IN22 · Time individual photons

TCSPC Fluorescence Lifetime Instrument

How can a decay curve separate fluorescence lifetime from detector timing response? Measure the timing spread of a prompt reference, then convolve that measured response with possible fluorescence decays and compare them with photon arrival times. Keep counts low: recording the first photon in a crowded cycle favors early arrivals, and detector dead time removes further events.

Time individual photons, measure a prompt response and background, then compare low-rate lifetime fits with first-photon pile-up and dead time.

Interactive modelTCSPC Fluorescence Lifetime Instrument

Each photon contributes one observed delay

Observed delays and measured prompt response

tphase/ns;log⁡10(1+Ntags)t_{\mathrm{phase}}/\mathrm{ns}\quad;\quad \log_{10}(1+N_{\mathrm{tags}})

Green marks sample counts, violet the background-subtracted prompt normalized to signal area, and gold dashes the low-rate candidate.

Empty cycles and detector busy time

Fit profile, timing convention and saved records

Counts and busy intervals are retained with the full number of reference cycles. Losing late photons does not turn those missing cycles into measured fast decays.

τ/ns;log⁡10(1+D−Dmin⁡)\tau/\mathrm{ns}\quad;\quad \log_{10}(1+D-D_{\min})

The periodic reference clock starts at zero. Each saved pulse ID names the immediately preceding reference pulse; old excitations may contribute in later cycles. Time tags are integer ten-picosecond ticks.

Frozen records
RecordPhotonsCalibrationInference

CSV retains every sample, prompt and background pulse ID and absolute timestamp, clock descriptors, hardware counters and processing version. The separate generated reference is labelled in JSON and excluded from the inverse.

Physics tutorial

Calibrate the clock spread before measuring the wait

BackgroundAfter a short excitation pulse, different emitters wait different times before emitting. The detector and electronics add their own timing spread.

Why it mattersA broad arrival-time peak can come from the specimen, the instrument, or both.

Start with the essentials

Focus question
Which part of a measured delay distribution belongs to the specimen lifetime?
One-sentence intuition
Measure a prompt response independently, then compare its periodic convolution with possible fluorescence decays against saved photon timestamps.

Core mathematical model

A single exponential lifetime

p(t)=1τe−t/τ,t≥0p(t)=\frac1\tau e^{-t/\tau},\qquad t\ge0

Lifetime is a mean waiting time, not a fixed delay for each photon. The selected model has one lifetime and ideal independent emitters.

A measured timing response broadens a decay

h(t)=∫−∞∞r(u) p(t−u) duh(t)=\int_{-\infty}^{\infty}r(u)\,p(t-u)\,\mathrm du

The measured prompt response r is combined with each possible lifetime. Repeated excitation adds contributions from earlier cycles; the fit uses the periodic version of this convolution.

The first detected photon is selected unevenly

pfirst(t)=λ(t)exp⁡ ⁣[−∫0tλ(u) du]p_{\mathrm{first}}(t)=\lambda(t)\exp\!\left[-\int_0^t\lambda(u)\,\mathrm du\right]

This unconditional density includes the probability that no earlier photon arrived. Its missing area represents empty cycles. A crowded cycle favors earlier photons before any detector dead-time losses are added.

Common difficulties

A wide response is not a long lifetime

Typical misconceptionThe width of the arrival-time peak directly equals the lifetime.

Better mental modelThe peak includes source width and detector/electronic response. A prompt reference measures these contributions.

More counts can make a fitted decay wrong

Typical misconceptionA brighter sample always gives a more accurate lifetime.

Better mental modelFirst-photon selection favors early arrivals and detector busy time hides other photons. The low-rate fit is withheld outside its stated limits.

Run the experiment

  1. 01

    Watch a photon journey

    Pause the selected journey and identify excitation, waiting, fluorescence collection and electrical timing.

    What to observe: The source, specimen and collector form one causal path. Empty gaps are skipped and each leg is slowed separately.
  2. 02

    Broaden the instrument response

    Choose the fast-decay preset and inspect the independently measured prompt curve.

    What to observe: A broad prompt response can conceal a much shorter specimen decay. The declared resolution criterion withholds that short lifetime.
  3. 03

    Cross the next clock pulse

    Choose the folded preset and compare the observed pulse IDs with the sample delay histogram.

    What to observe: A photon can be registered after a later reference pulse. The inferred model adds earlier-cycle fluorescence.
  4. 04

    Crowd the detector

    Compare the first-photon and detector-busy presets.

    What to observe: A per-cycle gate rejects later detector pulses. Nonparalyzable detector dead time removes arrivals across cycles and acts first.
  5. 05

    Change the actual cable

    Drag the gold cable handle and then measure fresh references.

    What to observe: The cable centreline length changes the electrical delay. Old saved records retain their own matching timing calibration.

Related experiments and articles

How physics instruments measure · Reading path

How physics instruments measure

Articles on this path that use this experiment, followed by the other experiments they suggest.

  1. ArticleHow individual photons reveal a fluorescence lifetime