Energy and dose
A negative stage bias and negative charging both lower the landing energy. If it falls to zero or below, the electrons never land and the scan records nothing.
E13 · Field emission and beam deceleration
Can electrons travel fast in the column and land gently? Yes. A negative voltage on the specimen slows the electrons just before they land. They travel down the column at high energy and land with low energy, so the interaction stays shallow, close to the surface.
Run the electron column at high energy, then slow the beam down with a negative bias on the specimen. Try to recover a thin-film pattern by trading off depth sensitivity, probe width, counting noise and the assumed charging.
Physics tutorial
BackgroundAn SEM can run its column at high energy and still land electrons gently by putting a negative bias on the specimen. Here you try to recover a thin-film pattern while trading off depth sensitivity, probe width, counting noise and the assumed charging.
Why it mattersThe task is to recover the film's harmonic amplitude to within 0.02 of the separate input. Along the way, keep the contrast-to-noise ratio at a representative pixel at 8 or more, the response FWHM at 8 nm or less, the film depth weight at 90 percent or more, and the assumed charging at 20 V or less.
Start with the essentials
A negative stage bias and negative charging both lower the landing energy. If it falls to zero or below, the electrons never land and the scan records nothing.
An exponential depth weighting and a Gaussian blur both weaken the film pattern in the signal.
Only the normalized counts enter the fit; the known film and instrument parameters are not used.
Typical misconceptionIf the image looks clearer, my interpretation must be right.
Better mental modelDisplay gain does not change the counts, and it cannot unmix the physical effects. The input reference is shown separately and is not a measurement.
Load the Buried substrate dominates preset and compare it with the known film pattern and depth kernel.
What to observe: Even a narrow beam can pick up mostly substrate signal.Compare a 0.5 keV column with a 5 keV column plus a 4.5 kV negative stage bias.
What to observe: Both beams land at the same energy and weight depth the same way, but their assumed probe widths differ.Load the Charging and frame shear preset, then raise the charge drainage. Watch the beam displacement and the fit to the stored rows.
What to observe: A longer dwell gives more counts, but the charge that builds up can also shift where the beam lands.Slow the beam with the bias, keep the column energy up, drain the charge and pick a balanced dwell. Then press Check target.
What to observe: Display gain changes how the image looks, but it cannot bring back a harmonic that was lost.