What they measured
The device is a straight channel in a silicon–glass chip, 377 µm wide and 157 µm high, with a piezo exciting a standing wave across the width. They filmed polystyrene particles 0.6 µm to 10 µm across in the horizontal symmetry plane at mid-height and used micro particle image velocimetry (micro-PIV) to get the velocity at each position.
There are four series: two in pure water at 1.940 MHz with the channel half a wavelength wide (the second repeats the first), one in pure water at 3.900 MHz with the channel one wavelength wide, and one in a 3 : 1 water–glycerol mixture at 2.027 MHz. Velocities taken at different voltages are rescaled to 1 V with the rule that velocity grows as voltage squared.
Source: R. Barnkob · M. Settnes and H. Bruus
The model they test
In the symmetry plane the horizontal particle velocity is the sum of the radiation drift and the streaming drag, and both are proportional to . So each particle size needs only one fitted amplitude . Theory says is a straight line in , with the intercept from streaming and the slope from radiation.
The paper corrects Rayleigh’s parallel-plate coefficient twice: by 0.82 for the rectangular cross-section and by 1.26 for thermoviscous effects, giving . The radiation term is also divided by a wall-drag factor , because particles near a wall feel extra drag.
Source: R. Barnkob · M. Settnes and H. Bruus
How the numbers support the conclusion
First, each of the four series shows a clear dependence (Fig. 8). The fitted energy density per volt differs by more than a factor of 700 between series, meaning the resonances were excited with very different strength, while the streaming coefficient falls between 0.18 and 0.25, with an uncertainty-weighted mean of , in line with the theoretical 0.194 (Table IV).
Second, once velocities are divided by and squared sizes by , 22 measurements from three series fall on one line (Fig. 9).
Third, the ratios of Table V. Treating 0.6 µm particles as moved by streaming alone and 10 µm particles by radiation alone, their velocity ratio is 49.6 at 1.940 MHz and 88.4 at 3.900 MHz: 1.8 times larger for 2.0 times the frequency. Water against water–glycerol gives about 3.0 (the text writes 48.6/16.4, the table 49.6), against 3.2 predicted from . And should be 0.25 and is measured at 0.27 to 0.29.
The paper lists its own sources of error: particle density and compressibility vary between makers and batches, size distributions are not exactly Gaussian, local concentration changes cause particle interactions, and the particles change the liquid viscosity. It also notes that real channel fields can be far less uniform than the one-dimensional standing wave used here, and that the streaming field still needs mapping over the whole cross-section and length.
Source: R. Barnkob
Worked example: recompute Table II and the critical diameter
From Table II: polystyrene and ; water and . Then , and , rounded to 0.17 in the table.
Table II gives as 5.25 mm²/s in water and 16.8 mm²/s in water–glycerol. With and : water at 1.940 MHz gives 1.42 µm, at 3.900 MHz 1.00 µm, and water–glycerol at 2.027 MHz 2.48 µm.
Now predict the first ratio of Table V. In water at 1.940 MHz, . With the measured diameters of 10.16 µm and 0.59 µm, Eq. 13 gives radiation-to-streaming ratios and . Under the paper’s assumption (large particles pure radiation, small ones pure streaming) the velocity ratio is 51.5. Keeping the other part of each particle’s motion, Eq. 12 gives . The measured 49.6 lies between the two. Neither prediction includes wall drag; this is our recomputation, not a number from the paper.
The acoustic trap experiment of this course recomputes the contrast factor and penetration depths of Table II, the numbers from Eqs. 5 and 14, and the wall factors of Table I. Its particle tracks are model output computed from the closed-form fields.
Source: R. Barnkob · M. Settnes and H. Bruus