Rectangular-pulse excitation
Detunings in the formula are frequencies in hertz; the scan displays kilohertz. Probe duration is a calibrated input.
SPECTROSCOPY ION / SHARED MOTION / LOGIC ION
Probe a spectroscopy ion, map its excitation through a shared motional quantum and count fluorescence from the auxiliary ion. Calibrate bright and dark records before inferring the spectral center.
The trap axis joins the two ions. The two ions and their separation are enlarged. Sideband beams address the ions in sequence; shared motion is not a photon traveling between them. The objective collects only auxiliary-ion fluorescence toward the counter. Sequence colors show source population expectations, not a measured trajectory.
SYNTHETIC OBSERVATIONS / REPRODUCIBLE RECORDS
Ready to acquire synthetic records.
A new run retains 64 known bright and 64 known dark preparations, followed by an interleaved 41-frequency scan. Each trial records detected photon count, not a hidden ion-state label. Source line shift, sideband pulse areas, delay and source photon means are excluded from the spectral estimator. Acquisition is accelerated.
RETAINED DATA
First 240 scan trials, excluding calibration. Green: current detected counts. Gray: exact saved counts. The inspection control selects one of these trials; the full run remains in CSV.
ESTIMATES FROM RECORDS
Fitted spectral center
Transferred population amplitude
Measured bright / dark mean photons
Calibrated false / true dark fractions
Inspected photon count
On-resonance source populations · spectroscopy / motion / logic
No line center is reported for incomplete acquisition, overlapping references, a weak transferred response or a fit at the search edge. The profile is a relative likelihood diagnostic, not a calibrated confidence interval.
DATA REDUCTION
Green points: classified dark fraction; brown line: response fitted from these counts with calibrated classification rates. No source center is drawn as a measurement.
Transfer amplitude is profiled at each candidate center. Relative deviance is capped at forty.
Columns: index, preparation, probe detuning in kilohertz, detected photons.
| Index | Preparation | Detuning | Photons |
|---|
Synthetic two-level ions and one initially cold transfer mode. Coherent red-sideband pulses with an illustrative phonon-loss channel; ideal confinement and preparation. No RF electrode solution, thermal modes, spontaneous decay, clock systematics or precision claim. The fluorescence spectrum determines transferred population, not coherence or transfer fidelity. Enlarged apparatus and pulse animation are schematic.
Rectangular-pulse excitation
Detunings in the formula are frequencies in hertz; the scan displays kilohertz. Probe duration is a calibrated input.
State mapping
Two resonant red-sideband pulses transfer excitation through the shared mode. The unexcited ground branch stays unchanged. The displayed source evolution uses a density matrix.
Population transfer
An illustrative zero-temperature amplitude-damping channel acts on the mode between pulses. This lifetime is a chosen parameter, not a measured apparatus value.
Calibrated observation model
Bright/dark reference counts determine the threshold and two classification rates. A binomial likelihood fits center and transfer amplitude from scan counts; source shift and pulse areas are excluded.
Physics tutorial
BackgroundProbe a spectroscopy ion, map its excitation through a shared motional quantum and count fluorescence from the auxiliary ion. Calibrate bright and dark records before inferring the spectral center.
Why it mattersA protocol connects a prepared quantum system to retained detector observations.
Start with the essentials
Detunings in the formula are frequencies in hertz; the scan displays kilohertz. Probe duration is a calibrated input.
Two resonant red-sideband pulses transfer excitation through the shared mode. The unexcited ground branch stays unchanged. The displayed source evolution uses a density matrix.
An illustrative zero-temperature amplitude-damping channel acts on the mode between pulses. This lifetime is a chosen parameter, not a measured apparatus value.
Bright/dark reference counts determine the threshold and two classification rates. A binomial likelihood fits center and transfer amplitude from scan counts; source shift and pulse areas are excluded.
Typical misconceptionRecovering a source parameter independently confirms the law.
Better mental modelThese synthetic records test a measurement and inference workflow. They are not historical observations, a parameter confidence interval, or external scientific certification.
Acquire a complete run; export the raw photon counts.
What to observe: The first 128 records alternate known bright and dark preparations; 41 probe frequencies follow.Save a comparison, then turn mapping off or add delay.
What to observe: The auxiliary fluorescence loses its spectral contrast even though the source probe excitation remains.Inspect the sequence stages and the fluorescence calibration.
What to observe: Stage populations are source predictions; the retained photon counts support population inference only. A Ramsey protocol would be needed to test transferred coherence.