Spontaneous Raman Spectrometer

Do molecular lines keep the same energy shift when the laser changes? Yes. A vibrational transition takes a fixed amount of photon energy, so changing the excitation moves the outgoing wavelength while preserving the Raman shift. Recovering it requires a calibrated spectral axis and enough resolution to separate nearby lines.

Calibrate a spectrometer, acquire weak scattered light and compare two excitation wavelengths. Separate a close pair of synthetic vibrational modes from fluorescence and counting noise.

Interactive modelSpontaneous Raman Spectrometer

What the CCD counted

Simulated measurements

Move a control or reset to start a fresh simulation.

Physical assumptions and sources

Two synthetic modes are broadened and integrated over physical wavelength-bin edges. A separate known reference estimates the wavelength offset. Background is estimated from edge bins, so broad backgrounds or unresolved overlap limit peak recovery. Intensities and widths are assumed; the model does not identify a real substance. Apparatus sizes and ray angles are schematic. Physical CCD columns use a fixed brightness tone map of the selected counts. Moving packets replay equal-mass quantiles of these recorded bins; their sizes and angles are display coding. Slit width changes only the prescribed instrumental response, at fixed throughput and intrinsic molecular width.

HORIBA · MacroRAM educational laboratory
NIST · spectral intensity calibration

Read a molecular energy difference

BackgroundA spectrograph records wavelengths. Molecular vibrations are compared by energy differences from the excitation light, so the laser wavelength belongs in the calibration.

Why it mattersA peak that moves when the laser changes can still represent the same vibration. Compare the stored spectra in wavelength and Raman-shift coordinates.

One-sentence intuition
A fixed vibrational energy gives a fixed reciprocal-wavelength difference, not a fixed wavelength difference.

Core mathematical model

Wavelength to Raman shift

Δν~=107(1λ0−1λs)  cm−1\Delta\widetilde\nu=10^7\left(\frac1{\lambda_0}-\frac1{\lambda_s}\right)\;\mathrm{cm}^{-1}

Both wavelengths are inserted in nanometres. Positive shift means the photon left energy in the sample.

Energy transferred to the vibration

ΔE=hc(100Δν~)\Delta E=hc(100\Delta\widetilde\nu)

The factor of one hundred converts inverse centimetres into inverse metres. The mode energy is held fixed when the excitation changes.

Independent photon counts

Ni∼Poisson⁡(μi),Var⁡Ni=μiN_i\sim\operatorname{Poisson}(\mu_i),\qquad\operatorname{Var}N_i=\mu_i

Each detector bin records its own signal and background photons. Subtracting an estimated background does not remove their counting noise.

Common difficulties

A smooth spectrum can hide two lines

Typical misconceptionEvery visible maximum is one molecular vibration, and smoothing can always recover a hidden pair.

Better mental modelFinite response blends nearby lines. The analysis requires two separated maxima with a resolved valley; otherwise it reports the pair as unresolved.

Run the experiment

  1. 01

    Calibrate before acquiring the sample

    Use Resolve the pair and Acquire calibration reference. Acquire the sample, then analyze stored samples.

    What to observe: The known reference estimates the wavelength offset. The sample analysis uses the stored counts and that frozen calibration.
  2. 02

    Change the laser

    Select Change excitation, calibrate the second laser, and acquire another sample. Switch the graph between wavelength and Raman shift.

    What to observe: The wavelength peaks move farther from the infrared source, but the recovered energy shifts agree within counting and calibration errors.
  3. 03

    Test resolution and background

    Acquire the Blurred doublet and Strong fluorescence presets.

    What to observe: Broadening can merge the pair. More background increases fluctuations even when its smooth trend is subtracted.

Related experiments and articles

What this page links to directly

Multidimensional spectroscopy · Reading path

Vibrations, echoes and frequency memory

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleWhat one infrared peak can tell us
  2. LabCoupled Vibrations in Two-Dimensional IRFind diagonal and cross features

Probes and measurement · Reading path

Molecules, forces and ultrafast dynamics

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleRaman: the shifted lines moved with the source
  2. LabRaman & Brillouin LasersCompare stimulated scattering
  3. LabVibrational EELS: Resolution & Aloof ProbesCompare electron energy-loss spectroscopy