Snell’s law with a negative index
Energy crosses the normal and leaves on the same side it came from. Inside the slab the phase moves back toward the source while the energy moves away from it.
Optics 098 · Near-field imaging
How does a negative-index slab bend light the wrong way and recover fine detail that normally fades out? In a negative-index slab, Snell's law puts the refracted ray on the same side of the normal as the incident ray, and the phase runs backward relative to the flow of energy. An ideal slab also amplifies evanescent waves, which carry the finest detail and normally die out before reaching an image. Real materials absorb, and that loss limits how sharp such a superlens can be.
Put a negative-index slab between two slits and a near-field probe. Watch rays bend the wrong way, then see two slits closer than a wavelength resolved, or blurred by loss.
Physics tutorial
BackgroundLight leaving a fine object carries its detail in waves that change quickly across the object. Those waves cannot travel: they fade within a fraction of a wavelength, so an ordinary lens never receives them and cannot image anything finer than about half a wavelength.
Why it mattersA slab whose permittivity and permeability are both minus one bends light the wrong way and, in theory, makes the fading waves grow again. The probe behind the slab records whether two slits 0.2 wavelengths apart arrive as two peaks or as one.
Start with the essentials
Energy crosses the normal and leaves on the same side it came from. Inside the slab the phase moves back toward the source while the energy moves away from it.
A transverse wavenumber larger than the vacuum wavenumber has no real . Its field falls off with distance, faster the finer the detail.
Here is the normal wavenumber inside the slab of thickness , and the formula includes every reflection between the faces. For light polarized with its magnetic field along the slits, compares the tangential electric fields inside and outside.
The slab undoes the decay of both gaps. With the mask a distance in front, the image sits behind the slab.
With the transfer at the image plane falls to one half where . For a loss in the permeability alone, Smith and colleagues (2003) give the same leading form, .
A path that passes every wavenumber up to some limit and nothing beyond has its edge at that limit. Narrow resonance peaks count no more than a passband of the same width.
With and the energy direction turns negative while the phase still runs forward. Energy bends the wrong way, yet fading waves are not amplified.
Typical misconceptionAny slab that bends light the wrong way will image beyond the diffraction limit.
Better mental modelBending propagating rays only refocuses waves that already travel. Restoring fine detail needs the fading waves to grow, which the indefinite slab here does not do.
Typical misconceptionIf the field grows inside the slab, the slab must be adding energy.
Better mental modelA fading wave carries no energy away from the object, so its field can grow without any energy gain, as Pendry (2000) noted. The energy flow of the propagating waves is conserved as usual.
Typical misconceptionA thicker negative-index slab restores more detail.
Better mental modelThe growth needed rises exponentially with thickness while the loss cap does not, so the finest restored wavenumber falls roughly as one over the thickness.
Keep the default preset, Near-perfect lens: two slits resolved, and watch the probe sweep.
What to observe: The probe scan shows two peaks with a dip of about 97 % between them. With vacuum in place of the slab, the same probe sees a single peak. The passband edge reads about six times the vacuum wavenumber.Load Glass slab: the detail fades.
What to observe: The rays bend toward the normal instead of across it, the passband edge drops to about 1.3 and the two slits merge into one peak.From the default, drag the material loss up to 0.01, then to 0.1.
What to observe: At 0.01 the dip is still about 85 % and the edge about 4.3. At 0.1 the edge falls to about 2.4 and the two peaks merge, which is the Lossy slab preset.From the default, drag the probe stage toward the slab and then away from it, or load Probe past the image plane.
What to observe: The dip is largest with the probe about 0.1 wavelengths behind the slab, where the image plane lies, and vanishes by 0.3 wavelengths.Load Negative refraction, forward phase.
What to observe: The energy rays bend the wrong way, by about 24 degrees, while the phase dots move forward. The passband edge stays near 1.4 and the slits still merge.Electron probes · Reading path
Articles on this path that use this experiment, followed by the other experiments they suggest.