# Why a xenon detector needs liquid and gas

The gas above liquid xenon turns a small amount of drifting charge into a measurable second flash. The earlier flash marks the interaction time. The delayed light reads out the surviving electrons. With calibration, their separation in time and the pattern across the sensors help locate the interaction.

Sandbox Physics · 2026-11-16

## One interaction leaves light and charge

XENONnT uses liquid and gaseous xenon at the underground Gran Sasso laboratory in Italy. The liquid packs a large number of target atoms into the chamber. Deposited energy can leave xenon in excited states, which temporarily store energy, or knock electrons out of atoms. The xenon emits ultraviolet light as some of that energy is released. Electrons that recombine with positively charged xenon add to the light.

This prompt signal is called S1. Photomultiplier tubes above and below the xenon convert the light into electrical pulses, which the electronics record as waveforms. The electrons left free drift upward under the applied field. Their much slower journey provides the delay that the detector can time.

Sources: [1] [2] [8]

## The electric field powers the second flash

A stronger field near the surface extracts electrons into the gas. Between collisions they gain energy from the field, then excite xenon atoms. One electron can excite atoms repeatedly as it crosses the gap, producing the S2 signal. The high-voltage supply provides this additional light energy. Calibration converts that light back into an estimate of the original charge.

Sources: [4] [1]

## A delay and a light pattern locate the interaction

Electrons from a deeper interaction generally take longer to reach the surface. The interval between S1 and S2 gives a depth estimate once the drift speed has been calibrated. Clock scale, timing zero and delays associated with extraction and light production all enter that measurement.

The pattern of S2 across the top array locates the electrons as they enter the gas. Together, that horizontal position and the drift time provide a three-dimensional location, the time-projection principle. Field curvature can move electrons sideways along the way, so calibration must map their extraction position back into the liquid. The resulting location also lets the experiment select an inner volume shielded by surrounding xenon.

Sources: [1] [3] [2]

## Purity and light collection need their own corrections

Impurities can capture electrons in the liquid. A longer journey gives them more chances, which can reduce S2 for deep interactions. Calibration events of known energy at different depths reveal this attenuation. Its characteristic survival time is called the electron lifetime.

Light collection varies across the chamber too. The same amount of emitted light can give different signal areas near the centre and the edge. Known calibration events at many positions measure that spatial response.

A further complication is that the local field changes the light and charge yield. If the field varies with depth, a lifetime fit can mix this effect with impurity losses. Correcting the field response and light collection puts interactions at different depths on a common energy scale.

Selection has a calibration too. A small flash may blend into electronic noise or fail a threshold, and its partner may then be paired incorrectly. Calibration samples measure how often signals of different sizes are found and selected, so the missed population enters the analysis.

Sources: [3] [7]

## Light and charge together set the energy scale

An electron recoil transfers energy mainly to electrons in the xenon. For the same deposited energy, more recombination can leave more light and fewer free electrons. Measuring both helps recover the energy. Several known calibration peaks establish the conversion from their combined signals.

Signal area is often expressed in photoelectron units: the average electrical pulse from one electron released at the photocathode sets the unit. A single ultraviolet photon can release two such electrons. The prompt-photon gain $g_1$ is the mean signal per emitted prompt scintillation photon and includes light collection and detector response.

The effective charge gain $g_2$ also includes extraction efficiency, the fraction of electrons reaching the liquid surface that enter the gas. It differs from the mean S2 produced by one electron already extracted, often called the single-electron gain. The charge being inferred has survived the early recombination but has yet to experience drift losses. Attachment corrections and the extraction fraction both enter that conversion.

Dividing the corrected signals by their gains estimates the scintillation photons and electrons left free at the start of drift. Their sum, multiplied by an adopted mean energy scale per quantum, gives the reconstructed electron-recoil energy. Counting fluctuations and gain uncertainties enter the estimate. A shared calibration uncertainty affects an entire set of events together.

$$
E_{\mathrm{ER}}=W\left(\frac{\mathrm{cS1}}{g_1}+\frac{\mathrm{cS2}}{g_2}\right)
$$

$E_{\mathrm{ER}}$ is the calibrated electron-recoil energy estimate. $\mathrm{cS1},\mathrm{cS2}$ are corrected S1 and S2 signal areas; S2 includes the correction for electron attachment during drift. $g_1,g_2$ are the mean gain factors described above. The two quotients estimate prompt scintillation photons and electrons left free at the start of drift. $W$ is an energy-per-quantum scale for the mean response. A photon’s energy is set by its wavelength. Individual signal sizes fluctuate, so the relation provides a calibrated energy estimate for an event.

