Skip to main content
Sandbox Physics

Optics 065 · Diffraction, Fourier optics, and computational imaging

Phase Microscopy Lab

Can a transparent cell be seen without staining, and is the relief it seems to have real? It can. Zernike phase contrast, darkfield and DIC all turn the delay the cell gives the light into brightness differences, with no stain. The relief is not real height, though. DIC shading depends on direction, so the cell looks lit from one side, and phase contrast adds halos around edges.

Use a phase microscope with four imaging modes to make transparent patterns visible. See where halos and false relief appear, and follow the same complex specimen field all the way to the sensor.

Interactive modelPhase Microscopy Lab
Peak input phase / coherent cutoff ϕpeak,νc\phi_{\rm peak},\nu_c0.500.50
Raw mean / empty-specimen background ⟨I⟩,Iempty\langle I\rangle,I_{\rm empty}50%50\%
Raw intensity extrema Imin⁡,Imax⁡I_{\min},I_{\max}0.00π0.00\pi
Model regimevalid model regime\text{valid model regime}

Physics tutorial

Audit the physics of Phase Microscopy Lab

BackgroundA thin transparent specimen changes the phase of the light passing through it but not its local intensity: ϕ=2π OPD/λ,t=eiϕ\phi=2\pi\,\mathrm{OPD}/\lambda,\qquad t=e^{i\phi}. The specimen is a known, programmed map of optical path difference (OPD), not a photograph. The same specimen is sent through four different optical setups.

Why it mattersCan a transparent cell be seen without staining, and is the relief it seems to have real?

Start with the essentials

Focus question
Can a transparent cell be seen without staining, and is the relief it seems to have real?
One-sentence intuition
Light from mutually incoherent illumination directions adds in intensity, not in amplitude: I=∑jwj∣Uj∣2I=\sum_j w_j|U_j|^2. Brightfield and Zernike phase contrast share the same ring of sixteen source points. Darkfield moves the direct illumination outside the angles the objective accepts. DIC uses an ideal polarized pair of fields on the axis instead, so its illumination is not the same partially coherent source.

Core mathematical model

The annular phase filter in the pupil

P(ν)=1∣ν∣≤νc⋅[1+(Teiδ−1)1ring]νc=NA/λ\begin{aligned}P(\boldsymbol\nu)&=\mathbf1_{|\boldsymbol\nu|\le\nu_c}\\&\quad\cdot\left[1+\left(\sqrt{T}e^{i\delta}-1\right)\mathbf1_{\rm ring}\right]\\\nu_c&=\mathrm{NA}/\lambda\end{aligned}

The ring is centred at sixty-five percent of the coherent pupil radius and has a full width of fifteen percent. The transmission is an intensity transmission, so the amplitude uses its square root. The ring phase is set for the chosen wavelength; it is not a fixed manufactured coating.

Ideal DIC: two shifted fields interfere

U=F−1{PclearF[t]}IDIC=14∣eiβ/2U(r+s/2)−e−iβ/2U(r−s/2)∣2\begin{aligned}U&=\mathcal F^{-1}\{P_{\rm clear}\mathcal F[t]\}\\I_{\rm DIC}&=\frac14\left|e^{i\beta/2}U(\mathbf r+\mathbf s/2)\right.\\&\quad\left.-e^{-i\beta/2}U(\mathbf r-\mathbf s/2)\right|^2\end{aligned}

An equal split and an ideal crossed analyzer give the factor of one quarter relative to the total incident intensity. The shear direction is measured clockwise from the right of the image. The fields are shifted with Fourier phase ramps, not by rounding to whole pixels. The pupil filter acts before the two fields interfere.

Why the shadow is directional

IDIC/I0=sin⁡2 ⁣(β+Δϕ2)Δϕ≈s⋅∇ϕIempty/I0=sin⁡2(β/2)\begin{aligned}I_{\rm DIC}/I_0&=\sin^2\!\left(\frac{\beta+\Delta\phi}{2}\right)\\\Delta\phi&\approx\mathbf s\cdot\nabla\phi\\I_{\rm empty}/I_0&=\sin^2(\beta/2)\end{aligned}

This pure-phase limit assumes the objective passes the field that matters. The gradient approximation needs a shear that is small compared with the features. The full model keeps the exponential specimen field even when the phase is not weak, and at larger OPD the contrast follows no general linear law.

Darkfield blocks the direct light; gain only changes the display

∣νDF∣≈1.2νc>νcB=min⁡(1,GI/I0)∣t∣2=1\begin{aligned}|\boldsymbol\nu_{\rm DF}|&\approx1.2\nu_c>\nu_c\\B&=\min(1,G I/I_0)\\|t|^2&=1\end{aligned}

A discrete outer ring of illumination lights the same pure-phase object. The objective accepts no direct light, so an empty specimen gives zero intensity. The display gain scales all four images equally; it does not represent exposure physics, camera noise, automatic normalization or a phase reconstruction. The finite periodic cell and the sixteen source directions can cause discretization artifacts.

Common difficulties

Visible contrast is not a map of specimen height

Typical misconceptionEvery phase microscope shows true thickness; a brighter halo means more material.

Better mental modelZernike halos and shade-off appear because some diffracted light also passes through the phase ring. DIC responds to how the OPD changes along the shear direction, so rotating the shear changes the apparent shading. A uniform phase offset is invisible in every mode. No single intensity image is presented as a recovered phase.

Run the experiment

  1. 01

    Scene 1: Nearly invisible brightfield

    Start with brightfield, then switch to phase contrast without changing the specimen. Reverse the ring phase and compare the same raw centre cut.

    What to observe: The annular condenser aperture and the phase plate in the back focal plane sit in conjugate pupil planes, not in planes conjugate to the specimen image. Raising the NA changes the physical stop and its matching ring.
  2. 02

    Scene 2: Phase ring: contrast and halo

    Use the wide phase disk to see the halo and shade-off. Then make the ring fully transmitting with zero delay, and the phase-contrast image turns back into brightfield.

    What to observe: Darkfield records only light scattered into the objective. With no specimen, the direct illumination stops on an opaque region of the pupil and never passes through the absorbing detector.
  3. 03

    Scene 3: DIC: direction is not height

    Choose DIC and rotate the shear by a quarter turn. Then select darkfield and press the labeled exposure button. It changes only the display gain, not the raw intensity.

    What to observe: All recorded images share one fixed linear gain and clipping level. The OPD colour map has a fixed range of positive and negative values. Neither shows a measured biological specimen.

Related experiments and articles

Probes and measurement · Reading path

Diffraction, phase and surface structure

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleZernike: a reference wave makes transparent structure visible
  2. LabThick Lens DesignerFind where a thick lens measures its focal length from
FeedbackSomething broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy