Skip to main content
Sandbox Physics

E63 · Electron excitation and surface spectroscopy

Inverse Photoemission: Empty States

What can emitted photons tell us about empty electron states? Where they lie in energy. An incoming electron drops into an empty state and emits a photon, and the detector accepts only one fixed photon energy. Scanning the electron energy then locates the unoccupied states, broadened by the instrument resolution.

Step the energy of the incoming electrons while a detector counts photons in one fixed energy band. Find the two peaks where electrons drop into empty states, then see how photon bandwidth, electron energy spread, exposure and energy calibration change them.

Interactive modelInverse Photoemission: Empty States
First count-derived centroid—\text{—}
Second count-derived centroid—\text{—}
Weaker acquired peak SNR—\text{—}
Combined instrument FWHM—\text{—}
More measurements and target
Selected stored photon count—\text{—}
Selected surface kinetic energy—\text{—}
Larger centroid error against input—\text{—}
Acquired electron-energy positions—\text{—}
Summed incident dose—\text{—}
Prescribed changed-response fraction—\text{—}
Experiment target—\text{—}

Physics tutorial

Inverse Photoemission: Empty States

BackgroundAn electron that enters a surface can drop into an empty state and give off a photon. Here you step the energy of the incoming electrons while a detector counts photons in one fixed band. You find two empty-state peaks in the counts, then test how photon bandwidth, electron energy spread, exposure and energy calibration affect them.

Why it mattersThe goal is to place each peak within 0.12 eV of its undamaged position. Each peak needs a signal-to-noise ratio of at least 10, the combined instrument FWHM must be at most 0.8 eV, and the set damage rule may change at most 15 percent of the response.

Start with the essentials

Focus question
What can emitted photons tell us about empty electron states?
One-sentence intuition
The electron brings in energy and the photon carries a fixed energy away; the difference tells you where the empty state sits.

Core mathematical model

Energy of the empty final state

εf=Ekin+Φ−hνEkin=εf+hν−Φ\begin{aligned}\varepsilon_f&=E_{kin}+\Phi-h\nu\\E_{kin}&=\varepsilon_f+h\nu-\Phi\end{aligned}

Kinetic energy is measured from the local vacuum level, and the final-state energy from the Fermi level. A calibration error shifts the whole energy axis.

Instrument resolution from two Gaussian widths

winst=we2+wγ2σ=winst22ln⁡2\begin{aligned}w_{inst}&=\sqrt{w_e^2+w_\gamma^2}\\\sigma&=\frac{w_{inst}}{2\sqrt{2\ln2}}\end{aligned}

The electron and photon widths are independent Gaussians, so they add in quadrature. The natural width of each state comes on top of this.

Expected photon counts at each electron energy

ε0=E+Φ−hνR(E)=∫0∞ρu(u)×Gσ(u−ε0) duμ(E)=bτ+IτeηAγR(E)\begin{aligned}\varepsilon_0&=E+\Phi-h\nu\\R(E)&=\int_0^\infty\rho_u(u)\\&\quad\times G_\sigma(u-\varepsilon_0)\,du\\\mu(E)&=b\tau+\frac{I\tau}{e}\eta A_\gamma R(E)\end{aligned}

A wider photon band has a larger area, so it collects more counts. The recorded scan is the empty-state spectrum broadened and weighted by the instrument; it is not a reconstruction of the ideal input.

Common difficulties

Reference curves are not data

Typical misconceptionThe ideal input curve is the result inferred from the scan.

Better mental modelThe gold curves show the known input and serve only as an independent check. The teal result comes from the stored counts alone.

Longer exposure does not fix every error

Typical misconceptionIf I collect enough counts, the answer will be right.

Better mental modelMore counts reduce counting noise. Errors from the continuum background, the energy calibration, the finite instrument width and the set damage remain.

Run the experiment

  1. 01

    Predict which energy is scanned

    Pick a stored electron-energy position and compare its photon count with the energy bookkeeping card.

    What to observe: The photon band the detector counts stays fixed; only the energy of the incoming electrons changes.
  2. 02

    Recalibrate the same counts

    Change the assumed work function from 3.5 to 4.5 eV. Then change the smoothing and check that the raw scan does not change.

    What to observe: A wrong work function shifts both inferred peaks by the same energy.
  3. 03

    Trade counts against resolution

    Compare a wide photon band with a broad spread of incoming electron energies. Look at the count rate and the combined FWHM.

    What to observe: More counts do not undo the broadening, and they cannot separate features that sit too close to resolve.
  4. 04

    Recover the two peaks

    Set narrow electron and photon widths, the calibrated work function and enough dwell time. Compare the centroid energies with the separate reference, then check the target.

    What to observe: The target checks signal-to-noise, energy error, instrument width and the set amount of damage.
FeedbackSomething broken, confusing, or missing on this page?Tell us
What kind of feedback?

Sent with this page’s address, its Lab ID, the site language and your device type. Please leave out personal details. Feedback is kept for up to 12 months.Privacy policy