Sources: [3] [8] [6] [5]

## Recoil populations overlap

A particle can instead strike a xenon nucleus. More of a nuclear recoil’s energy goes into atomic motion that the light and charge signals do not detect. Its light-to-charge response differs from an electron recoil: at comparable S1, nuclear recoils generally have smaller S2. The calibrated populations overlap, and the electron-recoil energy formula is not a direct measure of total nuclear-recoil energy.

Neutrons and neutrinos can produce nuclear recoils too. Rock overburden reduces cosmic rays, and shielding and outer veto detectors help reduce or tag some neutron and other particle backgrounds. Neutrino recoils remain part of the expected sample.

Unrelated S1 and S2 signals can also be paired by mistake. A dark-matter search compares candidates with calibrated signal and background distributions; the two flashes supply the position and response information that comparison needs.

Sources: [8] [1]

## [Compare two flashes and drift position](https://sandboxphysics.com/labs/dual-phase-xenon-tpc/)

Change deposit depth and transverse position, then acquire and compare the two light distributions.

Prescribed drift and ideal photoelectron counting provide conditional position and visible-energy estimates.

## Sources and further reading

1. [XENON Collaboration · Time Projection Chamber](https://xenonexperiment.org/time-projection-chamber/)

   Checked 14 November 2026: read the official explanation of S1, upward charge drift, gas S2, position and recoil response distributions. The older INFN host timed out; this is the current official domain. Calibration illustrations include XENON1T and are not adopted as current XENONnT performance. No image is copied.

2. [Aprile et al. · The XENONnT Dark Matter Experiment (2024)](https://arxiv.org/html/2402.10446v1)

   Checked 14 November 2026: read sections 2.1 and 2.1.1 of arXiv:2402.10446v1 on the liquid/gas regions, sensor arrays and electrodes, plus selected lifetime/gain passages in 3.3.3–3.3.4. Uses the detector principle, without adopting dimensions, voltages, channel counts or operating records. This is not a full-paper reading.

3. [Aprile et al. · XENONnT Analysis: Signal Reconstruction, Calibration and Event Selection (2024 preprint)](https://arxiv.org/html/2409.08778v1)

   Checked 14 November 2026: selected passages in section III on digitized signals, IV.1–IV.3 on timing/position, V.2 on S2 corrections, and VII on ER combined energy and calibration peaks in arXiv:2409.08778v1. Uses the measurement chain, not a reproduction of the full analysis, selection acceptance or current sensitivity.

4. [Monteiro et al. · Secondary Scintillation Yield in Pure Xenon (2007)](https://arxiv.org/pdf/physics/0702142)

   Checked 14 November 2026: read the abstract and PDF pages 1–2 on field-powered electron excitation, with selected method/result text on pages 4–8. Uses the energy source and proportional-scintillation principle. Room-temperature gas coefficients are not transferred to a cryogenic detector and no figure or table is copied.

5. [Faham et al. · Measurements of wavelength-dependent double photoelectron emission (2015)](https://arxiv.org/abs/1506.08748v2)

   Checked 14 November 2026: read the abstract and version metadata of arXiv:1506.08748v2 for the fact that one VUV photon can release two photoelectrons. No tube-specific probability is adopted; neither a complete paper reading nor access to its raw data is claimed.

6. [Aprile et al. · Low-energy calibration of XENON1T with an internal 37Ar source (2023)](https://link.springer.com/article/10.1140/epjc/s10052-023-11512-z)

   Checked 14 November 2026: read the section 3 S1–S2 discussion of W, gain conventions and energy calibration. Uses the need for a consistent mean scale and gains; XENON1T coefficients are not presented as current XENONnT measurements.

7. [XENON Collaboration · Design and performance of the field cage for the XENONnT experiment (2024)](https://link.springer.com/article/10.1140/epjc/s10052-023-12296-y)

   Checked 14 November 2026: read sections 1–3 on field shaping and selected 4.1–4.2 passages on field-dependent charge yield and lifetime fits. Uses that confounding mechanism; no COMSOL map is obtained/digitized and no run-specific value is adopted.

