Counts per pixel and total scan time
The current is per active beam. Each batch adds an assumed overhead of one millisecond, and the time to record the reference is not included.
E22 · Electron acquisition
Can a multi-beam SEM scan faster and still stitch its tiles into a seamless image? Only if the beams are calibrated and each pixel still gets enough electrons. Every beam has its own gain and its own position error, so before stitching, even out each tile's brightness against a uniform reference and place it using calibrated beam positions. A faster scan leaves fewer electrons per pixel.
Record nine overlapping tiles and a uniform reference image, then correct each beam's gain and calibrated position before stitching the tiles together.
Physics tutorial
BackgroundA multi-beam SEM scans with several beams at once, and each beam records its own tile. Here you record nine overlapping tiles and a uniform reference, then correct each beam's gain and calibrated position before stitching.
Why it mattersMore beams finish the scan sooner, but each beam has its own response and position, so the seams between tiles can disagree.
Start with the essentials
The current is per active beam. Each batch adds an assumed overhead of one millisecond, and the time to record the reference is not included.
The uniform reference has a known probability of one quarter. The corrected signal is an estimate that can be fractional, not a new integer count.
Only pixels covered by more than one tile contribute. Bilinear resampling makes the noise correlated, so this number is not a confidence interval.
Typical misconceptionA small residual proves the physical experiment is fully resolved.
Better mental modelThe residual only shows agreement inside this simplified model. The model counts electrons in nine channels, using a synthetic specimen, normalized Gaussian probes and independent flat-field counts. Fiducial positions are assumed to be known exactly; stitching resamples the recorded pixels and never estimates positions from the true specimen. Channel response is an effective collection probability, not an analog detector gain. Switching between 1, 3 and 9 beams at fixed current per beam changes the scan time, not the dose. Stitching does not remove focus mismatch. Calibration time, stage motion, cross-talk, drift, charging and real multi-beam electron optics are left out. The task covers a single section, not an array-tomography reconstruction. The planes shown are schematic.
Change channel response and offsets. Predict which boundaries will disagree.
What to observe: The same specimen is measured through nine different channels.Acquire the tiles and a separate uniform flat field, then move a slider.
What to observe: The stored counts and calibration stay fixed until you acquire again.Stitch with and without the gain and position corrections, and inspect the coverage.
What to observe: Blank edges are pixels that were never measured; interpolation cannot add coverage.Compare the serial, blurred and low-count examples. Check the target.
What to observe: Parallel beams save time on the ideal schedule, but blur and dose still limit the data.