The vacuum gap sets the current scale
At the same bias and on the same sample, increasing the gap by a fraction of a nanometer can cut the current sharply.
Why can an STM report a bump on a flat surface? At a fixed gap, extra electronic states can carry more tunnel current. In constant-current mode the tip lifts to bring the current back down, so a flat surface can appear raised. Two biases give different weights to those states and can help separate electronic contrast from height.
Scan a sample at constant height or constant current, then hold the tip for a local spectrum. Use two bias readings at each position to recover the surface profile.
Each scan point records two extra currents at and at the same tip position. The biases weight the sample resonance differently, so the two readings determine a height and an electronic contribution for this calibrated sample model.
The spectrum sweeps bias at the recorded lateral position and height, with feedback held. The feedback readout during a scan can lag behind either the geometric or the electronic change.
BackgroundA scanning tunneling microscope brings a conducting tip close to a conducting sample. The current depends on the gap and on which sample states can accept or supply electrons at the chosen voltage.
Why it mattersTwo regions of the same height can look different in a scan. A local spectrum gives a way to test where that contrast comes from.
At the same bias and on the same sample, increasing the gap by a fraction of a nanometer can cut the current sharply.
At fixed gap, positive sample voltage opens a window above the sample Fermi level. The thermal occupation changes smoothly at the edges.
Typical misconceptionEvery bump in a constant-current scan is a bump in the atoms.
Better mental modelFeedback also lifts the tip above a region that carries more current at the same gap. Two biases respond differently to the same local states and allow a model-based separation.
Choose “A bump on a flat surface”.
What to observe: The recorded tip height rises while the surface recovered from the bias pair stays flat.Switch to “Constant height”.
What to observe: The tip follows a horizontal line and a peak appears in the current at the same place.Drag the tip to the peak, then select “Local spectrum”.
What to observe: The tip stays at its recorded height while the bias changes. The conductance peaks when the bias is near the sample resonance.