8. [Aprile et al. · XENONnT WIMP Search: Signal & Background Modeling and Statistical Inference (2024 preprint)](https://arxiv.org/html/2406.13638v1)

   Checked 14 November 2026: selected II.1–II.2 on empirical light/charge response, appendix C on extraction and single-electron gain, and III.2–III.5 on recoil backgrounds and accidental pairing. Uses statistical discrimination and energy partition; no cross-section, rate, sensitivity or search likelihood is adopted.

[How a vernier reads between millimetres](https://sandboxphysics.com/stories/instruments/length-metrology/) [Air helps hold up the object on a balance](https://sandboxphysics.com/stories/instruments/precision-balance/) [How Cavendish Measured the Attraction Between Spheres](https://sandboxphysics.com/stories/instruments/cavendish-balance/) [Which pressure does a liquid column measure?](https://sandboxphysics.com/stories/instruments/manometer/) [The cup absorbs heat too](https://sandboxphysics.com/stories/instruments/calorimeter/) [Rotational rheometers need both torque and speed](https://sandboxphysics.com/stories/instruments/rotational-rheometer/) [Why the air beyond a pipe changes its resonance](https://sandboxphysics.com/stories/instruments/resonance-tube/) [How a seismometer reads the ground through motion inside its case](https://sandboxphysics.com/stories/instruments/seismometer/) [Why does a meter change the circuit?](https://sandboxphysics.com/stories/instruments/electrical-meters/) [How a null detector reveals a resistance](https://sandboxphysics.com/stories/instruments/resistance-bridges/) [Why a capacitor can behave like an inductor](https://sandboxphysics.com/stories/instruments/lcr-impedance/) [Where does the slow wave on an oscilloscope come from?](https://sandboxphysics.com/stories/instruments/digital-oscilloscope/) [How a lock-in amplifier finds a signal buried in noise](https://sandboxphysics.com/stories/instruments/lock-in-amplifier/) [Where a small peak in a spectrum can come from](https://sandboxphysics.com/stories/instruments/spectrum-analyzer/) [How a network analyzer separates the cable from the device](https://sandboxphysics.com/stories/instruments/vector-network-analyzer/) [When a slower quartz crystal can measure added mass](https://sandboxphysics.com/stories/instruments/quartz-microbalance/) [How an echelle spectrograph gives overlapping colors a second direction](https://sandboxphysics.com/stories/instruments/echelle-spectrograph/) [How total brightness reveals many infrared colors](https://sandboxphysics.com/stories/instruments/ftir-spectrometer/) [Measuring a Solution’s Concentration with Light](https://sandboxphysics.com/stories/instruments/uv-vis-spectrophotometer/) [How an ellipsometer reads thin-film thickness from reflected light](https://sandboxphysics.com/stories/instruments/spectroscopic-ellipsometer/) [How a polarimeter reveals mixed polarization](https://sandboxphysics.com/stories/instruments/mueller-polarimeter/) [How individual photons reveal a fluorescence lifetime](https://sandboxphysics.com/stories/instruments/tcspc-fluorescence-lifetime/) [Brillouin Microscopy Listens to Sound with Light](https://sandboxphysics.com/stories/instruments/brillouin-microscope/) [How a white sphere compares a lamp’s total output](https://sandboxphysics.com/stories/instruments/integrating-sphere-radiometer/) [How powder rings reveal atomic spacing](https://sandboxphysics.com/stories/instruments/powder-xray-diffractometer/) [How Small-Angle X-ray Scattering Reads a Particle Halo](https://sandboxphysics.com/stories/instruments/small-angle-xray-scattering/) [Why X-ray fluorescence depends on sample thickness](https://sandboxphysics.com/stories/instruments/xray-fluorescence-spectrometer/) [Why an AFM makes a bump look wider](https://sandboxphysics.com/stories/instruments/atomic-force-microscope/) [How a small indentation separates hardness from elasticity](https://sandboxphysics.com/stories/instruments/nanoindenter/) [How Differential Scanning Calorimetry Reads a Melting Peak](https://sandboxphysics.com/stories/instruments/differential-scanning-calorimeter/) [How