Normalized electrode difference
Independent electrical and mechanical standards calibrate ideal linear-cut electrodes to a centroid. A separate charge ring supplies absolute bunch charge.
Why can a tiny spot hide a divergent beam? A screen measures position, so a small spot can still contain a wide range of particle angles. Changing a quadrupole reveals several position–angle projections. Calibrated screen widths then recover the projected phase-space area, provided resolution, clipping and a changing beam do not hide it.
Read four-electrode signals, independent bunch charge and screen pixels. Scan a finite quadrupole to recover projected emittance and Twiss parameters.
Charge ring and four downstream electrodes; signed finite-gate integration uses independent electrical references.
The interval includes a three-standard-uncertainty margin and deterministic pixel-bin bounds. A statistical ± value alone does not include that binning bound.
Eight records per session. CSV retains each complete calibration and raw record in a quoted JSON cell. Reset appends a new default run. Runs have independent virtual time origins; parent links express provenance, not a serial wall-clock schedule.
BackgroundA profile screen measures where particles arrive. Their directions determine how quickly that spot changes farther down the beamline.
Why it mattersA quadrupole changes which combination of entrance position and angle reaches the screen. Several projections supply information missing from one photograph.
Independent electrical and mechanical standards calibrate ideal linear-cut electrodes to a centroid. A separate charge ring supplies absolute bunch charge.
The matrix includes the finite quadrupole and drift. Each strength has a newly acquired image; held-out strengths are excluded from the fit.
The moments are centred on the measured beam. Statistical standard uncertainty and deterministic pixel-bin bounds are separate; unresolved area is withheld.
Typical misconceptionA smaller bright spot always means a better beam.
Better mental modelParticles can share a narrow waist and leave at very different angles. Emittance includes that directional spread.
Typical misconceptionMore photographs remove every calibration error.
Better mental modelA shared scale or reference affects the entire scan. Repetition does not average that common error away.
Compare the default screen images and width curves.
What to observe: One beam makes different spots as the field changes. Horizontal and vertical focusing act in opposite directions.Choose Same spot, different angles.
What to observe: One projection can agree even when the full scan curves separate. The missing information is angular spread.Drag the camera dial, then keep a scan.
What to observe: Independent reference images are acquired again with the unknown images. Their measured widths support the resolution correction.Drag the source or choose the lost-beam demonstration.
What to observe: Electrodes reveal the displacement. A mismatch between independent charge and screen light withholds full-beam area and preserves the pixels.You are hereBeam Diagnostics Bench
How physics instruments measure · Reading path
Articles on this path that use this experiment, followed by the other experiments they suggest.