Internal mixture and HOM visibility
The two independent inputs have identical internal density operators. At a balanced splitter and matched envelope and polarization, a maximally mixed two-mode input has a coincidence floor of one quarter.
Q036 · Two photons meet at a beam splitter
When two photons enter a beam splitter from opposite sides, why do the two counters stop clicking together? Two photons in the same pure mode leave a balanced beam splitter together because the alternatives cancel. Identical mixed preparations still contain distinguishable modes, so coincidences remain. Delay and polarization mismatch can also weaken the dip.
Send two single photons into the same beam splitter and predict what happens when they overlap. Change their arrival delay or polarization, then collect detector clicks one by one and build the interference dip from them.
Simulated acquisition
Acquire events to begin.
Both in A
One in each port
Both in B
Gray: ideal single-pair reference before loss or electronic gating. Orange: coincidence count per emitted trial, with 95% Wilson intervals. The references and acquired points need not coincide.
Expand the fit and raw trials to inspect the numerical evidence.
Count-only Gaussian fit
Visibility / baseline from counts
Complete the scan, or acquire all four settings, to see an inference.
| Delay (ps) | Trials | Coincidences | Fraction |
|---|
Physics tutorial
BackgroundOne photon enters each input port. There are two ways to end up with one photon in each output: both photons reflect, or both pass through. When the splitter is balanced and the input modes are identical, the amplitudes for these two ways cancel.
Why it mattersThe coincidence dip tests how indistinguishable the two input modes are. Detector loss, accidental clicks and extra photon pairs can change the recorded dip, but they do not change the ideal interference law for a single pair.
Start with the essentials
The two independent inputs have identical internal density operators. At a balanced splitter and matched envelope and polarization, a maximally mixed two-mode input has a coincidence floor of one quarter.
Both inputs have Gaussian spectra with the same intensity standard deviation, in THz. Delay is in ps, so their product is dimensionless. The envelope view is only a sketch of the modes.
A relative delay, a frequency offset or a polarization difference each makes the inputs distinguishable. The Gaussian envelopes have equal width. A separate internal two-mode state multiplies this overlap by its density-operator purity; these internal labels factor from the continuous delay.
These are probabilities before detection, and they add up to one. The splitter is described by a real unitary matrix, which fixes the relative phase on reflection. The calculation does not subtract classical intensities.
Every emitted trial is kept, including trials with no click. Each detector either clicks or stays silent and cannot count photons, and the electronic jitter is much wider than the optical pulse. Background adds an independent chance of a click in each gate. The table and the CSV use exactly this rule.
For each width on a grid, the baseline and dip depth are fitted first, and the count likelihood then picks the width. The model bandwidth is not used. The fit is switched off when background or two-pair events are present, and weak fits or fits at the grid edge are reported as unresolved. No confidence interval is given for the fit.
Typical misconceptionThe two particles in the picture must pick the same route before they are measured.
Better mental modelThe scene shows input modes and output photon numbers. No hidden path is ever stored as if it had been observed.
Typical misconceptionNarrowing the gate makes the photons indistinguishable.
Better mental modelThe gate only filters click times that already exist. It cannot bring back amplitudes removed by orthogonal polarization or a spectral mismatch.
Typical misconceptionFor any source, the depth of the dip directly gives its purity or its timing resolution.
Better mental modelFor identical independent inputs with all other degrees of freedom matched, the ideal balanced HOM visibility equals internal-state purity. Mismatch, background and extra pairs also lower the contrast. Correlated pair preparation and calibrated device response require a source model.
Choose Arrive together, predict the output, then acquire one trial at a time, several times.
What to observe: In the ideal balanced case, only one counter clicks. That click cannot tell you how many photons arrived at its port.Choose Arrive apart, then Orthogonal polarization, and compare a batch from each case.
What to observe: Coincidences come back. The photons must overlap in time and spectrum, and they must also share the same polarization.Start over and acquire a delay scan. Inspect the bins and export the raw CSV.
What to observe: The orange points are computed from those clicks. The Wilson ranges show the uncertainty of a finite number of yes-or-no counts, and the gray line is a separate ideal reference.Open the source and detector settings. Change the efficiency, gate width, background or two-pair fraction, then acquire again.
What to observe: Loss and a narrow gate lower the recorded counts. Background and four-photon input can fill in the dip. When the single-pair assumptions fail, the fit is switched off.Quantum evidence · Reading path
Articles on this path that use this experiment, followed by the other experiments they suggest.
Time-frequency quantum optics · Reading path
Articles on this path that use this experiment, followed by the other experiments they suggest.