Beam Diagnostics Bench

Why can a tiny spot hide a divergent beam? A screen measures position, so a small spot can still contain a wide range of particle angles. Changing a quadrupole reveals several position–angle projections. Calibrated screen widths then recover the projected phase-space area, provided resolution, clipping and a changing beam do not hide it.

Read four-electrode signals, independent bunch charge and screen pixels. Scan a finite quadrupole to recover projected emittance and Twiss parameters.

Interactive modelBeam Diagnostics Bench

From one spot to positions and angles

BackgroundA profile screen measures where particles arrive. Their directions determine how quickly that spot changes farther down the beamline.

Why it mattersA quadrupole changes which combination of entrance position and angle reaches the screen. Several projections supply information missing from one photograph.

Focus question
How can equally small spots hide different divergence?
One-sentence intuition
Projected phase-space area depends on position spread, angular spread and their correlation.

Core mathematical model

Normalized electrode difference

dx=AR−ALAR+ALd_x=\frac{A_R-A_L}{A_R+A_L}

Independent electrical and mechanical standards calibrate ideal linear-cut electrodes to a centroid. A separate charge ring supplies absolute bunch charge.

The screen is a transported projection

σs2=R112Σ11+2R11R12Σ12+R122Σ22\sigma_s^2=R_{11}^2\Sigma_{11}+2R_{11}R_{12}\Sigma_{12}+R_{12}^2\Sigma_{22}

The matrix includes the finite quadrupole and drift. Each strength has a newly acquired image; held-out strengths are excluded from the fit.

Projected RMS emittance

εrms=Σ11Σ22−Σ122\varepsilon_{\mathrm{rms}}=\sqrt{\Sigma_{11}\Sigma_{22}-\Sigma_{12}^2}

The moments are centred on the measured beam. Statistical standard uncertainty and deterministic pixel-bin bounds are separate; unresolved area is withheld.

Common difficulties

Waist and area

Typical misconceptionA smaller bright spot always means a better beam.

Better mental modelParticles can share a narrow waist and leave at very different angles. Emittance includes that directional spread.

Common reference errors

Typical misconceptionMore photographs remove every calibration error.

Better mental modelA shared scale or reference affects the entire scan. Repetition does not average that common error away.

Run the experiment

  1. 01

    Watch a scan

    Compare the default screen images and width curves.

    What to observe: One beam makes different spots as the field changes. Horizontal and vertical focusing act in opposite directions.
  2. 02

    Compare equal spots

    Choose Same spot, different angles.

    What to observe: One projection can agree even when the full scan curves separate. The missing information is angular spread.
  3. 03

    Blur the optics

    Drag the camera dial, then keep a scan.

    What to observe: Independent reference images are acquired again with the unknown images. Their measured widths support the resolution correction.
  4. 04

    Steer off the screen

    Drag the source or choose the lost-beam demonstration.

    What to observe: Electrodes reveal the displacement. A mismatch between independent charge and screen light withholds full-beam area and preserves the pixels.

Related experiments and articles

What this page links to directly

How physics instruments measure · Reading path

How physics instruments measure

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleThe directions hidden inside a small beam spot