[Volume 5 · Heat, Probability and Matter](https://sandboxphysics.com/learn/heat-and-statistics/)

# When does a thermometer represent the sample temperature?

Check thermal equilibrium, calibration and probe response before treating a thermometer reading as the sample temperature.

Before you start: [Volume 1 Chapter 2](https://sandboxphysics.com/learn/mechanics/measuring-motion-with-a-camera/) and [Chapter 3](https://sandboxphysics.com/learn/mechanics/testing-constant-acceleration/) on measurement, calibration and model checks. Basic algebra.

After this chapter you can: Use a thermometer scale, calculate an equilibrium temperature and explain the lag between a probe and its sample.

A thermometer represents a sample when it is at the same temperature as the part of the sample it touches. Give the probe time to exchange heat, and check that the probe scarcely changes the sample's temperature. The reading also needs a calibration that converts the sensor's signal to a temperature.

Volume 1, Chapters 2 and 3, compared instrument readings with reference values and tracked changes in time. Here we apply the same ideas to a thermometer in a bath.

## 1.1 Bodies in thermal equilibrium have the same temperature

Put a cooler thermometer into a warmer bath. Energy passes into the thermometer as heat. The temperature difference shrinks as the thermometer warms. When the two are in thermal equilibrium, there is no net heat transfer between them.

The zeroth law says that if two bodies are each in thermal equilibrium with a third body, they are in thermal equilibrium with each other. A thermometer can serve as that third body. Two baths that bring it to the same equilibrium reading have the same temperature at the points of contact. For this comparison, the thermometer's surroundings and contact must let it settle to the bath temperature. A probe attached to a cold support may settle to a steady reading even while heat continues to cross it. [OpenStax, University Physics Volume 2, Section 1.1](https://openstax.org/books/university-physics-volume-2/pages/1-1-temperature-and-thermal-equilibrium).

Heat can travel through a contact by conduction. Moving fluid can transport energy by convection. Radiation carries energy across space, even without a material connection. Stirring a bath can reduce temperature differences within it. A probe at one position measures a local temperature, so check more than one position if the bath is uneven. [OpenStax, University Physics Volume 2, Section 1.6](https://openstax.org/books/university-physics-volume-2/pages/1-6-mechanisms-of-heat-transfer).

## 1.2 Calibration turns a sensor reading into degrees

A thermometer may report a voltage, a resistance or a raw number that changes with temperature. In a narrow working range, suppose its raw reading $r$ depends linearly on the numerical Celsius value $T_{\mathrm C}$. Then $r=aT_{\mathrm C}+b$, where the gain $a$ sets the change per degree and the intercept $b$ sets the reading at zero Celsius. Two reference temperatures determine both. For a new reading,

$$
T_{\mathrm C}=\frac{r-b}{a}.
$$

The kelvin value is $T_{\mathrm K}=T_{\mathrm C}+273.15$. A change of one Celsius degree has the same size as a change of one kelvin. The two scales put zero at different points, so use the absolute kelvin temperature when a law calls for a ratio of absolute temperatures. [OpenStax, University Physics Volume 2, Section 1.2](https://openstax.org/books/university-physics-volume-2/pages/1-2-thermometers-and-temperature-scales).

The simulated sensor in Section 1.4 reads $20.24$ at a reference of $20\,^\circ\mathrm C$ and $40.30$ at $40\,^\circ\mathrm C$. The change in raw reading divided by the reference change gives $a=1.003$. The first reference gives $b=0.18$. The calibration corrects scale and zero. A reading can still lag while the probe warms.

## 1.3 Heat capacity sets how much a sample changes during contact

A body with heat capacity $C$ needs energy $Q=C\Delta T$ to change its temperature by $\Delta T$, when its heat capacity is nearly constant, the body stays in the same phase and mechanical work is negligible. Heat capacity is measured in joules per kelvin. Specific heat $c$ is the heat capacity per unit mass, so a body of mass $m$ has $C=mc$. [OpenStax, University Physics Volume 2, Section 1.4](https://openstax.org/books/university-physics-volume-2/pages/1-4-heat-transfer-specific-heat-and-calorimetry).

Bring two bodies starting at $T_1$ and $T_2$ into contact and insulate them from the room. With nearly constant heat capacities $C_1$ and $C_2$, one body's energy loss is the other's gain. Their common final temperature $T_f$ satisfies

$$
C_1(T_f-T_1)+C_2(T_f-T_2)=0.
$$

Solving for the final temperature weights each starting temperature by its heat capacity.

$$
T_f=\frac{C_1T_1+C_2T_2}{C_1+C_2}.
$$

For example, mix $0.100\,\mathrm{kg}$ of water at $60.0\,^\circ\mathrm C$ with $0.200\,\mathrm{kg}$ at $20.0\,^\circ\mathrm C$. With a cup of negligible heat capacity, the two amounts have the same specific heat. Their final temperature is about $33.3\,^\circ\mathrm C$. It lies closer to the colder starting temperature because there is twice as much cold water.

