IN18 · Many colors, one detector

Fourier Transform Infrared Spectrometer

How can a detector that sees total brightness recover an entire spectrum? Moving the mirror changes how far the two light paths differ. Each infrared wavelength then contributes its own bright-dark rhythm to the total detector signal. A Fourier transform separates these rhythms into a spectrum. Comparing sample and reference spectra reveals absorption where the reference supplies enough light.

Move a Michelson mirror and recover infrared absorption from saved reference and sample detector counts.

Interactive modelFourier Transform Infrared Spectrometer

Saved scans retain their positions and calibration

The two raw interferograms

xencoder/mm;C/103x_{\mathrm{encoder}}/\mathrm{mm}\quad;\quad C/10^3

Turquoise is the reference; violet is the sample. The moving marker replays a saved point only when the apparatus matches this record.

Spectra recovered from these counts

σ/cm−1;Sσ/[10−6 W/(cm−1)]\sigma/\mathrm{cm}^{-1}\quad;\quad S_\sigma/[10^{-6}\,\mathrm W/(\mathrm{cm}^{-1})]

Absorbance after background division

σ/cm−1;A=−log⁡10T\sigma/\mathrm{cm}^{-1}\quad;\quad A=-\log_{10}T

Violet is recovered from the raw pair. Dashed grey is the separate ideal sample; finite resolution can change the measured band depth. Unresolved bins are omitted.

Calibration, checks and retained pairs

Saved reference/sample pairs
PairPointsCalibrationWindowValid

Exports include both raw count arrays, every encoder position, independent dark/span and line observations, the withheld check, processing, model version and seed. Ideal source and sample data are explicitly separate.

Physics tutorial

The detector sees a sum of fringe rhythms

BackgroundA beam splitter sends infrared light down two paths. A moving mirror changes the return distance in one arm. Each wavelength alternately reinforces and cancels at a different displacement rhythm.

Why it mattersA broadband detector records total brightness, yet a position-tagged scan can retain enough information to recover a spectrum.

Start with the essentials

Focus question
How can a detector that sees total brightness recover an entire spectrum?
One-sentence intuition
Different wavelengths acquire phase at different rates as the mirror moves.

Core mathematical model

One bright output after a common-path sample

P+(δ)=12∫Sσ(σ)T(σ)[1+cos⁡(2πσδ)] dσ,δ=2xP_+(\delta)=\frac12\int S_\sigma(\sigma)T(\sigma)\left[1+\cos(2\pi\sigma\delta)\right]\,\mathrm d\sigma,\qquad\delta=2x

Wavenumber is in inverse centimetres and optical path difference in centimetres. The mirror travels half the optical path difference. The sample follows recombination, so its intensity transmission multiplies the whole output at each frequency. Different optical frequencies add powers.

Divide two recovered spectra where background is resolved

T^(σ)=S^sample(σ)S^reference(σ),A^=−log⁡10T^\widehat T(\sigma)=\frac{\widehat S_{\mathrm{sample}}(\sigma)}{\widehat S_{\mathrm{reference}}(\sigma)},\qquad\widehat A=-\log_{10}\widehat T

Both spectra are phase-corrected real Fourier transforms of saved counts, with the same finite window. Their ratio is an instrument-broadened estimate; it need not equal the ideal local sample transmission exactly. Weak-reference, alias-risk and saturated data cannot support absorbance.

Sample spacing and maximum path difference do different jobs

σNyq=12h,Δσbox=12D\sigma_{\mathrm{Nyq}}=\frac{1}{2h},\qquad\Delta\sigma_{\mathrm{box}}=\frac{1}{2D}

The calibrated optical path step sets Nyquist. The nominal half range is DD; the box-window first zero is the stated resolution criterion. A triangular window doubles that first-zero width. Zero padding adds displayed bins without narrowing it.

Common difficulties

A broad absorption band needs broad illumination

Typical misconceptionTwo isolated source lines reveal the whole sample band.

Better mental modelOnly illuminated wavenumbers carry sample information. The default adds a continuum to the two narrow source lines.

An interferogram is already an intensity measurement

Typical misconceptionSquare its Fourier transform to obtain the measured light spectrum.

Better mental modelThe positive-frequency transform is proportional to spectral intensity. Squaring it again changes peak ratios and units.

Run the experiment

  1. 01

    Inspect an actual mirror position

    Drag the scanning mirror and watch the detector count and current interferogram point.

    What to observe: The physical mirror and double-pass path share the selected encoder position. Motion replays the preview scan; it does not rewrite a saved scan.
  2. 02

    Insert the common-path sample

    Drag the sample holder out of the recombined output, then back in.

    What to observe: Both arm fields have already recombined before they enter the sample. Its band reduces only the corresponding spectral contributions.
  3. 03

    Compare the two scan lengths

    Compare Two colors beat together and A short scan blurs the pair.

    What to observe: Shortening maximum optical path difference broadens the instrument line shape. More zero padding leaves that line shape unchanged.
  4. 04

    Retain before changing processing

    Keep this pair, change the window in the collapsed group and export the saved records.

    What to observe: Old reference/sample counts, encoder coordinates and calibration remain fixed. The preview uses a new processing choice until explicitly kept.

Related experiments and articles

How physics instruments measure · Reading path

How physics instruments measure

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleHow total brightness reveals many infrared colors
FeedbackSomething broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy