IN41 · Timing an ion flight

Time-of-Flight Mass Spectrometer

How do arrival times separate an ion beam by mass-to-charge ratio? Ions with the same charge gain the same energy across a voltage drop. Lighter ions travel faster. Known reference ions calibrate the start delay and time scale before arrival times become mass-to-charge values. Charge state needs separate information.

Follow finite ion flights through straight and reflected paths. Use independent references to reconstruct mass-to-charge peaks and widths.

Interactive modelTime-of-Flight Mass Spectrometer

Detected mass-to-charge spectrum

Detected events00
Supported peaks00
First peak00
First-peak resolving power00

μ=(m/u)/z(Da/e)\mu=(m/u)/z\quad(\mathrm{Da}/e)

Detected counts are preserved. A single broad band can contain unresolved components; one mass-to-charge peak does not determine charge state.

Raw clocks, calibration and limits

Saved unknown arrival times

(tarrival−ttrigger)/μs(t_{\mathrm{arrival}}-t_{\mathrm{trigger}})/\mathrm{\mu s}

Fs=1.00 GHz,ηref=0.65F_s=1.00\,\mathrm{GHz},\quad \eta_{\mathrm{ref}}=0.65

ΔμFWHM\Delta\mu_{\mathrm{FWHM}}

Values after ± are standard uncertainties. Shared reference errors propagate together; the common clock scale cancels from mass-to-charge. Width uncertainty uses 256 resampled pulse blocks plus bin resolution. An unresolved mean clock-cell correction is retained instead of averaging it away. Approximate compatibility intervals are conditional on the retained band and stationary independent pulses, with no certified coverage claim.

Known mass standards fit time delay and scale; independent held-out standards test them. Blanks estimate background, and separate measured charge/count references check efficiency. A frozen calibration precedes all held-out and unknown events.

Each acquisition is a separate virtual experiment with its own counter origin and UTC anchor; comparison histories can overlap in simulated physical time. Whole records include every pulse gate and voltage readback, all reference and unknown event ticks, uncertainties and lineage. Unsupported conditions retain raw events and withhold mass and width results.

Physics tutorial

Time the whole ion flight

BackgroundElectric fields accelerate ions into a field-free tube. Equal charges gain similar energies, so light ions arrive before heavy ones.

Why it mattersSource positions, launch velocities and electronic delays broaden or shift the arrivals. Reference peaks establish the mass-to-charge scale.

Start with the essentials

Focus question
How can arrival times reveal an ion beam’s mass-to-charge ratios?
One-sentence intuition
Keep detected arrival times and calibrate them independently. Peak position gives mass-to-charge ratio; observed width measures resolving power.

Core mathematical model

The extraction gap takes time too

mv˙=qE,tflight=tgap+tdrift+tmirrorm\dot{\mathbf v}=q\mathbf E,\quad t_{\mathrm{flight}}=t_{\mathrm{gap}}+t_{\mathrm{drift}}+t_{\mathrm{mirror}}

Every physical segment contributes to the recorded time. Mirror operation includes deceleration, reversal and the return flight.

Reference ions establish the time scale

t=β+αμV,μ=m/uzt=\beta+\alpha\sqrt{\frac{\mu}{V}},\qquad \mu=\frac{m/u}{z}

Known reference ratios fit the delay and scale. Separate held-out references test that relation before unknown events use the frozen calibration and measured voltage.

Resolving power comes from the observed width

R=μpeakΔμFWHMR=\frac{\mu_{\mathrm{peak}}}{\Delta\mu_{\mathrm{FWHM}}}

Full width at half maximum is estimated from detected events. An unresolved broad band remains an unresolved result.

Common difficulties

Acceleration has a finite duration

Typical misconceptionTube length divided by final speed gives the entire flight time.

Better mental modelIons also cross an extraction gap. Starting position and velocity change both that time and the exit energy.

Detected counts and source abundance

Typical misconceptionPeak area directly counts all ions emitted by the source.

Better mental modelApertures, detection efficiency, gates and deadtime remove events. Separate charge references check efficiency; unknown incident counts are absent from the raw record.

Run the experiment

  1. 01

    Start with two peaks

    Select Two neighbouring peaks, then drag the gold voltage knob.

    What to observe: Ion speeds change before release. A new acquisition supplies new reference and arrival records.
  2. 02

    Give energetic ions a longer route

    Compare A wide starting region, Refocus with an ion mirror and An untuned mirror.

    What to observe: Faster ions penetrate deeper and spend longer in the mirror. A suitable field narrows arrival differences; mistuning broadens them.
  3. 03

    Interpret a single peak

    Select Different ions, same ratio and open the raw times and calibration.

    What to observe: Different masses and charge states can share one mass-to-charge ratio and one arrival peak.
  4. 04

    Keep a record and reset

    Acquire, export JSON or CSV, then reset the controls.

    What to observe: Reset appends a default acquisition. Existing events, standards and exports stay unchanged. An excessive beam retains raw events and withholds inference.

Related experiments and articles

How physics instruments measure · Reading path

How physics instruments measure

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleHow an ion flight becomes a mass spectrum