thermogravimetry separates mass loss from the gas around it](https://sandboxphysics.com/stories/instruments/thermogravimetric-analyzer/) [How a diamond anvil cell reads pressure in ruby light](https://sandboxphysics.com/stories/instruments/diamond-anvil-cell/) [How diamond fluorescence reveals a magnetic-field direction](https://sandboxphysics.com/stories/instruments/nv-center-magnetometer/) [Reading a magnetic field in the rhythm of atoms](https://sandboxphysics.com/stories/instruments/optically-pumped-magnetometer/) [Keeping a laser in time with atoms](https://sandboxphysics.com/stories/instruments/optical-lattice-clock/) [Why a Kibble balance needs two experiments](https://sandboxphysics.com/stories/instruments/kibble-balance/) [How a dilution refrigerator keeps carrying heat away](https://sandboxphysics.com/stories/instruments/dilution-refrigerator/) [Why a bigger vacuum pump may barely help the sample](https://sandboxphysics.com/stories/instruments/ultra-high-vacuum-system/) [How microwaves read a superconducting qubit](https://sandboxphysics.com/stories/instruments/superconducting-qubit-readout/) [How current leaves a voltage across a thin sheet](https://sandboxphysics.com/stories/instruments/four-probe-hall/) [How an ion flight becomes a mass spectrum](https://sandboxphysics.com/stories/instruments/time-of-flight-mass-spectrometer/) [How four straight rods select ion mass](https://sandboxphysics.com/stories/instruments/quadrupole-mass-filter/) [How a rapidly reversing field can hold ions](https://sandboxphysics.com/stories/instruments/paul-ion-trap-mass-spectrometer/) [How a cyclotron keeps a proton on the accelerating beat](https://sandboxphysics.com/stories/instruments/cyclotron-synchrocyclotron/) [Why the drift tubes in an RF linac get longer](https://sandboxphysics.com/stories/instruments/radio-frequency-linac/) [The directions hidden inside a small beam spot](https://sandboxphysics.com/stories/instruments/beam-diagnostics/) [How defocusing magnets help a beam stay in a ring](https://sandboxphysics.com/stories/instruments/synchrotron-storage-ring/) [Why a collider cannot keep every collision](https://sandboxphysics.com/stories/instruments/collider-trigger-daq/) [Getting synchrotron light onto a sample](https://sandboxphysics.com/stories/instruments/synchrotron-beamline/) [How a free-electron laser gathers electrons into a brighter beam](https://sandboxphysics.com/stories/instruments/xray-free-electron-laser/) [Following energy exchange with neutron flight times](https://sandboxphysics.com/stories/instruments/pulsed-neutron-time-of-flight/) [How a tokamak holds heat and measures where it goes](https://sandboxphysics.com/stories/instruments/tokamak-iter/) [Why stellarator coils bend in three dimensions](https://sandboxphysics.com/stories/instruments/w7x-stellarator/) [How NIF turns a laser pulse into an implosion](https://sandboxphysics.com/stories/instruments/nif-inertial-fusion/) [How ELI brings an intense pulse to a target](https://sandboxphysics.com/stories/instruments/eli-ultra-intense-laser/) [How LIGO measures a gravitational wave with mirrors that move](https://sandboxphysics.com/stories/instruments/ligo-feedback-calibration/) [How delayed laser records let LISA measure gravitational waves](https://sandboxphysics.com/stories/instruments/lisa-delayed-links/) [How FAST brings faint radio signals into focus](https://sandboxphysics.com/stories/instruments/fast-radio-telescope/) [How ALMA keeps separated antennas in step](https://sandboxphysics.com/stories/instruments/alma-interferometer/) [How SKA-Low steers a beam without moving its antennas](https://sandboxphysics.com/stories/instruments/ska-beamforming/) [How EHT connects telescopes with clocks and hard drives](https://sandboxphysics.com/stories/instruments/eht-vlbi/) [How Webb separates a faint spectrum from infrared background](https://sandboxphysics.com/stories/instruments/jwst-infrared-spectroscopy/) [How LHAASO finds an incoming direction from ground particles](https://sandboxphysics.com/stories/instruments/lhaaso-air-showers/)

## Keep reading

How physics instruments measure · Reading path

### How physics instruments measure

1. [Article **How a vernier reads between millimetres**](https://sandboxphysics.com/stories/instruments/length-metrology/)
2. [Article **Air helps hold up the object on a balance**](https://sandboxphysics.com/stories/instruments/precision-balance/)
3. [Article **How Cavendish Measured the Attraction Between Spheres**](https://sandboxphysics.com/stories/instruments/cavendish-balance/)
4. [Article **Which pressure does a liquid column measure?**](https://sandboxphysics.com/stories/instruments/manometer/)
5. [Article **The cup absorbs heat too**](https://sandboxphysics.com/stories/instruments/calorimeter/)
6. [Article **Rotational rheometers need both torque and speed**](https://sandboxphysics.com/stories/instruments/rotational-rheometer/)
7. [Article **Why the air beyond a pipe changes its resonance**](https://sandboxphysics.com/stories/instruments/resonance-tube/)
8. [Article **How a seismometer reads the ground through motion inside its case**](https://sandboxphysics.com/stories/instruments/seismometer/)
9. [Article **Why does a meter change the circuit?**](https://sandboxphysics.com/stories/instruments/electrical-meters/)
10. [Article **How a null detector reveals a resistance**](https://sandboxphysics.com/stories/instruments/resistance-bridges/)
11. [Article **Why a capacitor can behave like an inductor**](https://sandboxphysics.com/stories/instruments/lcr-impedance/)
12. [Article **Where does the slow wave on an oscilloscope come from?**](https://sandboxphysics.com/stories/instruments/digital-oscilloscope/)
13. [Article **How a lock-in amplifier finds a signal buried in noise**](https://sandboxphysics.com/stories/instruments/lock-in-amplifier/)
14. [Article **Where a small peak in a spectrum can come from**](https://sandboxphysics.com/stories/instruments/spectrum-analyzer/)
15. [Article **How a network analyzer separates the cable from the device**](https://sandboxphysics.com/stories/instruments/vector-network-analyzer/)
16. [Article **When a slower quartz crystal can measure added mass**](https://sandboxphysics.com/stories/instruments/quartz-microbalance/)
17. [Article **How an echelle spectrograph gives overlapping colors a second direction**](https://sandboxphysics.com/stories/instruments/echelle-spectrograph/)
18. [Article **How total brightness reveals many infrared colors**](https://sandboxphysics.com/stories/instruments/ftir-spectrometer/)
19. [Article **Measuring a Solution’s Concentration with Light**](https://sandboxphysics.com/stories/instruments/uv-vis-spectrophotometer/)
20. [Article **How an ellipsometer reads thin-film thickness from reflected light**](https://sandboxphysics.com/stories/instruments/spectroscopic-ellipsometer/)
21. [Article **How a polarimeter reveals mixed polarization**](https://sandboxphysics.com/stories/instruments/mueller-polarimeter/)
22. [Article **How individual photons reveal a fluorescence lifetime**](https://sandboxphysics.com/stories/instruments/tcspc-fluorescence-lifetime/)
23. [Article **Brillouin Microscopy Listens to Sound with Light**](https://sandboxphysics.com/stories/instruments/brillouin-microscope/)
24. [Article **How a white sphere compares a lamp’s total output**](https://sandboxphysics.com/stories/instruments/integrating-sphere-radiometer/)
25. [Article **How powder rings reveal atomic spacing**](https://sandboxphysics.com/stories/instruments/powder-xray-diffractometer/)
26. [Article **How Small-Angle X-ray Scattering Reads a Particle Halo**](https://sandboxphysics.com/stories/instruments/small-angle-xray-scattering/)
27. [Article **Why X-ray fluorescence depends on sample thickness**](https://sandboxphysics.com/stories/instruments/xray-fluorescence-spectrometer/)
28. [Article **Why an AFM makes a bump look wider**](https://sandboxphysics.com/stories/instruments/atomic-force-microscope/)
29. [Article **How a small indentation separates hardness from elasticity**](https://sandboxphysics.com/stories/instruments/nanoindenter/)
30. [Article **How Differential Scanning Calorimetry Reads a Melting Peak**](https://sandboxphysics.com/stories/instruments/differential-scanning-calorimeter/)
31. [Article **How thermogravimetry separates mass loss from the gas around it**](https://sandboxphysics.com/stories/instruments/thermogravimetric-analyzer/)
32. [Article **How a diamond anvil cell reads pressure in ruby light**](https://sandboxphysics.com/stories/instruments/diamond-anvil-cell/)
