title: "Cryo-EM reconstruction and half-map agreement | Sandbox Physics"
description: "Use projections from multiple directions to fit a three-dimensional map. Compare directional coverage and correlation between two half maps."
canonical_url: "https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/cryo-em-reconstruction/"
language: "en"
page_type: "story"
alternate_languages:
  zh-CN: "https://sandboxphysics.com/zh/learn/experiments-on-the-microscopic-world/cryo-em-reconstruction/"
last_modified: "2026-11-25"
---

[Volume 7 · Learning the Microscopic World from Experiments](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/)

# How do two-dimensional images reveal a particle’s three-dimensional shape?

Use projections from multiple directions to fit a three-dimensional map. Compare directional coverage and correlation between two half maps.

Before you start: [Volume 2 Chapter 6](https://sandboxphysics.com/learn/oscillations-and-waves/standing-waves-and-fourier/) on Fourier modes, [Volume 4 Chapter 4](https://sandboxphysics.com/learn/light-and-imaging/diffraction-and-resolution/) on spatial frequency and resolution, and [Chapter 1](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/signals-calibration-and-background/) and [Chapter 2](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/diffraction-and-crystal-spacing/) on calibration and diffraction.

After this chapter you can: Relate each projection to a Fourier slice and interpret half-map correlation alongside directional coverage.

A microscope can recover a three-dimensional shape by recording many copies of a particle in different orientations. Each view constrains a different slice of the particle’s spatial frequencies. When most copies face one way, two reconstructions can agree closely and still leave the same direction poorly determined.

## 3.1 A projection adds density along the beam

A view along the electron beam collapses depth into a plane. For an ideal projection, let the unit vector $\mathbf n$ point along the beam. Let $\mathbf e_u$ and $\mathbf e_v$ point along the two image axes. The coordinates $u$, $v$ and the distance $t$ along the beam are in nanometres. If $\rho$ is the model’s effective scattering density per unit length, the projected density is its sum along that line.

$$
P_{\mathbf n}(u,v)
=\int\rho(u\mathbf e_u+v\mathbf e_v+t\mathbf n)\,\mathrm dt.
$$

Consider two columns of two equal-depth voxels. The amounts in one column are $3$ and $1$ in the same density units. The other column contains $2$ and $2$. A beam along either column records the same projected sum of $4$. A side view places the two voxels at different positions on the image, where the difference can show up.

## 3.2 Rotating the beam selects another frequency slice

The Fourier transform of a projection gives a central plane of the particle’s three-dimensional Fourier transform. Let $k_u$ and $k_v$ be spatial frequencies along the image axes, measured in cycles per nanometre. Write $\widehat P_{\mathbf n}$ for the two-dimensional transform and $\widehat\rho$ for the three-dimensional transform. In the image transform, the exponential depends on $u$ and $v$ alone. Insert the line integral and combine the three spatial integrals. The result is the volume transform evaluated at a frequency with zero component along $\mathbf n$. [Scheres, A Bayesian View on Cryo-EM Structure Determination](https://extranet.fredhutch.org/content/dam/extranet/u/shared-resources/electron-microscopy/RELION-theory.pdf).

$$
\widehat P_{\mathbf n}(k_u,k_v)
=\widehat\rho(k_u\mathbf e_u+k_v\mathbf e_v).
$$

A beam along the $z$ axis covers the plane $k_z=0$. A beam along the $x$ axis covers $k_x=0$. More independent images from one direction can reduce the noise in that slice by averaging. Views from other directions are needed to add slices through more of the frequency volume.

## 3.3 The reconstruction fits images after alignment and calibration

The electron detector records counts on pixels over several frames. The teaching model supplies the frame shifts, so the images can be moved back to the same pixel coordinates before averaging. A particle picker locates candidate windows in the aligned field. The model estimates defocus from the power spectrum of a support strip whose texture is specified.

The contrast transfer function, or CTF, describes how the optics change the amplitude and sign of a spatial Fourier component. Let $D$ denote the Fourier transform of the calibrated image, $H$ the CTF and $\eta$ the noise. The weak-phase model relates the image to the ideal projection by

$$
D=H\widehat P_{\mathbf n}+\eta.
$$

Each term refers to the same spatial frequency. At a zero of the CTF, the image carries only noise at that frequency in this model. Recording at a defocus that transfers the component can give the reconstruction extra information.

After screening, the teaching model assigns a viewing direction and an in-plane angle by matching each window against the same reference. It starts the three-dimensional map with zero density, makes trial projections through the model optics and compares them with the class averages. Each update reduces a penalized mismatch under a nonnegative density constraint. The reference supplies orientation candidates. The update uses the particle images to estimate density.

## 3.4 Half maps can agree under the same coverage limit

A half map is a reconstruction made from one part of a split collection of particle images. Compare two half maps by taking their Fourier transforms, $A$ and $B$. For a shell $S$ of frequencies at similar distance from the origin, the Fourier shell correlation, or FSC, is the normalized agreement of the complex amplitudes in that shell.

$$
\mathrm{FSC}(S)=
\frac{\operatorname{Re}\sum_{S} A B^*}
{\sqrt{\sum_{S}|A|^2\,\sum_{S}|B|^2}}.
$$

The star denotes complex conjugation. In the teaching model the grid has $16$ voxels across, each $0.8\,\mathrm{nm}$ wide. The fourth shell is labelled by $4/(16\times0.8\,\mathrm{nm})=0.3125\,\mathrm{nm}^{-1}$. Its associated spatial period is $3.20\,\mathrm{nm}$. A shell pools different directions at that scale into one number.

Independent refinement of the two half maps helps prevent common fitting errors from inflating the correlation. The teaching model splits its particle images and fits each half separately, using the same orientation reference and optical calibration. Its FSC describes agreement under those shared choices. Read the directional coverage too, since both halves can lack the same views. [Electron Microscopy Data Bank validation guide](https://www.ebi.ac.uk/emdb/va/help/).

## 3.5 Compare two collections with the same requested particle count

Open [Single-particle Cryo-EM](https://sandboxphysics.com/labs/single-particle-cryo-em/) and select **Find particles and estimate optics**. Record the number of accepted windows, the inferred directions and the viewing-normal eigenvalue ratio. The coverage ratio compares the smallest and largest mean squared projections of the viewing normals onto a unit direction. It is zero when all the normals point along one line.

Predict how many directions will remain if all the particles face the same way. Keep **Independent particle copies** at $96$ and leave the other acquisition settings at their starting values. Set **Copies forced into one viewing direction** to $100\%$. Select **Acquire next movies**, then **Align, classify and reconstruct**. Compare the fourth-shell agreement in the two runs.

- Broad views
- Preferred views

**Figure 3.1** Accepted windows by inferred viewing direction and half-map correlation by spatial frequency. The two simulated records request the same number of particles. One uses the default orientation distribution and the other sets the orientation preference to $100\%$.

**Table 3.1** The saved runs use the same requested count and the same dose, ice and defocus settings. The picker finds $91$ windows in the default run and $96$ in the second run. All picked windows pass screening in these two records.

| Orientation preference | Accepted windows | Inferred directions | Coverage ratio | Fourth-shell FSC |
| --- | --- | --- | --- | --- |
| Default | 91 | 12 | 0.734 | 0.921 |
| One direction | 96 | 1 | 0 | 0.962 |

The preferred orientation leaves a single viewing normal in the data. Both halves then have much the same information about one plane of spatial frequencies. Their agreement is higher on the fourth shell in this run. The direction counts show which views to acquire next.

## 3.6 Exercises

1. Two equal-depth columns have voxel amounts $(3,1)$ and $(2,2)$. Why can a side view distinguish them if the view along the columns gives the same sum?
2. You have beams along $x$ and $z$. Does either central plane contain the frequency $(0.10,0.10,0.10)\,\mathrm{nm}^{-1}$?
3. Compute the frequency and period at shell $4$ for a grid $16$ voxels wide with pitch $0.8\,\mathrm{nm}$.
4. A shell has two real coefficients. The two half maps give $(1,2)$ and $(2,1)$. Compute the FSC.
5. You average sixteen copies at one orientation instead of four. If each copy has independent noise with the same variance, how does the standard deviation of the mean change?
6. The CTF vanishes at a particular spatial frequency. What kind of additional acquisition can supply that component?
7. In Figure 3.1, which collection would you extend by changing viewing directions? Explain what you expect to happen to the occupied direction count.
8. A reconstruction program repeatedly refines one shared reference from all the particle images before calculating half maps. How would you redesign the comparison to reduce the risk of common fitting errors?

