a,\qquad \eta_{\rm col}=\frac12\int_0^{\theta_{m,\max}}T_{md}T_{da}\sin\theta_m\,d\theta_m

Refraction through the medium and diamond sets the accepted cone. Fresnel transmission factors enter the photon fraction. The objective collimates that cone, and the dichroic redirects it onto the actual spectrometer face.

Compare the individually measured sites

ΔPsites=max⁡iPi−min⁡iPi\Delta P_{\rm sites}=\max_i P_i-\min_i P_i

The pressure span is computed from the five acquired spectra. A small span at those grains does not prove every part of the specimen has the same stress.

Common difficulties

External force does not determine the local pressure by itself

Typical misconceptionDividing the applied force by the diamond tip area tells the specimen pressure.

Better mental modelThe gasket supports and redistributes force. A calibrated local marker measures the enclosed conditions separately.

A redshift can have more than one cause

Typical misconceptionEvery redshift means that pressure increased.

Better mental modelTemperature and nonhydrostatic stress also change ruby spectra. A matched zero, temperature measurement and qualified pressure scale are part of the inference.

Run the experiment

  1. 01

    Follow the light through the cell

    Switch between the cell, ruby chamber, laser and collector views.

    What to observe: The green beam crosses the upper diamond; red fluorescence returns through the same opening and reaches the spectrometer.
  2. 02

    Increase the applied load

    Drag the upper diamond while holding the pointer, then keep the spectra.

    What to observe: The gasket gap changes and the measured doublet shifts. The local pressure is fitted from light, rather than supplied by the load control.
  3. 03

    Compare five ruby grains

    Choose the spatial-gradient demonstration and inspect the local-pressure plot.

    What to observe: The fitted pressures differ across the chamber even though one external load is applied.
  4. 04

    Test the pressure conditions

    Try the rigid-medium and warm demonstrations, then open reference checks.

    What to observe: Broadened lines or unmatched temperature withhold a pressure conclusion while the counted spectra remain visible.

Related experiments and articles

How physics instruments measure · Reading path

How physics instruments measure

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleHow a diamond anvil cell reads pressure in ruby light
FeedbackSomething broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy