Steady-state absorption and dispersion
These are the normalized magnetization components in the rotating frame. The detuning is the sample frequency minus the drive frequency, and the bridge phase puts the absorption signal in the in-phase channel.
Continuous nuclear resonance / Line shape
How do you find the center and width of a nuclear resonance with a continuous radio-frequency scan? Step the drive frequency and record two signals, one in phase with the drive and one a quarter cycle out of phase. Absorption peaks at resonance; dispersion changes sign there. Both share one center and width. Remove the baseline, then measure the width at half height. A strong drive saturates the spins and broadens the line, so only a weak drive reveals the intrinsic transverse relaxation time.
Balance a radio-frequency bridge, step the drive frequency, and record the absorption and dispersion voltages. From those readings, estimate the center of the resonance and its power-broadened line width.
Nominal vertical static field, transverse sample coil. Continuous RF and a matched reference arm feed a balanced bridge; the residual reaches a phase-referenced receiver. Ideal cancellation and flat calibrated gain, not a solved RF circuit. Cable motion is schematic.
Continuous drive, voltages averaged at steady state
121 independent settled exposures span the drive offset from minus 180 to plus 180 hertz. Each point uses the selected fixed RF amplitude. This is a stepped steady-state scan; the accelerated playback does not model rapid passage or finite dwell.
Ready. Predict, then acquire.
Raw voltages, with the source expectation shown separately
Teal: absorptive in-phase voltage. Gold: dispersive quadrature voltage. Dots: observations. Thin lines: source expectations. Purple dashes: record-only fit. Gray: exact saved records. Constant receiver baselines are retained and fitted.
Residuals computed from the retained voltages
Only recorded coordinates appear, and no residual is drawn until a supported fit exists. A small residual means the data are consistent with this family of responses. It does not independently confirm the physical law.
Inference without the source parameters
Inferred resonance offset
Inferred full width at half height
Inferred absorption peak above baseline
Retained complex exposures
Voltage residual RMS
Unknown channel amplitudes and baselines are profiled out. The line width includes drive saturation, so no lifetime is inferred from it. Search bounds and weak-information thresholds are diagnostic heuristics, not confidence intervals.
| Index |
|---|
Saved comparison, model limits and sources
The model has a single homogeneous, noninteracting resonance with exponential Bloch relaxation. It leaves out static field distribution, molecular shifts or coupling, heating, diffusion, radiation damping, finite settling, bridge imbalance, drive leakage, phase error and RF circuit loading. Receiver gain is assumed flat over the narrow carrier span. The bridge measures response voltage, not absorbed watts or individual nuclear outcomes.
Physics tutorial
BackgroundBalance a radio-frequency bridge, step the drive frequency, and record the absorption and dispersion voltages. From those readings, estimate the center of the resonance and its power-broadened line width.
Why it mattersYou cannot read a constant straight off the instrument. You infer it from the measurements you keep, using the calibrated response.
Start with the essentials
These are the normalized magnetization components in the rotating frame. The detuning is the sample frequency minus the drive frequency, and the bridge phase puts the absorption signal in the in-phase channel.
This is the full width at half maximum of the absorption peak after the baseline is subtracted. Only for a weak drive and a homogeneous line does the inverse width give the intrinsic transverse time.
The amplitude and constant baseline of each channel are unknown and are fitted along the way. The center and half-width are searched together; neither the source relaxation times nor the drive enters the fit.
Typical misconceptionGetting the source parameter back independently confirms the law.
Better mental modelThe data were generated from that law, so getting the parameter back tests the measurement and estimation steps. The data are not historical observations, they do not give a confidence interval for the parameter, and they are not external scientific certification.
Choose the Reference line preset and scan. Compare where the absorption peaks with where the dispersion signal, after baseline subtraction, changes sign.
What to observe: Both signals share one center and one width. The dots are the recorded voltages; the thin curves show the source expectation separately for comparison.Save the run, choose the Power broadened preset and scan again.
What to observe: The line gets wider even though the source transverse time has not changed. The estimate reports this broadened width.Turn the RF drive off and run the same scan.
What to observe: A baseline and noise alone cannot locate a resonance, so the weak-information check withholds the line estimate.Quantum evidence · Reading path