Wave aberration with signed coefficients
Angles are in radians and every coefficient is a length; spherical aberration is entered in micrometres. The sign of positive defocus follows this formula.
E31 · Electron wavefront
Can a wider cone of electrons focus to a sharper probe? Only if the leftover aberrations are tuned small. A wider convergence angle can give a sharper probe, but the outer rays carry uncorrected wave aberrations that spread intensity out of the central peak. A Ronchigram that looks well tuned still does not prove the resolution you will get in an experiment.
Adjust the residual aberrations, each with its sign, record a Ronchigram, and compare it with the calculated probe at the specimen.
Physics tutorial
BackgroundA corrected electron lens still has small residual aberrations, and they decide how sharp the focused probe can be. Here you tune those signed residuals, record a Ronchigram and compare it with the calculated probe at the specimen.
Why it mattersA wider cone of electrons can focus to a smaller spot, but the residual aberrations grow with angle and can wash out the coherent peak.
Start with the essentials
Angles are in radians and every coefficient is a length; spherical aberration is entered in micrometres. The sign of positive defocus follows this formula.
A unitary discrete Fourier transform carries the normalized probe through one fixed, synthetic amorphous phase sheet.
This on-axis Strehl ratio compares the peak with that of the same aperture without aberrations. It is calculated, not inferred from the camera.
Typical misconceptionA low residual or sharp calculated probe certifies an experimental image.
Better mental modelThe residuals and the calculated probe describe only this model. This is a coherent 200 kV model with a thin phase screen as the specimen. It includes astigmatism and coma along fixed axes, signed defocus and third-order spherical aberration. The coefficients are residual aberrations, not corrector currents. Ronchigrams are simulated recorded diffraction counts. The probe intensity, aperture phase, Strehl ratio and encircled radius are calculated separately as references and are never reconstructed from the counts. A separate known-grating scan shares a fixed total electron count across five field positions and six frequencies. Its measured modulation and three-standard-error threshold define a sampled information limit and a usable field at 2.5 inverse nanometres. An assumed quadratic field-curvature law changes local focus; these are teaching calibration results, not microscope specifications. The model leaves out chromatic spread, source-size blur, higher-order aberrations, dose damage and material calibration. Angles and planes are magnified schematically.
Adjust residual aberrations and convergence. Predict how the probe changes.
What to observe: A small aperture can hide aberrations, but it makes the probe wider.Acquire a Ronchigram, then change a slider.
What to observe: Each lens change records fresh counts and references; changing display gain preserves them.Inspect the measured diffraction plane and the separately calculated specimen-plane probe.
What to observe: The aperture phase plot is a view into the model, not a second detector.Compare the corrected, wide-field and low-electron presets. Open the recorded field and dose test.
What to observe: The same three-standard-error rule tests every grating. A sharp reference probe can coexist with an undetectable modulation or a poor field edge.Check the target, then compare astigmatism, spherical aberration and low-count cases.
What to observe: Correcting the lens and collecting enough counts are two separate requirements.