Skip to main content
Sandbox Physics

E13 · Field emission and beam deceleration

Low-energy SEM: Slow at the Surface

Can electrons travel fast in the column and land gently? Yes. A negative voltage on the specimen slows the electrons just before they land. They travel down the column at high energy and land with low energy, so the interaction stays shallow, close to the surface.

Run the electron column at high energy, then slow the beam down with a negative bias on the specimen. Try to recover a thin-film pattern by trading off depth sensitivity, probe width, counting noise and the assumed charging.

Interactive modelLow-energy SEM: Slow at the Surface
Actual landing energy—\text{—}
Prescribed film depth weight—\text{—}
Modulation to representative pixel noise—\text{—}
Prescribed negative charging magnitude—\text{—}
More measurements and target
Selected stored detector counts—\text{—}
Count-derived film harmonic amplitude—\text{—}
Amplitude error against separate film—\text{—}
Prescribed frame displacement span—\text{—}
Combined Gaussian response FWHM—\text{—}
Incident dose per pixel—\text{—}
Experiment target—\text{—}

Physics tutorial

Low-energy SEM: Slow at the Surface

BackgroundAn SEM can run its column at high energy and still land electrons gently by putting a negative bias on the specimen. Here you try to recover a thin-film pattern while trading off depth sensitivity, probe width, counting noise and the assumed charging.

Why it mattersThe task is to recover the film's harmonic amplitude to within 0.02 of the separate input. Along the way, keep the contrast-to-noise ratio at a representative pixel at 8 or more, the response FWHM at 8 nm or less, the film depth weight at 90 percent or more, and the assumed charging at 20 V or less.

Start with the essentials

Focus question
Can electrons travel fast in the column and land gently?
One-sentence intuition
The landing energy decides how deep the signal comes from, but the column energy still helps set the probe width.

Core mathematical model

Energy and dose

El=Ec−e∣Vb∣−e∣Vq∣Ninc=Iτ/eD=Ninc/p2\begin{aligned}E_l&=E_c-e|V_b|-e|V_q|\\N_{inc}&=I\tau/e\\D&=N_{inc}/p^2\end{aligned}

A negative stage bias and negative charging both lower the landing energy. If it falls to zero or below, the electrons never land and the scan records nothing.

Assumed depth weighting and blur

L=12 nm(El/keV)1.5wt=1−e−t/LH(k)=e−k2σ2/2\begin{aligned}L&=12\,\mathrm{nm}(E_l/\mathrm{keV})^{1.5}\\w_t&=1-e^{-t/L}\\H(k)&=e^{-k^2\sigma^2/2}\end{aligned}

An exponential depth weighting and a Gaussian blur both weaken the film pattern in the signal.

Harmonic from recorded counts

C^=2n∑jyjcos⁡(kxj)S^=2n∑jyjsin⁡(kxj)A^=C^2+S^2\begin{aligned}\widehat C&=\frac{2}{n}\sum_jy_j\cos(kx_j)\\\widehat S&=\frac{2}{n}\sum_jy_j\sin(kx_j)\\\widehat A&=\sqrt{\widehat C^2+\widehat S^2}\end{aligned}

Only the normalized counts enter the fit; the known film and instrument parameters are not used.

Common difficulties

A clearer image does not carry more information

Typical misconceptionIf the image looks clearer, my interpretation must be right.

Better mental modelDisplay gain does not change the counts, and it cannot unmix the physical effects. The input reference is shown separately and is not a measurement.

Run the experiment

  1. 01

    Predict surface sensitivity

    Load the Buried substrate dominates preset and compare it with the known film pattern and depth kernel.

    What to observe: Even a narrow beam can pick up mostly substrate signal.
  2. 02

    Separate column from landing

    Compare a 0.5 keV column with a 5 keV column plus a 4.5 kV negative stage bias.

    What to observe: Both beams land at the same energy and weight depth the same way, but their assumed probe widths differ.
  3. 03

    See how charging limits the scan

    Load the Charging and frame shear preset, then raise the charge drainage. Watch the beam displacement and the fit to the stored rows.

    What to observe: A longer dwell gives more counts, but the charge that builds up can also shift where the beam lands.
  4. 04

    Recover the surface pattern

    Slow the beam with the bias, keep the column energy up, drain the charge and pick a balanced dwell. Then press Check target.

    What to observe: Display gain changes how the image looks, but it cannot bring back a harmonic that was lost.
FeedbackSomething broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy