IN20 · Thickness changes reflected polarization

Spectroscopic Ellipsometer

Can reflected polarization tell film thickness and refractive index apart? Several wavelengths and angles often help separate thickness from refractive index because each changes the reflected polarization differently. A single color can repeat the same response at several thicknesses. Matching the substrate index or strongly absorbing the returning light can leave thickness poorly determined even with a good fit.

Rotate two optical arms, keep intensity scans and reference calibration, then compare competing film fits across wavelengths and angles.

Interactive modelSpectroscopic Ellipsometer
One fitted thickness119.6 nm119.6\,\mathrm{nm}
Candidate index1.5001.500
Candidate extinction0.0140.014
Normalized ratio RMS5.72×10−45.72\times10^{-4}

Intensity scans determine the polarization change

Raw central intensity scan

A/∘;NADC/countsA/{}^\circ\quad;\quad N_{\mathrm{ADC}}/\mathrm{counts}

Turquoise: compensator at zero degrees; violet: forty-five degrees. The marker and physical display replay one stored exposure.

Choose incidence angle

Amplitude angle Ψ\Psi

λvac/nm;Ψ/∘\lambda_{\mathrm{vac}}/\mathrm{nm}\quad;\quad\Psi/{}^\circ

Phase difference Δ\Delta

λvac/nm;Δ/∘\lambda_{\mathrm{vac}}/\mathrm{nm}\quad;\quad\Delta/{}^\circ

Phase wraps leave a gap.

Filled marks come from raw reduction; dashes or open marks show one candidate, with each selected-angle panel scaled independently.

Competing films, reference checks and saved scans

Thickness profile at the candidate optical constants

d/nm;log⁡10Ld/\mathrm{nm}\quad;\quad\log_{10}L

L=1N∑j∣ρfit,j−ρ^j∣21+∣ρ^j∣2L=\frac1N\sum_j\frac{|\rho_{\mathrm{fit},j}-\widehat\rho_j|^2}{1+|\widehat\rho_j|^2}
Distinct numerical candidates
d/nmd/\mathrm{nm}nnκ\kappaCost

The profile holds candidate optical constants fixed; it is not a confidence interval or a full marginalization over correlated parameters. Several starting points reduce local-minimum risk but do not prove global uniqueness.

Frozen raw scans
RecordConditionsCalibrationFit

Exports retain all reference/dark/check and sample ADC counts, commanded angles, relative exposure times, seeds, conventions, model family and fitted candidates. Preview changes leave kept raw and processing snapshots unchanged.

Physics tutorial

Read the phase before choosing a film

BackgroundA thin film sends back light from its upper and lower boundaries. Their relative timing changes the polarization of the reflected beam.

Why it mattersTwo films can give nearly the same brightness yet change the reflected electric-field ellipse differently.

Start with the essentials

Focus question
When do several colors and angles distinguish thickness from refractive index?
One-sentence intuition
A measured polarization change constrains an optical model; several wavelengths can separate film candidates that coincide at one color.

Core mathematical model

The complex ellipsometric ratio

ρ=rprs=tan⁡Ψ eiΔ\rho=\frac{r_p}{r_s}=\tan\Psi\,e^{\mathrm i\Delta}

The reflection amplitudes use right-handed local p, s and propagation axes. Psi specifies their amplitude ratio and Delta their phase difference. A detector observes intensities at known compensator and analyzer angles.

One transparent-film round trip

β=2πdλvacN2−sin⁡2θ,N=n+iκ\beta=\frac{2\pi d}{\lambda_{\mathrm{vac}}}\sqrt{N^2-\sin^2\theta},\qquad N=n+\mathrm i\kappa

The phase convention is spatial phase minus time phase. A round trip contributes twice beta. For a transparent film, adding a whole round-trip cycle repeats the single-color reflection. Absorption attenuates the returning field.

Common difficulties

A small residual can have several solutions

Typical misconceptionThe lowest displayed cost proves a unique thickness.

Better mental modelThe film model may contain several similar minima or a broad weak-sensitivity region. The numerical comparison tolerance is not a confidence interval.

A linear analyzer can lose part of the phase

Typical misconceptionRotating a single polarizer always gives the full polarization ellipse.

Better mental modelThe compensator converts circular sensitivity into a measurable intensity change. Parking it can make the reference matrix rank deficient.

Run the experiment

  1. 01

    Watch one exposure

    Pause the replay and compare the physical analyzer display with the marker on the raw scan.

    What to observe: Both refer to the same stored detector count and commanded analyzer angle.
  2. 02

    Compare one color with several

    Choose the repeated-thickness preset, then restore several wavelengths and angles.

    What to observe: Several thicknesses can fit a single color. Additional conditions usually separate those branches.
  3. 03

    Change the receiving optics

    Drag the compensator or analyzer mount to change its zero offset, then measure new reference states.

    What to observe: A hardware change invalidates the old calibration. Six known input states calibrate the receiver; a different ellipse checks it.
  4. 04

    Remove the lower reflection

    Compare the index-matched and strongly absorbing presets.

    What to observe: A film can become insensitive to thickness because the lower interface vanishes or its returning light becomes too weak.

Related experiments and articles

How physics instruments measure · Reading path

How physics instruments measure

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleHow an ellipsometer reads thin-film thickness from reflected light
FeedbackSomething broken, confusing, or missing on this page?Tell us