Skip to main content
Sandbox Physics

E52 · Photoelectron spectroscopy

UPS: Valence Bands & Work Function

Apply a negative bias, acquire a UV photoelectron spectrum and an independent metallic reference, then position the fitting windows. Recover the work function and occupied-band onset while testing a missing Fermi edge, an inaccessible cutoff and low counts.

Interactive modelUPS: Valence Bands & Work Function
Work function from measured edges—\text{—}
Fitted secondary-electron cutoff—\text{—}
Fitted Fermi reference—\text{—}
Measured cutoff-to-Fermi energy span—\text{—}
Measured occupied-edge binding offset—\text{—}
Measured ionization energy—\text{—}
Instrument Gaussian-equivalent FWHM—\text{—}
Applied signed sample bias—\text{—}
Recorded sample counts—\text{—}
Independent reference counts—\text{—}
Fermi-zero source—\text{—}
Experiment target—\text{—}

Physics tutorial

Measure a span, not an absolute cutoff

BackgroundUV photons expose occupied states and produce low-energy secondary electrons. The Fermi edge and secondary cutoff delimit an energy span set by photon energy and sample work function.

Why it mattersA low-energy cutoff can fall outside the analyzer acceptance. An occupied semiconductor band edge can also be mistaken for the Fermi level.

Start with the essentials

Focus question
Which two edges measure a work function?
One-sentence intuition
Negative bias moves both edges upward by the same energy. Their difference cancels bias and analyzer work function. A separate metallic reference supplies the Fermi zero when the sample has no Fermi edge.

Core mathematical model

Analyzer-side energies under negative sample bias

KF=hν−ϕa+e∣Vs∣Kc=ϕs−ϕa+e∣Vs∣\begin{gathered}K_F=h\nu-\phi_a+e|V_s|\\ K_c=\phi_s-\phi_a+e|V_s|\end{gathered}

Energies are measured in the analyzer vacuum. Sample and reference share the applied bias. The same bias offset enters the Fermi edge and cutoff.

Work function from a measured span

ϕs=hν−(KF−Kc)\phi_s=h\nu-(K_F-K_c)

The measured difference removes analyzer work function and bias. He I photon energy is fixed at 21.218 eV. Both edges must be inside the accepted energy range and correctly identified.

Occupied-edge binding offset and ionization energy

EV=KF−KVI=ϕs+EV\begin{gathered}E_V=K_F-K_V\\ I=\phi_s+E_V\end{gathered}

The semiconductor-like occupied edge is below Fermi. Its offset comes from sample counts referenced to the separately measured metal edge. In a metal this offset approaches zero.

Acquisition and processing

Di∼Poisson(μi),EB,i=KF,fit−KiD_i\sim\mathrm{Poisson}(\mu_i),\qquad E_{B,i}=K_{F,\mathrm{fit}}-K_i

The energy conversion reuses acquired counts. Expected rates and display gain are separate from measured spectra; changing fitting windows does not collect another scan.

Assumed Gaussian thermal surrogate

ΔEF,thermal≃3.53kBT\Delta E_{F,\mathrm{thermal}}\simeq3.53k_BT

This matches the Fermi derivative FWHM with a Gaussian surrogate. It illustrates thermal broadening without claiming the exact Fermi–Dirac line shape.

Common difficulties

Absolute cutoff is not enough

Typical misconceptionThe cutoff energy alone equals sample work function.

Better mental modelAnalyzer vacuum offset and sample bias enter the absolute cutoff. Measure the Fermi-to-cutoff span.

The first occupied state is not Fermi

Typical misconceptionAny high-energy edge is a Fermi edge.

Better mental modelA semiconductor-like sample has an occupied-band offset. Using that edge as Fermi overestimates work function by the offset.

Valence spectra are weighted

Typical misconceptionThe displayed valence curve is the exact density of states.

Better mental modelPhotoemission weights and response influence the intensity. This Lab uses invented valence peaks and performs no DOS inversion.

Run the experiment

  1. 01

    Recover detector access

    Start without bias, then select Measure a metal.

    What to observe: Negative bias lifts the sample cutoff above the 1 eV detector threshold; both edges move together.
  2. 02

    Position the fitting windows

    Move cutoff and Fermi windows across their visible measured edges; compare Wrong cutoff window.

    What to observe: The fitted edge needs counts and two-sided support. A smooth continuum is not an accepted cutoff.
  3. 03

    Separate Fermi and occupied-band edges

    Use Occupied edge below Fermi, then switch off the metal reference.

    What to observe: The sample lacks a Fermi edge. Restoring the independent reference recovers work function and ionization energy.
  4. 04

    Complete and stress the target

    Use sufficient counts and narrow instrument response, then check. Change bias and adjust the windows to follow the edges.

    What to observe: The edge positions change while the recovered work function stays near the separate known value.