Exercise 1

The sums along the columns equal $4$. A side projection assigns the separate amounts to separate image positions. The first image shows unequal amounts there, and the second shows equal amounts.

Exercise 2

The planes are $k_x=0$ and $k_z=0$. The specified point has positive values for both components. A view whose normal is perpendicular to that frequency vector would put it in the central plane.

Exercise 3

The frequency is $4/(16\times0.8\,\mathrm{nm})=0.3125\,\mathrm{nm}^{-1}$. Its reciprocal is $3.20\,\mathrm{nm}$. This period labels a frequency shell shared by many directions.

Exercise 4

The numerator is $1\times2+2\times1=4$. Each vector has squared length $5$, so their length product is $5$. The FSC is $4/5=0.80$.

Exercise 5

Dividing the variance of one copy by the number of independent copies gives the variance of the mean. The standard deviation is proportional to $1/\sqrt N$, so it halves. The average still uses the same viewing direction.

Exercise 6

Pick another defocus whose CTF is nonzero at that frequency. A calibrated image at that defocus gives the corresponding projection component a nonzero weight in the fit.

Exercise 7

Add new directions to the preferred collection. Its $96$ accepted windows occupy just one direction, so the new views can populate the missing classes. If the assignments are correct, a count above one gives the fit images from additional frequency planes.

Exercise 8

Split the particle images before refining either orientation reference. Refine the halves independently and compare their maps after the fits. Then inspect the angles as well as the FSC to check which directions support the agreement.

## Test it in the Lab

Change one setting at a time, then compare what you see with the tables and answers in this chapter.

[Single-particle Cryo-EM: From Movies to a Map → Find particles in a counted micrograph, estimate the CTF and reconstruct their map. Compare wrong optics, false picks and restricted reference views with genuinely missing viewing directions.](https://sandboxphysics.com/labs/single-particle-cryo-em/)

## Sources and reading scope

1. [Sjors H. W. Scheres, A Bayesian View on Cryo-EM Structure Determination](https://extranet.fredhutch.org/content/dam/extranet/u/shared-resources/electron-microscopy/RELION-theory.pdf)

   The image model uses a Fourier slice through the volume and multiplies it by the contrast transfer function for each image.

2. [Electron Microscopy Data Bank, Validation Analysis Help](https://www.ebi.ac.uk/emdb/va/help/)

   The guide explains how the FSC plot compares independently refined half maps at different spatial frequencies and how masks affect the comparison.

[← Chapter 2 · How does diffraction reveal crystal plane spacing?](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/diffraction-and-crystal-spacing/) [Chapter 4 · How does an X-ray spectrum identify elements? →](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/xray-fluorescence-and-element-lines/)

Chapters in this volume

- [1. What does an instrument record directly?](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/signals-calibration-and-background/)
- [2. How does diffraction reveal crystal plane spacing?](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/diffraction-and-crystal-spacing/)
- [3. How do two-dimensional images reveal a particle’s three-dimensional shape?](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/cryo-em-reconstruction/)
- [4. How does an X-ray spectrum identify elements?](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/xray-fluorescence-and-element-lines/)
- [5. How do counts reveal neutrino oscillations?](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/counts-and-neutrino-oscillations/)
- [6. How does the end of a beta spectrum constrain neutrino mass?](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/beta-endpoint-and-neutrino-mass/)

## Keep reading

Interactive Physics Textbook · Reading path

### Learning the Microscopic World from Experiments

1. [Article **What does an instrument record directly?**](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/signals-calibration-and-background/)
2. [Article **How does diffraction reveal crystal plane spacing?**](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/diffraction-and-crystal-spacing/)
3. [You are here **How do two-dimensional images reveal a particle’s three-dimensional shape?**](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/cryo-em-reconstruction/)
4. [Article **How does an X-ray spectrum identify elements?**](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/xray-fluorescence-and-element-lines/)
5. [Article **How do counts reveal neutrino oscillations?**](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/counts-and-neutrino-oscillations/)
6. [Article **How does the end of a beta spectrum constrain neutrino mass?**](https://sandboxphysics.com/learn/experiments-on-the-microscopic-world/beta-endpoint-and-neutrino-mass/)