The same balance applies to a sample and its probe. If the probe has much less heat capacity than the sample, their final temperature stays near the sample's initial temperature. For a tiny sample, the probe can shift the final temperature noticeably.

## 1.4 A probe follows a changing bath with a delay

Hold a bath at temperature $T_b$. Suppose a probe at $T_p$ absorbs heat at a rate $G(T_b-T_p)$, with thermal conductance $G$ in watts per kelvin. The probe has heat capacity $C_p$, so its temperature follows

$$
\begin{aligned}
C_p\frac{\mathrm dT_p}{\mathrm dt} &= G(T_b-T_p),\\
T_p(t)&=T_b+[T_p(0)-T_b]\mathrm e^{-t/\tau},\qquad
\tau=\frac{C_p}{G}.
\end{aligned}
$$

The temperature gap loses the same fraction in each equal time interval. After time $\tau$, the remaining gap is about thirty-seven percent of the initial gap. For a probe that starts at $22\,^\circ\mathrm C$ and enters a bath held at $32\,^\circ\mathrm C$, it reads about $28.32\,^\circ\mathrm C$ after one response time. The curve assumes constant conductance and a fixed bath temperature.

- Maintained bath
- Four-second probe
- Twenty-four-second probe

**Figure 1.1** Calculated thermometer response in a maintained 32 °C bath. Both probes start at 22 °C. Their response times are four and twenty-four seconds.

Open [Electrical Calorimeter](https://sandboxphysics.com/labs/calorimeter/) and choose **Reset**. Under **Raw measurements, probe calibration and model**, find **Probe response in a fixed-temperature bath**. The default record gives a fitted response time of about $4.00\,\mathrm s$. Choose **A slow thermometer** to acquire a second record and compare the step curve. Its fitted response time is about $24.0\,\mathrm s$. The bath inside this separate calibration stays at the same target temperature.

**Table 1.1** Calibrated probe temperatures from the simulated fixed-bath step records. The bath stays at 32 °C. The probe samples include a small reproducible perturbation.

| Time (s) | Default probe (°C) | Slow probe (°C) |
| --- | --- | --- |
| 4 | 28.321 | 23.535 |
| 24 | 31.976 | 28.322 |
| 120 | 31.999 | 31.932 |

Return to the main warming curve. The heater now brings the bath temperature up gradually. In the model, the sample, water and container share a single bath temperature. The probe has its own temperature. After $120\,\mathrm s$ the ideal bath is about $24.57\,^\circ\mathrm C$ for either probe. The calibrated detector readings are about $24.49\,^\circ\mathrm C$ and $24.10\,^\circ\mathrm C$. The lab's **JSON** export keeps the ideal temperatures and the separate raw measurements. During heating, a calibrated thermometer with a longer response time can read farther below the sample temperature.

## 1.5 Exercises

1. A thermometer settles to the same reading in two large, well-stirred baths. State the law that relates the bath temperatures and what contact condition is needed.
2. Convert $25.0\,^\circ\mathrm C$ to kelvins. If the temperature rises by $5.0$ Celsius degrees, what is the rise in kelvins?
3. A sensor gives raw numbers of $15.5$ and $35.7$ at reference temperatures of $15.0\,^\circ\mathrm C$ and $35.0\,^\circ\mathrm C$. Find the gain and intercept. What Celsius temperature corresponds to a raw value of $25.6$?
4. An object of heat capacity $240\,\mathrm{J\,K^{-1}}$ absorbs $1200\,\mathrm J$ of heat without doing appreciable work. Find its temperature change.
5. Mix equal masses of water at $20.0\,^\circ\mathrm C$ and $50.0\,^\circ\mathrm C$ in an insulated container of negligible heat capacity. Find the final temperature and explain why the arithmetic average applies here.
6. Two probes start at $22\,^\circ\mathrm C$ and move to a bath held at $32\,^\circ\mathrm C$. Their response times are four and twenty-four seconds. Which probe is closer to the bath after four seconds? Use Table 1.1 to compare.
7. A small sample feels heat from both a probe and the room. The probe reading is steady. What else must you check before treating it as the sample temperature?
8. A probe trails below a warming sample's temperature, then the heater turns off and the bath cools. Can the probe later read higher than the bath? Explain using the temperature gap.

## 1.6 Answers and explanations

Exercise 1

The zeroth law relates the two baths by thermal equilibrium with the same thermometer. The probe must come to the bath temperature at each contact with negligible influence from its surroundings.

Exercise 2

The absolute temperature is $298.15\,\mathrm K$, or $298.2\,\mathrm K$ to the precision of the Celsius reading. The increase is $5.0\,\mathrm K$ because the offset cancels in a difference.

Exercise 3

The gain is the raw change divided by the reference change, so $a=20.2/20.0=1.01$. Using the first reference gives $b=0.35$. The calibrated value is $(25.6-0.35)/1.01=25.0\,^\circ\mathrm C$.

Exercise 4

The increase is $\Delta T=Q/C=5.0\,\mathrm K$. A positive heat input raises the temperature when the body remains in the same phase and work can be neglected.

Exercise 5

The final temperature is $35.0\,^\circ\mathrm C$. Equal masses of water have equal heat capacities here. Losing and gaining the same energy therefore require the same magnitude of temperature change.

Exercise 6

The default probe is closer. Its gap after four seconds is about $3.68\,\mathrm K$. The slow probe is still about $8.47\,\mathrm K$ below the bath because it has completed only one sixth of its response time.

Exercise 7

Check that the probe is in good thermal contact with the sample and that little heat flows from it to the room or its mount. Check that the probe did not appreciably change the sample temperature. A steady reading can arise when ongoing heat flows balance at the probe.

Exercise 8

Yes. While the bath cools, its temperature can drop below the slower probe. The gap then points from the probe to the bath. The probe gives heat to the bath as it approaches the new temperature.

## Test it in the Lab

Change one setting at a time, then compare what you see with the tables and answers in this chapter.

[Electrical Calorimeter → Heat a sample in a stirred water bath. Record electrical input and thermometer response, compare a separate water-only blank, and recover specific heat with heat-loss and probe-response corrections.](https://sandboxphysics.com/labs/calorimeter/)

## Sources and reading scope

1. [OpenStax, University Physics Volume 2, §1.1 Temperature and Thermal Equilibrium](https://openstax.org/books/university-physics-volume-2/pages/1-1-temperature-and-thermal-equilibrium)

   The section uses the zeroth law to explain how two bodies in thermal equilibrium with a thermometer share a temperature.

2. [OpenStax, University Physics Volume 2, §1.2 Thermometers and Temperature Scales](https://openstax.org/books/university-physics-volume-2/pages/1-2-thermometers-and-temperature-scales)

   The section explains how thermometers use a measurable property and compares Celsius and Kelvin scales.

3. [OpenStax, University Physics Volume 2, §1.4 Heat Transfer, Specific Heat, and Calorimetry](https://openstax.org/books/university-physics-volume-2/pages/1-4-heat-transfer-specific-heat-and-calorimetry)

   Heat capacity connects an energy transfer to a temperature change, and calorimetry uses the energy balance to infer the final temperature.

4. [OpenStax, University Physics Volume 2, §1.6 Mechanisms of Heat Transfer](https://openstax.org/books/university-physics-volume-2/pages/1-6-mechanisms-of-heat-transfer)

   The section distinguishes heat transfer by conduction, convection and radiation and gives examples of what sets the transfer rate.

[← Volume 5 · Heat, Probability and Matter](https://sandboxphysics.com/learn/heat-and-statistics/) [Chapter 2 · How do molecular collisions create pressure? →](https://sandboxphysics.com/learn/heat-and-statistics/molecular-motion-and-pressure/)

Chapters in this volume

- [1. When does a thermometer represent the sample temperature?](https://sandboxphysics.com/learn/heat-and-statistics/temperature-and-thermal-equilibrium/)
- [2. How do molecular collisions create pressure?](https://sandboxphysics.com/learn/heat-and-statistics/molecular-motion-and-pressure/)
- [3. How do heat and work change internal energy?](https://sandboxphysics.com/learn/heat-and-statistics/first-law-of-thermodynamics/)
- [4. Why does heat flow spontaneously from hot to cold?](https://sandboxphysics.com/learn/heat-and-statistics/entropy-and-irreversibility/)
- [5. How much work can a heat engine deliver in one cycle?](https://sandboxphysics.com/learn/heat-and-statistics/heat-engines-and-cycles/)
- [6. How do probabilities determine a system's thermal averages?](https://sandboxphysics.com/learn/heat-and-statistics/boltzmann-weights-and-fluctuations/)

## Keep reading

Interactive Physics Textbook · Reading path

### Heat, Probability and Matter

1. [You are here **When does a thermometer represent the sample temperature?**](https://sandboxphysics.com/learn/heat-and-statistics/temperature-and-thermal-equilibrium/)
2. [Article **How do molecular collisions create pressure?**](https://sandboxphysics.com/learn/heat-and-statistics/molecular-motion-and-pressure/)
3. [Article **How do heat and work change internal energy?**](https://sandboxphysics.com/learn/heat-and-statistics/first-law-of-thermodynamics/)
4. [Article **Why does heat flow spontaneously from hot to cold?**](https://sandboxphysics.com/learn/heat-and-statistics/entropy-and-irreversibility/)
5. [Article **How much work can a heat engine deliver in one cycle?**](https://sandboxphysics.com/learn/heat-and-statistics/heat-engines-and-cycles/)
6. [Article **How do probabilities determine a system's thermal averages?**](https://sandboxphysics.com/learn/heat-and-statistics/boltzmann-weights-and-fluctuations/)

1 more group

### Other articles that use the same experiments

Each experiment from this article that another series also uses, followed by the articles there.

1. [Lab **Electrical Calorimeter**](https://sandboxphysics.com/labs/calorimeter/)
2. [Article **The cup absorbs heat too**](https://sandboxphysics.com/stories/instruments/calorimeter/)