IN17 · Give overlapping orders another direction

Cross-dispersed Echelle Spectrograph

How can high diffraction orders separate fine colors without piling them onto the same pixels? High orders spread nearby wavelengths farther apart, but different orders can send different colors in the same direction. A second disperser separates them in the perpendicular direction. The camera then records separate spectral stripes, each calibrated with known wavelengths.

Rotate an echelle grating, separate two orders across a real camera plane, and recover wavelength from stored detector pixels and independent lamp frames.

Interactive modelCross-dispersed Echelle Spectrograph
Order 111 largest bin548.55 nm548.55\,\mathrm{nm}
Order 110 largest bin553.56 nm553.56\,\mathrm{nm}
Minimum trace spacing41.0 px41.0\,\mathrm{px}
Largest withheld-line error0.000125 nm0.000125\,\mathrm{nm}

Pixels first, wavelengths after calibration

Raw camera observation

c/px;r/px(ADU)c/\mathrm{px}\quad;\quad r/\mathrm{px}\quad(\mathrm{ADU})

Log display above a 50-ADU display pedestal; row zero is at the bottom. Colored outlines show the actual extraction apertures. Drag vertically to change their half-width.

Two extracted spectra

λvac/nm;Cflat/ADU\lambda_{\mathrm{vac}}/\mathrm{nm}\quad;\quad C_{\mathrm{flat}}/\mathrm{ADU}

Turquoise is order 111; amber is order 110. Largest-bin labels do not split unresolved source lines. Invalid fits leave only the raw camera visible.

Lamp frames, checks and saved observations

λ^(c)=a0+a1ξ+a2ξ2,ξ=c−63.564\widehat\lambda(c)=a_0+a_1\xi+a_2\xi^2,\qquad\xi=\frac{c-63.5}{64}

Four line centroids per order determine the quadratic. The fifth wavelength is checked only afterwards. Exports retain all six raw frames, axes, seeds, standard definitions, pixel windows and processing results.

Frozen observations
ObservationCalibrationApertureValid

Physics tutorial

One camera direction is not enough for overlapping orders

BackgroundA ruled reflective surface sends light into several bright directions. At a high diffraction order, two nearby wavelengths leave at more distinct angles. Another order can send a different wavelength along that same path.

Why it mattersA camera needs a second way to tell those coincident colors apart before it can measure a narrow spectral line.

Start with the essentials

Focus question
How can high diffraction orders separate fine colors without piling them onto the same pixels?
One-sentence intuition
A perpendicular disperser turns coincident orders into separate camera stripes.

Core mathematical model

Directions of a reflective grating

mλ=d(sin⁡α+sin⁡β),d=(31.6 mm−1)−1m\lambda=d(\sin\alpha+\sin\beta),\qquad d=(31.6\,\mathrm{mm}^{-1})^{-1}

Angles are measured from the outward grating normal, with incidence and reflection on opposite tangent sides. Wavelength is vacuum wavelength; use the same length unit for pitch and wavelength. Order is an integer. The scene uses a fixed 80-degree incident beam and a small echelle rotation; the two arms have a separate 8-degree narrow-beam mount clearance.

Perpendicular wavevector change

uout=uin,vout=vin+λnc,wout=1−uout2−vout2u_{\mathrm{out}}=u_{\mathrm{in}},\qquad v_{\mathrm{out}}=v_{\mathrm{in}}+\lambda n_c,\qquad w_{\mathrm{out}}=\sqrt{1-u_{\mathrm{out}}^2-v_{\mathrm{out}}^2}

The thin transmission grating has grooves along its U direction and adds tangential wavevector along V. Direction cosines share its orthogonal local frame; wavelength in millimetres times rulings per millimetre is dimensionless. The positive root transmits forward. A linked camera mount follows the fixed central reference wavelength as ruling density changes.

Extract from corrected camera pixels

Cm(c)=∑r∈Am(c)[D(r,c)−B(r,c)F(r,c)−bm(c)]C_m(c)=\sum_{r\in\mathcal A_m(c)}\left[\frac{D(r,c)-B(r,c)}{F(r,c)}-b_m(c)\right]

Raw ADU minus a saved equal-exposure dark is divided by the normalized uniform detector flat. Actual row windows follow quadratic traces fitted from continuum pixels. Median side rows estimate background. The sum has flat-corrected ADU units and retains finite-aperture losses. Four known line centroids per order fit wavelength; a fifth line checks it independently.

Common difficulties

The camera is not a wavelength ruler yet

Typical misconceptionA bright pixel already comes with its wavelength.

Better mental modelPixel coordinates become wavelengths only after known-line calibration and an independent line check.

A flat frame corrects detector response

Typical misconceptionDetector flat correction makes the spectrum an absolute photon flux.

Better mental modelThe independent uniform flat removes relative pixel response. Blaze, aperture throughput and other optical losses remain; the extracted spectrum is not absolutely flux calibrated.

Run the experiment

  1. 01

    Follow the two arms

    Open Separated arms, then drag the gold echelle slowly.

    What to observe: The incident and reflected beams are distinct. Camera points shift with the same ray equations; changed optics require fresh lamp frames.
  2. 02

    Bring the stripes together

    Compare Two clean orders, Orders move closer and Orders can no longer be extracted.

    What to observe: Reducing cross dispersion decreases camera-row separation. Once the continuum no longer resolves two traces, extraction stops.
  3. 03

    Blur within an order

    Compare A wide slit blurs colors and A pair merges within one order.

    What to observe: Separating orders does not remove the slit and diffraction width within each stripe. The processing never inserts hidden source-line identities.
  4. 04

    Change processing after saving

    Keep this observation, then drag vertically on the raw image to change the aperture.

    What to observe: The preview refits and extracts the same pixels using a new window. Saved pixels, lamp frames and processing remain unchanged in CSV and JSON.

Related experiments and articles

How physics instruments measure · Reading path

How physics instruments measure

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleHow an echelle spectrograph gives overlapping colors a second direction