Wheatstone null
The left arms run from A through R1 and R2 to D; the right arms use R3 and Rx. At a true null their midpoint voltages agree, independently of finite detector resistance.
IN10 · A null reveals a resistance
Why can zero detector current reveal an unknown resistance? At a bridge null, the detector connects two points at equal potential. Known arm ratios then determine the unknown resistance. The detector need only resolve which way the imbalance points; the standards and actual contact locations still determine measurement accuracy.
Balance a finite-input bridge, move sense contacts and reverse excitation to measure resistance and temperature-corrected strain.
Simulated ADC observations
| Polarity | Detector code | Source monitor code |
|---|
The two measured source magnitudes need not match. Their slope removes a thermal voltage stable over the pair. Bridge inversion uses known arms and the finite detector; wire measurements report the segment between the actual clips, including any lead inside it. The even intercept is a circuit-loaded offset, not an independent thermometer. Noise, ADC quantization and standard errors remain; no total uncertainty or confidence interval is claimed.
The gauge fixture uses an ideal maintained current of five milliamperes in each direction. Its known input value, ampere unit and fixture version are saved. Fixture raw current codes are monitoring diagnostics; the gauge fit uses the saved known current. Main measurement pairs instead use their measured excitation difference.
Linear, ohmic, isothermal teaching components. Gauge resistance is multiplicative in strain and measured temperature, with factor and temperature coefficient fitted from independent fixture observations. Self-heating, lead/contact drift during reversal, leakage to earth and nonlinear contacts are omitted. A current marker is schematic; the board layout identifies the actual solved nodes.
Physics tutorial
BackgroundTwo divider midpoints can sit at the same voltage even while current flows in each arm. A finite-input detector compares those points.
Why it mattersA small resistance can be hidden by its leads, and a small strain can be hidden by temperature.
Start with the essentials
The left arms run from A through R1 and R2 to D; the right arms use R3 and Rx. At a true null their midpoint voltages agree, independently of finite detector resistance.
Main path A–B–C–D contains Rx, link r and standard S. The main ratio path runs A–E–D; the auxiliary path runs B–F–C. The detector compares E to F. Matching ratios cancels the link term.
The slope uses measured source currents, not assumed equal commands. A finite sense-input current is corrected. Leads within the chosen segment remain part of that measured resistance.
A known-current fixture measures zero strain, known strain and a second temperature. These observations identify the reference resistance, factor and temperature coefficient used by the inverse.
Typical misconceptionIf the detector current is zero, no current flows anywhere.
Better mental modelThe arms carry current. Only the comparison branch has equal potentials at its endpoints.
Drag the gold ratio dial, then save a reversed pair.
What to observe: The detector changes sign. Its input current vanishes at balance, while current still flows in the bridge arms.Choose Sense at the specimen, then drag each clip to an outer post.
What to observe: Every included lead adds its resistance to the measured segment. The force current and voltage contacts have distinct paths.Choose Two ratios disagree, then change the auxiliary mismatch in advanced settings.
What to observe: The complete double bridge changes its detector response; a main-ratio-only interpretation would miss the link term.Choose Stretch a warm gauge. Change temperature, acquire independent standards and save a new pair.
What to observe: Known strain/current/temperature fixture observations set the gauge conversion. Old data retain their original calibration.How physics instruments measure · Reading path