33. [Article **How diamond fluorescence reveals a magnetic-field direction**](https://sandboxphysics.com/stories/instruments/nv-center-magnetometer/)
34. [Article **Reading a magnetic field in the rhythm of atoms**](https://sandboxphysics.com/stories/instruments/optically-pumped-magnetometer/)
35. [Article **Keeping a laser in time with atoms**](https://sandboxphysics.com/stories/instruments/optical-lattice-clock/)
36. [Article **Why a Kibble balance needs two experiments**](https://sandboxphysics.com/stories/instruments/kibble-balance/)
37. [Article **How a dilution refrigerator keeps carrying heat away**](https://sandboxphysics.com/stories/instruments/dilution-refrigerator/)
38. [Article **Why a bigger vacuum pump may barely help the sample**](https://sandboxphysics.com/stories/instruments/ultra-high-vacuum-system/)
39. [Article **How microwaves read a superconducting qubit**](https://sandboxphysics.com/stories/instruments/superconducting-qubit-readout/)
40. [Article **How current leaves a voltage across a thin sheet**](https://sandboxphysics.com/stories/instruments/four-probe-hall/)
41. [Article **How an ion flight becomes a mass spectrum**](https://sandboxphysics.com/stories/instruments/time-of-flight-mass-spectrometer/)
42. [Article **How four straight rods select ion mass**](https://sandboxphysics.com/stories/instruments/quadrupole-mass-filter/)
43. [Article **How a rapidly reversing field can hold ions**](https://sandboxphysics.com/stories/instruments/paul-ion-trap-mass-spectrometer/)
44. [Article **How a cyclotron keeps a proton on the accelerating beat**](https://sandboxphysics.com/stories/instruments/cyclotron-synchrocyclotron/)
45. [Article **Why the drift tubes in an RF linac get longer**](https://sandboxphysics.com/stories/instruments/radio-frequency-linac/)
46. [Article **The directions hidden inside a small beam spot**](https://sandboxphysics.com/stories/instruments/beam-diagnostics/)
47. [Article **How defocusing magnets help a beam stay in a ring**](https://sandboxphysics.com/stories/instruments/synchrotron-storage-ring/)
48. [Article **Why a collider cannot keep every collision**](https://sandboxphysics.com/stories/instruments/collider-trigger-daq/)
49. [Article **Getting synchrotron light onto a sample**](https://sandboxphysics.com/stories/instruments/synchrotron-beamline/)
50. [Article **How a free-electron laser gathers electrons into a brighter beam**](https://sandboxphysics.com/stories/instruments/xray-free-electron-laser/)
51. [Article **Following energy exchange with neutron flight times**](https://sandboxphysics.com/stories/instruments/pulsed-neutron-time-of-flight/)
52. [Article **How a tokamak holds heat and measures where it goes**](https://sandboxphysics.com/stories/instruments/tokamak-iter/)
53. [Article **Why stellarator coils bend in three dimensions**](https://sandboxphysics.com/stories/instruments/w7x-stellarator/)
54. [Article **How NIF turns a laser pulse into an implosion**](https://sandboxphysics.com/stories/instruments/nif-inertial-fusion/)
55. [Article **How ELI brings an intense pulse to a target**](https://sandboxphysics.com/stories/instruments/eli-ultra-intense-laser/)
56. [You are here **Why a xenon detector needs liquid and gas**](https://sandboxphysics.com/stories/instruments/xenonnt-dual-phase-tpc/)
57. [Article **How LIGO measures a gravitational wave with mirrors that move**](https://sandboxphysics.com/stories/instruments/ligo-feedback-calibration/)
58. [Article **How delayed laser records let LISA measure gravitational waves**](https://sandboxphysics.com/stories/instruments/lisa-delayed-links/)
59. [Article **How FAST brings faint radio signals into focus**](https://sandboxphysics.com/stories/instruments/fast-radio-telescope/)
60. [Article **How ALMA keeps separated antennas in step**](https://sandboxphysics.com/stories/instruments/alma-interferometer/)
61. [Article **How SKA-Low steers a beam without moving its antennas**](https://sandboxphysics.com/stories/instruments/ska-beamforming/)
62. [Article **How EHT connects telescopes with clocks and hard drives**](https://sandboxphysics.com/stories/instruments/eht-vlbi/)
63. [Article **How Webb separates a faint spectrum from infrared background**](https://sandboxphysics.com/stories/instruments/jwst-infrared-spectroscopy/)
64. [Article **How LHAASO finds an incoming direction from ground particles**](https://sandboxphysics.com/stories/instruments/lhaaso-air-showers/)